Valiant Navion™ VNMC4337C200TU (UDI 00763000261054)Recalls & reported problems
Medtronic, Inc. · STENT GRAFT VNMC4337C200TU VAL NAV CS TP · Class 3 · Implant
All models of Valiant Navion (137)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00763000261054), received between Feb 16, 2021 and Feb 16, 2021 — classified as 1 injury.
The most frequently reported specific device problem is “Malposition of Device” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Feb 12, 2021Class IOpen, Classified
Due to the presence of type IIIb endoleaks, stent fractures, and stent ring enlargement.
Z-1201-2021 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Malposition of Device1
- Material Deformation1
Patient problems
- Insufficient Information1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryFeb 16, 2021Malposition of Device; Material Deformation; Malposition of Device; Material Deformation
A VALIANT NAVION STENT GRAFT WAS ATTEMPTED TO BE IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF A 30 MM THORACIC AORTIC DISSECTION. IT WAS REPORTED THAT DURING THE INDEX PROCEDURE, A VNMC4337C200TU TO TREAT A TBAD. THE PROXIMAL TEAR WAS 7CM FROM THE LSA, AND THE PROXIMAL LANDING ZONE WAS AT THE LSA. THE DEVICE WAS DEPLOYED ACCORDING TO IFU, BUT THE BLOOD PRESSURE REMAINED AT 114 MM/HG. IT WAS STATED THAT THE DEPLOYMENT OF THE GRAFT AND THE NOSE CONE WERE COMPLETED WITHOUT INCIDENT, AND THE DELIVERY SYSTEM WAS REMOVED UNDER FLURO WITHOUT NOTICEABLY GETTING CAUGHT ON THE GRAFT. AN ANGIOGRAM WAS PERFORMED, WHERE IT WAS NOTED THAT PROXIMAL END OF GRAFT APPEARING TO BE MISALIGNED AND SEVERAL STENT SEGMENTS BUNCHES UP AND TWISTED ALONG THE INNER CURVE. INSTEAD OF THE 3 PROXIMAL RO MARKETS ALIGNING, ONE OF THE RO MARKETS APPEARED SIGNIFICANT LH FURTHER BACK FROM THE OTHER 2, DESPITE THE GRAFT BEING DEPLOYED WITH PARALLAX ELIMINATED. THE PHYSICIAN THEN DECIDED TO IMPLANT A SECOND VNMC4337C200TU TO RELINE THE FIRST GRAFT TO ENSURE THE AORTA WAS WIDELY PATENT AND THE FALSE LUMEN WAS CLOSED OFF. THE SECOND DEVICE WAS DEPLOYED WITHOUT INCIDENT AND THE FINAL ANGIE SHOWED SUCCESSFUL EX
Patient problems: Insufficient Information
MAUDE report 11332231 ↗
Other SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page