Skip to content

In.Pact™ 018 IPU05015013PRecalls & reported problems

Medtronic, Inc. · DCB 018 IPU05015013P D5.0 L150 UL0130 · Class 3

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00763000291488), received between Feb 18, 2026 and Sep 21, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Burst Container or Vessel” (2 reports).

1 other recall cover devices cleared under the same FDA submission (P140010); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Burst Container or Vessel2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 21, 2026Burst Container or Vessel

    It was reported that during the procedure, during the first balloon inflation at nominal pressure, the InPact 018 drug coated balloon incurred a pinhole rupture. The pressure was slowly increased up to 6 atmospheres. Calcification and mild tortuosity were reported. A 6Fr 45cm non-Medtronic sheath and a non-Medtronic guidewire were used. Embolic protection was not used. No damage was noted to the packaging and no issues were noted when removing the device from the hoop/tray. The device was prepped as per the IFU. The balloon did not detach/fragment. The balloon did not pass through a previously deployed stent. No resistance was encountered during delivery. The procedure was completed using a non-Medtronic balloon. No patient symptoms or complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50837762 ↗
  • MalfunctionFeb 18, 2026Burst Container or Vessel

    IT WAS REPORTED THAT DURING A PROCEDURE TO TREAT PLAQUE IN THE MID SUPERFICIAL FEMORAL ARTERY USING THE INPACT 018 DRUG COATED BALLOON, A PIN HOLE BALLOON BURST OCCURRED DURING THE FIRST INFLATION. A 6FR NON-MEDTRONIC SHEATH AND A 018 GUIDEWIRE WERE USED. THE VESSEL WAS PRE-DILATED. 50/50 CONTRAST INFLATION FLUID WAS USED. THE DEVICE WAS NOT PASSED THROUGH A PREVIOUSLY DEPLOYED STENT. NO RESISTANCE WAS ENCOUNTERED WHEN ADVANCING THE DEVICE. THE BALLOON DID NOT DETACH DURING THE EVENT. NO CALCIUM WAS NOTED IN THE AREA OF INFLATION. THE BALLOON WAS SAFELY REMOVED FROM THE PATIENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24375820 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P140010. One submission often covers many models, so check the product and lot details in each record.

  • Z-1042-2022 · Mar 23, 2022 · Open, Classified
    Medtronic IN.PACT Admiral, Paclitaxel-coated PTA Balloon Catheter, Sterile EO, Rx Only for the following model Numbers:ADM04004013P ADM04008013P ADM04008013P ADM04008013P ADM040120

Other Drug-Eluting Peripheral Transluminal Angioplasty Catheter devices

All Drug-Eluting Peripheral Transluminal Angioplasty Catheter devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page