Skip to content

Stealth Autoguide™ System 29631Recalls & reported problems

Medtronic Navigation, Inc. · SYSTEM 29631 STEALTH AUTOGUIDE · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00763000293659), received between Aug 21, 2020 and Jun 26, 2025 — classified as 2 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Application Program Freezes or Fails to Launch” (1 report).

1 other recall cover devices cleared under the same FDA submission (K191597); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Application Program Freezes or Fails to Launch1
  • Device Sensing Problem1
  • Environmental Compatibility Problem1
  • Mechanical Problem1
  • Positioning Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 26, 2025Positioning Problem; Positioning Problem

    MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED DURING A CRANIAL BIOPSY PROCEDURE. IT WAS REPORTED THAT THE AUTOMATED TRAJECTORY GUIDANCE UNIT SETUP WAS CORRECT, BUT THE TARGETING UNIT WAS UNABLE TO MOVE. NO ERROR CODE OR ANY WARNING SYMBOL ON THE AUTOMATED TRAJECTORY GUIDANCE UNIT. TECHNICAL SERVICES (TS) HAD A MANUFACTURER REPRESENTATIVE (REP) MOVE THE TARGETING UNIT TO A MORE FAVORABLE POSITION IN RELATION TO THEIR PLAN, THE UNIT MOVED TO PLANNED TRAJECTORY, AND THE SITE WAS ABLE TO MOVE FORWARD WITH SURGERY. THERE WAS NO DELAY AND NO IMPACT ON PATIENT OUTCOME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22339039 ↗
  • InjuryAug 22, 2022Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch

    MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION DEVICE BEING USED DURING A BRAIN PROCEDURE. IT WAS REPORTED THAT THE SITE'S ROBOTIC GUIDANCE WAS DISPLAYING TWO DIFFERENT ERROR CODES. FIRST, IT BOOTED INTO ERROR 0021 (CABLING FAULT ON THE COMMUNICATION LINE TO THE ROBOTIC GUIDANCE CONTROL UNIT). ALL CORDS WERE RE-SEATED AND SYSTEM REBOOTED, BUT THEN THE TARGETING UNIT ONLY BOOTED INTO A SCREEN WITH ERROR 0072 (ROBOTIC GUIDANCE TARGETING UNIT DOES NOT START PROPERLY: CONFIGURATION ERROR OR SELF TEST ERROR). MULTIPLE REBOOTS WERE ATTEMPTED WITH NO CHANGE. DELAY INFORMATION WAS UNKNOWN. THERE WAS NO REPORTED IMPACT TO THE PATIENT. IT WAS NOTED THAT NAVIGATION WAS ABORTED, AS WELL AS THAT THE PROCEDURE WAS RESCHEDULED. THIS IS CONFLICTING INFORMATION. ADDITIONAL INFORMATION WAS RECEIVED. IT WAS REPORTED THAT THE SUSPECTED CAUSE OF THE ISSUE WAS THE TARGETING UNIT RUNNING INTO THE ARM. THE PROCEDURE WAS RESCHEDULED. THE SITE HAD ALREADY MADE THE BURR HOLE WHEN THE TARGETING UNIT WENT DOWN. THE DELAY WAS ABOUT 20MINS OR SO PRIOR TO ABORTING THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15266222 ↗
  • MalfunctionApr 4, 2022Environmental Compatibility Problem; Environmental Compatibility Problem

    MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION DEVICE BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT THERE WAS A DAMAGED CONTROL UNIT CABLE. THE SYSTEM WOULD POWER OFF WHEN IT WAS ANGLED A CERTAIN WAY. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13998885 ↗
  • MalfunctionMar 23, 2021Device Sensing Problem; Device Sensing Problem

    ADDITIONAL INFORMATION WAS RECEIVED STATING THAT THE ISSUE WAS RESOLVE BY REPLACING THE SSD AND THE CONTROL UNIT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11552980 ↗
  • InjuryAug 21, 2020Mechanical Problem; Mechanical Problem

    MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM AND AUTOMATED TRAJECTORY GUIDANCE UNIT BEING USED FOR A THERMAL THERAPY LASER PLACEMENT. IT WAS REPORTED THAT INTRA-OPERATIVELY, THE SURGEON USED THE AUTOMATED TRAJECTORY GUIDANCE UNIT AND THERMAL THERAPY SYSTEM DRILL GUIDE SET TO DRILL A HOLE AND THE HOLE ENDED UP LARGER THAN EXPECTED AND THE BONE ANCHORS WOULD NOT FIT THE HOLE. THEY HAD TO DRILL ANOTHER HOLE USING A SHORTER DRILL GUIDE. THE PROCEDURE WAS COMPLETED WITH A THIRTY MINUTE DELAY. THERE WAS NO REPORTED IMPACT TO PATIENT OUTCOME. THE SURGEON SUSPECTED THAT IT OVERBORED THE HOLE DUE TO SOFT BONE, THE DRILL HAD TO BE PULSED IN AND OUT TO MAKE SURE THEY DIDN'T GO TOO DEEP, AND THE SURGEON ALLEGED THAT THE LONGER THERMAL THERAPY DRILL GUIDE WAS LESS STABLE THAN THE SHORT DRILL GUIDE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10437081 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K191597. One submission often covers many models, so check the product and lot details in each record.

  • Z-0417-2022 · Nov 9, 2021 · Open, Classified
    Stealth Autoguide Tracker, Model: 28248, which is an instrument in the Stealth Autoguide Basic Instrument Kit, Mode: 9736188, and the Bundle Autoguide Reg Kit.

Other Neurological stereotaxic Instrument devices

All Neurological stereotaxic Instrument devices →