Skip to content

PipelineTM Flex Embolization Device with Shield TechnologyTM PED2-375-20Recalls & reported problems

Micro Therapeutics, Inc. · STENT PED2-375-20 V06 E-IFU · Class 3 · Implant

All models of PipelineTM Flex Embolization Device with Shield TechnologyTM (79)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00763000284275), received between Feb 17, 2023 and Jan 21, 2026 — classified as 2 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Activation Failure” (3 reports).

12 other recalls cover devices cleared under the same FDA submission (P100018); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Activation Failure3
  • Material Deformation3
  • Adverse Event Without Identified Device or Use Problem2
  • Improper or Incorrect Procedure or Method1
  • No Apparent Adverse Event1
  • Physical Resistance/Sticking1
  • Structural Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Vasoconstriction2

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 21, 2026Improper or Incorrect Procedure or Method; Material Deformation; Activation Failure

    ADDITIONAL INFORMATION RECEIVED REPORTED THAT THE CAUSE OF THE STENT NOT OPENING, EVEN AFTER RETRIEVING THE DAMAGE PART, WAS UNDETERMINED. IT WAS CLARIFIED THAT THE MIDDLE SECTION OF THE PIPELINE DID NOT OPEN, THE PIPELINE HAD NOT BEEN PLACED IN A VESSEL BEND WHEN IT FAILED TO OPEN, AND THERE WERE NO ADDITIONAL STEPS OR OTHER DEVICES ATTEMPTED TO OPEN THE PIPELINE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24132266 ↗
  • MalfunctionMar 13, 2025Material Deformation; Activation Failure; Material Deformation; Activation Failure

    ADDITIONAL INFORMATION RECEIVED REPORTED THAT THE CAUSE OF THE FAILURE TO OPEN/PIPELINE DAMAGE WAS NOT DETERMINED. THE PIPELINE HAD NOT BEEN POSITIONED IN A VESSEL BEND WHEN IT FAILED TO OPEN. THEY HAD PUSHED THE INTERMEDIATE CATHETER TO GO UPPER IN AN ATTEMPT TO OPEN THE PIPELINE. THERE WAS NO DAMAGE OR ISSUE WITH THE PHENOM CATHETER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21593388 ↗
  • MalfunctionNov 22, 2024Structural Problem; Material Deformation; Activation Failure; Structural Problem; Material Deformation; Activation Failure

    IT WAS REPORTED THAT DURING A NEUROVASCULAR FLOW DIVERSION PROCEDURE USING A PIPELINE EMBOLIZATION DEVICE, THE MIDDLE SECTION OF THE STENT BECAME KINKED DURING DEPLOYMENT. THE PROCEDURE TOOK PLACE AT THE C6 OPHTHALMIC ARTERY FOR AN UNRUPTURED, AMORPHOUS ANEURYSM WITH A MAXIMUM DIAMETER OF 4.3 MILLIMETERS AND A NECK DIAMETER OF 1.88 MILLIMETERS. THE DISTAL LANDING ZONE MEASURED 3.5 MILLIMETERS AND THE PROXIMAL LANDING ZONE MEASURED 3.77 MILLIMETERS, WITH MODERATE VESSEL TORTUOSITY. POST-PROCEDURE, SLOWED BLOOD FLOW WAS OBSERVED. DUAL ANTIPLATELET TREATMENT WAS ADMINISTERED, THOUGH THE PLATELET REACTIVITY UNIT LEVEL WAS UNKNOWN. THE STENT WAS RETRIEVED AND REDEPLOYED, BUT IT DID NOT OPEN PROPERLY. UPON REMOVAL FROM THE BODY, THE MIDDLE SECTION OF THE STENT WAS SLIGHTLY DEFORMED. THE STENT WAS REPLACED, AND THE SURGERY WAS SUCCESSFULLY COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20758796 ↗
  • InjuryApr 29, 2024Adverse Event Without Identified Device or Use Problem; No Apparent Adverse Event; Adverse Event Without Identified Device or Use Problem; No Apparent Adverse Event

    MEDTRONIC RECEIVED A REPORT THAT A PIPELINE PATIENT EXPERIENCED A CEREBRAL VASOSPASM. THE SUBJECT EXPERIENCED MILD TO MODERATE VASOSPASMS IN BILATERAL ICAS. THE PATIENT WAS TREATED WITH VERAPAMIL, RECOVERED APPROPRIATELY. END DATE OF THE HOSPITALIZATION: (B)(6) 2022.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Vasoconstriction

    MAUDE report 19203572 ↗
  • MalfunctionMar 23, 2023Physical Resistance/Sticking; Physical Resistance/Sticking

