HyperForm 104-4415Recalls & reported problems
Micro Therapeutics, Inc. · BALLOON SYSTEM 104-4415 V20 HYPERFORM · Class 2
All models of HyperForm (7)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00763000294205), received between Feb 22, 2023 and Jul 14, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Inflation Problem” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Inflation Problem2
- Burst Container or Vessel1
- Physical Resistance/Sticking1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 14, 2025Burst Container or Vessel; Inflation Problem; Burst Container or Vessel; Inflation Problem
ADDITIONAL INFORMATION RECEIVED REPORTED THE BALLOON RUPTURED. DURING PROCEDURE, WAS THERE NOT EXTENSIVE GUIDEWIRE MANIPULATION. ITS OWN GUIDEWIRE X-PEDION WAS USED WITH THE BALLOON. DURING THE INFLATION, THE GUIDEWIRE TIP ADVANCED 10CM OUT OF THE CATHETER TIP. THE GUIDEWIRE TIP WAS NOT SHAPED. 50:50 CONTRAST RATIO WAS USED. THE INJECTION RATE WAS SLOW. WITHDREW THE GUIDEWIRE TO DEFLATE THE BALLOON. IT WAS NOT INFLATED MULTIPLE TIMES. BLOOD DID NOT ENTER THE CATHETER LUMEN. CONTRAST AGENT LEAKAGE WAS FOUND. THE CATHETER WAS NOT KINKED. THE BALLOON WAS TESTED IN VITRO. THE BALLOON PERFORMED AS INTENDED. A CARTRIDGE SYRINGE WAS USED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22493401 ↗MalfunctionFeb 22, 2023Inflation Problem; Physical Resistance/Sticking; Inflation Problem; Physical Resistance/Sticking
MEDTRONIC RECEIVED A REPORT THAT THE PATIENT WAS A PATIENT WITH RECURRENT NASOPHARYNGEAL CARCINOMA. A BALLOON OCCLUSION EXPERIMENT WAS PERFORMED IN THE LEFT INTERNAL CAROTID ARTERY, AND 6F GUIDING WAS USED AS THE PROXIMAL SUPPORT CATHETER. AFTER THE HYPERFORM BALLOON WAS HYDRATED IN VITRO, IT WAS DEFLATED AND PRE-FILLED. WHEN THE X-PEDION-10 GUIDE WIRE CAME OUT OF THE DISTAL END OF THE BALLOON, THERE WAS A STUCK. THE GUIDE WIRE CAME OUT ABOUT 2 CM FROM THE TIP OF THE BALLOON CATHETER, AND THE SURGEON INFLATED THE BALLOON WITH A CARTRIDGE SYRINGE, BUT FOUND THE TIP OF BALLOON LEAKED WATER, AND THE BALLOON COULDN'T BE INFLATED. THE SURGEON SAID THAT THE BALLOON HAD QUALITY PROBLEMS AND DID NOT CHARGE FOR IT, AND REQUESTED A REPORT FOR DAMAGE TESTING. NO PATIENT SYMPTOMS OR FURTHER COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16416571 ↗
FDA clearance and classification
Cleared through 510(k) K101570 on Jul 13, 2010 (decision: Substantially Equivalent). Submitted by Micro Therapeutics Dba Ev3 Neurovascular as “Hyperglide And Hyperform Occlusion Balloon Systems”. The FDA review took 39 days. FDA record ↗
Cleared through 510(k) K091458 on Oct 15, 2009 (decision: Substantially Equivalent). Submitted by Ev3 Neurovascular as “Hyperglide Occlusion Balloon System, Hyperform Occlusion Balloon System, Model 104-4112, 104-4113, 104-4127,104-4132,104”. The FDA review took 150 days. FDA record ↗
The FDA regulates product code MJN (catheter, intravascular occluding, temporary) under 21 CFR 870.4450 as a Class II device. 21 CFR 870.4450 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the HyperForm 104-4415
Has the HyperForm 104-4415 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00763000294205), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the HyperForm 104-4415?
The FDA's MAUDE database lists 2 adverse event reports for this device, received between Feb 22, 2023 and Jul 14, 2025. In those reports, the most common specific device problems are “Inflation Problem” (2), “Burst Container or Vessel” (1) and “Physical Resistance/Sticking” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the HyperForm 104-4415?
None of the reports are classified as deaths, 0 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the HyperForm 104-4415 FDA approved?
It was cleared through 510(k) K091458 on Oct 15, 2009; its record lists 2 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the HyperForm 104-4415, and is it still sold?
The FDA's device database lists Micro Therapeutics, Inc. as the company responsible for it (brand HyperForm, model 104-4415). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Catheter, intravascular occluding, temporary devices
- Pruitt F3 Carotid Shunt 2012-10Lemaitre Vascular, Inc. · Catheter, Intravascular Occluding, Temporary1 death158 reports
- GORE Molding Occlusion Balloon Catheter MOB37W. L. Gore & Associates, Inc. · Catheter, Intravascular Occluding, Temporary6 deaths152 reports
- Bridge Occlusion Balloon 590-001Spectranetics Corporation · Catheter, Intravascular Occluding, Temporary4 deaths57 reports
- Pruitt F3 Carotid Shunt e2012-10Lemaitre Vascular, Inc. · Catheter, Intravascular Occluding, Temporary53 reports
- Pruitt F3 Carotid Shunt 2013-10Lemaitre Vascular, Inc. · Catheter, Intravascular Occluding, Temporary46 reports
- Equalizer™ M001171100Boston Scientific Corporation · Catheter, intravascular occluding, temporary18 reports
- Pruitt F3 Carotid Shunt e2013-10Lemaitre Vascular, Inc. · Catheter, Intravascular Occluding, Temporary17 reports
- TRANSFORM M003SSC07070Stryker Corporation · Catheter, intravascular occluding, temporary17 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page