PipelineTM Flex Embolization Device with Shield TechnologyTM PED2-250-10Recalls & reported problems
Micro Therapeutics, Inc. · STENT PED2-250-10 V06 E-IFU · Class 3 · Implant
All models of PipelineTM Flex Embolization Device with Shield TechnologyTM (79)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00763000283834), received between Apr 1, 2022 and Jul 22, 2022 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
12 other recalls cover devices cleared under the same FDA submission (P100018); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Break1
- Physical Resistance/Sticking1
Patient problems
- High Blood Pressure/ Hypertension1
- Hypersensitivity/Allergic reaction1
- No Clinical Signs, Symptoms or Conditions1
- Paralysis1
- Swelling/ Edema1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 22, 2022Break; Physical Resistance/Sticking; Break; Physical Resistance/Sticking
MEDTRONIC RECEIVED A REPORT THAT THE PIPELINE EXPERIENCED RESISTANCE IN THE DISTAL PART OF THE PHENOM 27 MICROCATHETER AND THE PUSHWIRE SEPARATED. THE PATIENT WAS UNDERGOING TREATMENT FOR AN UNRUPTURED, SACCULAR ANEURYSM LOCATED IN THE RIGHT MIDDLE CEREBRAL ARTERY (MCA). THE MAX DIAMETER WAS 3MM, AND THE NECK DIAMETER WAS 2MM. THE PATIENT'S VESSEL TORTUOSITY WAS MODERATE. THE LANDING ZONE WAS 1.9MM DISTAL AND 2.7MM PROXIMAL. THE ACCESS VESSEL WAS THE RIGHT MCA, WHICH WAS 2MM. IT WAS REPORTED THAT THERE WAS EXTREME PRESSURE WHEN PUSHING THE DEVICE UP THROUGH THE PHENOM 27 MICROCATHETER. THEY UN-SHEATHED THE TIP COIL AND FELT EXTREME RESISTANCE DEPLOYING DEVICE. THE DEVICE FELT LIKE IT LOCKED UP ON IN THE PHENOM 27. THE PHYSICIAN RELEASED THE LOAD/SLACK IN THE SYSTEM, BUT HIS DID NOT RESOLVE THE ISSUE. THE DISTAL PART OF THE MICROCATHETER WAS KINKED, AND THE DISTAL PART OF THE PIPELINE PUSHWIRE SEPARATED. THE PHYSICIAN DECIDED TO ABORT DEPLOYMENT AND PULLED THE SYSTEM OUT, AND CUT THE MICROCATHETER ON THE BACK TABLE TO EXPOSE DEVICE. THE PATIENT WAS TREATED WITH STENT COIL. THE PATIENT DID NOT EXPERIENCE ANY INJURY OR COMPLICATIONS. THE DEVICES WERE PREPARED AND FLUSHED ACCORDING TO
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15086110 ↗InjuryApr 1, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
MEDTRONIC RECEIVED A REPORT THAT A PIPELINE PATIENT EXPERIENCED HEMIPARESIS FOLLOWING THE PROCEDURE. THE PATIENT WAS UNDERGOING TREATMENT FOR AN UNRUPTURED, RECURRENT NECK ANEURYSM LOCATED IN THE RIGHT ANTERIOR COMMUNICATING ARTERY THAT HAD BEEN PREVIOUSLY COILED ANEURYSM. THE PATIENT WAS KNOWN TO HAVE FIBROMUSCULAR DYSPLASIA (FMD) IN THE PROXIMAL INTERNAL CAROTID ARTERY. THE MAX DIAMETER WAS 2.75MM, AND THE NECK DIAMETER WAS 3.5MM. THE PATIENT'S VESSEL TORTUOSITY WAS SEVERE. THE LANDING ZONE WAS 2.7MM DISTAL AND 2.5MM PROXIMAL. DUAL ANTIPLATELET TREATMENT WAS ADMINISTERED. IT WAS REPORTED THAT THE PATIENT WAS TREATED WITH VERAPAMIL DUE TO THE PROXIMAL FMD. THE PIPELINE WAS PLACED WITHOUT INCIDENT OR TECHNICAL DIFFICULTIES. HOWEVER, UPON FOLLOW UP THE NEXT DAY, THE PATIENT WAS UNABLE TO MOVE THEIR LEFT ARM AND ALSO HAD SOME HYPERPERFUSION. THE DOCTOR SUSPECTED THAT A RESPONSE TO THE USE OF THE PIPELINE OR THE HYDROPHILIC COATING OF THE PHENOM 27 MICROCATHETER MAY HAVE CAUSED THE ISSUE. THE DOCTOR HAD SEEN THIS PREVIOUSLY, WITH EXTRAVASATION OF FLUID AND SWELLING FROM THE PARENT TREATED VESSEL DIRECTLY DISTAL TO THE PLACEMENT OF THE PIPELINE 3-13 HOURS FOLLOWING IMPLANT. IN ALL CASE
Patient problems: High Blood Pressure/ Hypertension; Hypersensitivity/Allergic reaction; Paralysis; Swelling/ Edema
MAUDE report 13984638 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P100018. One submission often covers many models, so check the product and lot details in each record.
