PipelineTM Flex Embolization Device with Shield TechnologyTM PED2-400-30Recalls & reported problems
Micro Therapeutics, Inc. · STENT PED2-400-30 V06 E-IFU · Class 3 · Implant
All models of PipelineTM Flex Embolization Device with Shield TechnologyTM (79)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 1
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00763000284381), received between May 29, 2025 and Dec 4, 2025 — classified as 1 death, 2 malfunctions.
The most frequently reported specific device problem is “Activation Failure” (2 reports).
12 other recalls cover devices cleared under the same FDA submission (P100018); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Activation Failure2
- Activation, Positioning or Separation Problem1
- Material Deformation1
- Material Twisted/Bent1
- Physical Resistance/Sticking1
- Positioning Failure1
- Retraction Problem1
- Structural Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Hemorrhage/Blood Loss/Bleeding1
- High Blood Pressure/ Hypertension1
- Intracranial Hemorrhage1
- Rupture1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 4, 2025Activation Failure; Activation Failure
MEDTRONIC RECEIVED A REPORT THAT THE PATIENT WAS UNDERGOING TREATMENT FOR A SACCULAR, UNRUPTURED LEFT CAVERNOUS SINUS SEGMENT ANEURYSM WITH A MAX DIAMETER OF 12.52 MM AND A 9.27 MM NECK DIAMETER LOCATED AT THE C4 SEGMENT OF THE INTRACRANIAL INTERNAL CAROTID ARTERY. IT WAS NOTED THAT THE PATIENT'S VESSEL TORTUOSITY WAS SEVERE. THE LANDING ZONE WAS 4.06MM DISTALLY AND 3.93MM PROXIMALLY. THE ACCESS VESSEL WAS THE FEMORAL ARTERY, WITH 9.86MM DIAMETER. DUAL ANTIPLATELET TREATMENT (DAPT) WAS ADMINISTERED. THE PRU LEVEL WAS NORMAL. THE ANGIOGRAPHIC RESULT POST PROCEDURE WAS WELL. IT WAS REPORTED THAT A PED2-400-30 WAS SELECTED FOR IMPLANTATION. AFTER ADEQUATE HYDRATION, THE PED2-400-30 WAS DELIVERED TO THE STENT CATHETER. HOWEVER, THE DENSE MESH STENT DID NOT OPEN WELL AT THE MIDSECTION BEND. THE OPERATOR MADE MULTIPLE ATTEMPTS, INCLUDING REPEATED PUSHING, PULLING, REPOSITIONING, AND REDEPLOYING, BUT THE STENT STILL DID NOT OPEN PROPERLY. CONSIDERING THE POSSIBILITY OF INTIMAL INJURY DUE TO REPEATED MANIPULATION AND THE COMPLEXITY OF THE PROCEDURE, THE STENT AND MICROCATHETER WERE WITHDRAWN. AFTER DISCUSSING WITH THE PATIENT'S FAMILY, THE SURGERY WAS CANCELED. THE PIPELINE WAS USED FOR AN
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23715701 ↗MalfunctionNov 3, 2025Material Deformation; Material Twisted/Bent; Activation Failure; Material Deformation; Material Twisted/Bent; Activation Failure
ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DURING STENT DEPLOYMENT, THE STENT TWISTED AND OPENED ABNORMALLY. THE MIDDLE SECTION OF THE PIPELINE DID NOT OPEN. THE PIPELINE HAD NOT BEEN PLACED IN A VESSEL BEND WHEN IT FAILED TO OPEN. THERE WAS NO FRICTION OR DIFFICULTY DURING DELIVERY OR POSITIONING. THERE WERE NO DEVICE ISSUES WITH THE PHENOM 27. THE CAUSE OF THE EVENT WAS NOT DETERMINED. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT THE PHYSICIAN PUSHED THE INTERMEDIATE CATHETER TO GO UPPER IN ANOTHER ATTEMPT TO OPEN THE PIPELINE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23451213 ↗DeathMay 29, 2025Positioning Failure; Retraction Problem; Structural Problem; Activation, Positioning or Separation Problem; Physical Resistance/Sticking; Positioning Failure; Retraction Problem; Structural Problem; Activation, Positioning or Separation Problem; Physical Resistance/Sticking
ADDITIONAL INFORMATION WAS RECEIVED REPORTING THAT THE PROBLEM OCCURRED THAT DAY AND THE PATIENT DIED. THE PATIENT PASSED AWAY AT (B)(6), DOCTOR INFORMED THAT THE PATIENT DIED IN THE HOSPITAL. NO AUTOPSY RECORDS WILL BE AVAILABLE. THE PATIENT'S DEATH MAY BE RELATED TO THE PIPELINE SHIELD, AND IT MAY ALSO BE RELATED TO THE OPERATOR'S OPERATION. THERE WAS INTRACRANIAL ANEURYSM LEADING UP TO THE PATIENT'S DEATH. THE CAUSE OF THE RESISTANCE DURING RETRIEVAL MAY BE VASCULAR TORTUOSITY.