    MEDTRONIC RECEIVED A REPORT THAT THE PIPELINE WAS STUCK IN THE PHENOM. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF A SACCULAR, UNRUPTURED RIGHT SUPER CLINOID ICA ANEURYSM WITH A MAX DIAMETER OF 3.5MM AND A 5MM NECK DIAMETER. THE LANDING ZONE WAS 3.25MM AT THE DISTAL END AND 3.6MM AT THE PROXIMAL END. IT WAS NOTED THE PATIENT'S VESSEL TORTUOSITYWAS MINIMAL. THE ACCESS VESSEL WAS THE ICA WITH 3.6MM DIAMETER. DAPT (DUAL ANTIPLATELET TREATMENT) WAS ADMINISTERED, PRU LEVEL WAS THERPEUTIC. NO PATIENT SYMPTOMS OR FURTHER COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT. IT WAS REPORTED THAT THE DEVICE WAS EXTREMELY HARD TO PUSH THROUGH THE PHENOM27, AND THE DOCTOR COULDN¿T GET IT TO EXIT THE CATHETER SO THE DOCTOR TRIED TO REMOVE THE PIPELINE THEN IT GOT STUCK IN THE CATHETER HUB - THEY REMOVED THE PHENOM AND PIPELINE. THE PIPELINE ENCOUNTERED RESISTANCE IN THE MIDDLE SECTION OF THE CATHETER. THE CATHETER WAS FLUSHED CONTINUOUSLY WITH HEPARANIZED SALINE. THE PIPELINE THEN BECAME STUCK AT THE CATHETER HUB DURING DELIVERY. THE PIPELINE NEVER EXITED THE CATHETER INTO THE BODY, COULDN'T GET THE DEVICE TO PUSH TO THE END OF THE CATHETER TO EXIT AND DEPLOY. THE CATHETER AND PUSHWIR

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16601206 ↗
  • InjuryFeb 17, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE EVENT RESULTED IN A PROLONGATION OF EXISTING HOSPITALIZATION, WHICH ENDED ON (B)(6) 2022.

    Patient problems: Vasoconstriction

    MAUDE report 16399510 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P100018. One submission often covers many models, so check the product and lot details in each record.

  • Z-1245-2025 · Jan 30, 2025 · Open, Classified
    Medtronic Pipeline Vantage Embolization Device with Shield Technology, REF: PED3-027-350-12, PED3-027-350-14, PED3-027-350-16, PED3-027-350-20, PED3-027-350-25, PED3-027-400-12, PE
  • Z-1246-2025 · Jan 30, 2025 · Open, Classified
    Medtronic Pipeline Vantage Embolization Device with Shield Technology, REF: PED3-021-350-16, PED3-021-250-14, PED3-021-350-20, PED3-021-250-20, PED3-021-300-16, PED3-021-300-20, PE
  • Z-2402-2021 · Jul 13, 2021 · Open, Classified
    Pipeline Flex Embolization Device
  • Z-2403-2021 · Jul 13, 2021 · Open, Classified
    EV3 Pipeline Flex Embolization Device with Flex Shield Technology
  • Z-1527-2020 · Feb 14, 2020 · Terminated
    ev3 Pipeline Flex Embolization Device. For neurological endovascular use.
  • Z-0310-2017 · Oct 5, 2016 · Terminated
    X-Celerator Hydrophilic Exchange Guidewire Catalog No. 103-0601-300 103-0601-350V10 103-0602-350V06 103-0601-300V01 103-0602-300 103-0602-350V09 103-0601-300V02 103-0602-300V01 10
  • Z-0308-2017 · Oct 5, 2016 · Terminated
    Alligator Retrieval Device The device is intended for use in the peripheral and neurovasculature for foreign body removal. Catalog No. FA-88810-20V04 FA-88810-20V06 FA-88810-30V04
  • Z-0309-2017 · Oct 5, 2016 · Terminated
    Marathon Flow Directed Micro Catheter, UltraFlow HPC Flow Directed Micro Catheter Marathon Flow: The device is intended to access periphera and neurovasculature for the controlled
  • Z-0307-2017 · Oct 5, 2016 · Terminated
    Pipeline Embolization Device (Pipeline Classic) The device is indicated for the endovascular treatment of adults (22 year of age or older) with large or giant widenecked intracrani
  • Z-0120-2016 · Sep 4, 2015 · Terminated
    -Pipeline Embolization Device -Pipeline Flex Embolization Device The Pipeline and Pipeline Flex Embolization Devices consist of a permanent implant combined with a guidewire based
  • Z-1453-2014 · Apr 1, 2014 · Terminated
    Pipeline Embolization Device (PED). Used endovascular treatment of adults with intracranial aneurysms in the internal carotid artery.
  • Z-2045-2013 · Jul 19, 2013 · Terminated
    Pipeline Embolization Device (PED) The Pipeline Embolization Device is indicated for the endovascular treatment of adults (22 years of age and older) with large or giant wide-neck

Other Intracranial aneurysm flow diverter devices

All Intracranial aneurysm flow diverter devices →

Sources and updates