- Z-1246-2025 · Jan 30, 2025 · Open, ClassifiedMedtronic Pipeline Vantage Embolization Device with Shield Technology, REF: PED3-021-350-16, PED3-021-250-14, PED3-021-350-20, PED3-021-250-20, PED3-021-300-16, PED3-021-300-20, PE
- Z-1245-2025 · Jan 30, 2025 · Open, ClassifiedMedtronic Pipeline Vantage Embolization Device with Shield Technology, REF: PED3-027-350-12, PED3-027-350-14, PED3-027-350-16, PED3-027-350-20, PED3-027-350-25, PED3-027-400-12, PE
- Z-2403-2021 · Jul 13, 2021 · Open, ClassifiedEV3 Pipeline Flex Embolization Device with Flex Shield Technology
- Z-2402-2021 · Jul 13, 2021 · Open, ClassifiedPipeline Flex Embolization Device
- Z-1527-2020 · Feb 14, 2020 · Terminatedev3 Pipeline Flex Embolization Device. For neurological endovascular use.
- Z-0307-2017 · Oct 5, 2016 · TerminatedPipeline Embolization Device (Pipeline Classic) The device is indicated for the endovascular treatment of adults (22 year of age or older) with large or giant widenecked intracrani
- Z-0308-2017 · Oct 5, 2016 · TerminatedAlligator Retrieval Device The device is intended for use in the peripheral and neurovasculature for foreign body removal. Catalog No. FA-88810-20V04 FA-88810-20V06 FA-88810-30V04
- Z-0309-2017 · Oct 5, 2016 · TerminatedMarathon Flow Directed Micro Catheter, UltraFlow HPC Flow Directed Micro Catheter Marathon Flow: The device is intended to access periphera and neurovasculature for the controlled
- Z-0310-2017 · Oct 5, 2016 · TerminatedX-Celerator Hydrophilic Exchange Guidewire Catalog No. 103-0601-300 103-0601-350V10 103-0602-350V06 103-0601-300V01 103-0602-300 103-0602-350V09 103-0601-300V02 103-0602-300V01 10
- Z-0120-2016 · Sep 4, 2015 · Terminated-Pipeline Embolization Device -Pipeline Flex Embolization Device The Pipeline and Pipeline Flex Embolization Devices consist of a permanent implant combined with a guidewire based
- Z-1453-2014 · Apr 1, 2014 · TerminatedPipeline Embolization Device (PED). Used endovascular treatment of adults with intracranial aneurysms in the internal carotid artery.
- Z-2045-2013 · Jul 19, 2013 · TerminatedPipeline Embolization Device (PED) The Pipeline Embolization Device is indicated for the endovascular treatment of adults (22 years of age and older) with large or giant wide-neck
Other Intracranial aneurysm flow diverter devices
- Pipeline TM Flex Embolization Device with Shield Technology TM PED2-300-12Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter2 recalls1 report
- PIPELINE PED-500-20Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall31 reports
- PIPELINE PED-500-35Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall1 death29 reports
- PIPELINE PED-500-16Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall2 deaths27 reports
- PIPELINE PED-500-25Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall1 death23 reports
- PIPELINE PED-500-30Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall21 reports
- PIPELINE PED-450-25Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall20 reports
- PIPELINE PED-475-30Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall1 death19 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page