Patient problems: Hemorrhage/Blood Loss/Bleeding; High Blood Pressure/ Hypertension; Intracranial Hemorrhage; Rupture
MAUDE report 22104668 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P100018. One submission often covers many models, so check the product and lot details in each record.
- Z-1245-2025 · Jan 30, 2025 · Open, ClassifiedMedtronic Pipeline Vantage Embolization Device with Shield Technology, REF: PED3-027-350-12, PED3-027-350-14, PED3-027-350-16, PED3-027-350-20, PED3-027-350-25, PED3-027-400-12, PE
- Z-1246-2025 · Jan 30, 2025 · Open, ClassifiedMedtronic Pipeline Vantage Embolization Device with Shield Technology, REF: PED3-021-350-16, PED3-021-250-14, PED3-021-350-20, PED3-021-250-20, PED3-021-300-16, PED3-021-300-20, PE
- Z-2402-2021 · Jul 13, 2021 · Open, ClassifiedPipeline Flex Embolization Device
- Z-2403-2021 · Jul 13, 2021 · Open, ClassifiedEV3 Pipeline Flex Embolization Device with Flex Shield Technology
- Z-1527-2020 · Feb 14, 2020 · Terminatedev3 Pipeline Flex Embolization Device. For neurological endovascular use.
- Z-0309-2017 · Oct 5, 2016 · TerminatedMarathon Flow Directed Micro Catheter, UltraFlow HPC Flow Directed Micro Catheter Marathon Flow: The device is intended to access periphera and neurovasculature for the controlled
- Z-0310-2017 · Oct 5, 2016 · TerminatedX-Celerator Hydrophilic Exchange Guidewire Catalog No. 103-0601-300 103-0601-350V10 103-0602-350V06 103-0601-300V01 103-0602-300 103-0602-350V09 103-0601-300V02 103-0602-300V01 10
- Z-0308-2017 · Oct 5, 2016 · TerminatedAlligator Retrieval Device The device is intended for use in the peripheral and neurovasculature for foreign body removal. Catalog No. FA-88810-20V04 FA-88810-20V06 FA-88810-30V04
- Z-0307-2017 · Oct 5, 2016 · TerminatedPipeline Embolization Device (Pipeline Classic) The device is indicated for the endovascular treatment of adults (22 year of age or older) with large or giant widenecked intracrani
- Z-0120-2016 · Sep 4, 2015 · Terminated-Pipeline Embolization Device -Pipeline Flex Embolization Device The Pipeline and Pipeline Flex Embolization Devices consist of a permanent implant combined with a guidewire based
- Z-1453-2014 · Apr 1, 2014 · TerminatedPipeline Embolization Device (PED). Used endovascular treatment of adults with intracranial aneurysms in the internal carotid artery.
- Z-2045-2013 · Jul 19, 2013 · TerminatedPipeline Embolization Device (PED) The Pipeline Embolization Device is indicated for the endovascular treatment of adults (22 years of age and older) with large or giant wide-neck
Other Intracranial aneurysm flow diverter devices
- Pipeline TM Flex Embolization Device with Shield Technology TM PED2-300-12Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter2 recalls1 report
- PIPELINE PED-500-20Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall31 reports
- PIPELINE PED-500-35Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall1 death29 reports
- PIPELINE PED-500-16Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall2 deaths27 reports
- PIPELINE PED-500-25Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall1 death23 reports
- PIPELINE PED-500-30Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall21 reports
- PIPELINE PED-450-25Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall20 reports
- PIPELINE PED-475-30Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall1 death19 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page