HyperForm 104-4470 (UDI 00763000294267)Recalls & reported problems
Micro Therapeutics, Inc. · BALLOON SYSTEM 104-4470 V20 HYPERFORM · Class 2
All models of HyperForm (7)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00763000294267), received between Jul 31, 2023 and Nov 19, 2025 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Inflation Problem” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Inflation Problem3
- Leak/Splash3
- Burst Container or Vessel2
- Physical Resistance/Sticking1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 19, 2025Burst Container or Vessel; Burst Container or Vessel
ADDITIONAL INFORMATION WAS RECEIVED STATING THAT THE VESSEL TORTUOSITY WAS NORMAL. THE DEVICE WAS PREPARED AS INDICATED IN THE IFU (INSPECTION ,HYDRATION, ETC.). THE PROCEDURE WAS COMPLETED USING ANOTHER PRODUCT. LEAKAGE OCCURRED DURING SET UP. THE ISSUE WAS DISCOVERED DURING IN VITRO TESTING. THE CAUSE OF THE LEAK WAS NOT DETERMINED. THE BALLOON PERFORMED AS INTENDED DURING TESTING. THE GUIDEWIRE TIP WAS ADVANCED 7CM OUT OF THE CATHETER TIP DURING THE INFLATION. BALLOON LEAK WAS AT THE TIP. THE PHYSICIAN DID NOT SHAPE THE GUIDEWIRE TIP. THE BALLOON WAS INFLATED AND DEFLATED 1 TIME. INJECTION RATE WAS SLOW. A 1ML SYRINGE WAS USED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23593894 ↗MalfunctionNov 11, 2025Inflation Problem; Inflation Problem
MEDTRONIC RECEIVED A REPORT THAT THE AXIUM COIL WAS STRETCHED. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF A SACCULAR, UNRUPTURED ANEURYSM IN THE INTERNAL CAROTID ARTERY WITH A MAX DIAMETER OF 9 MM AND A 5 MM NECK DIAMETER. THE ACCESS VESSEL WAS THE FEMORAL ARTERY. IT WAS NOTED THE PATIENT'S BLOOD FLOW WAS NORMAL AND VESSEL TORTUOSITY WAS NORMAL. IT WAS REPORTED THAT THE AXIUM COIL WAS STRETCHED AND REPLACED AND THEN THE NEW COIL WAS IMPLANTED SMOOTHLY. AFTER THE ECHELON MICROCATHETER WAS SHAPED, IT WAS FOUND THAT THE TIP OF THE MICROCATHETER WAS DAMAGED. AFTER REPLACEMENT, THE OPERATION WAS SUCCESSFULLY COMPLETED. THE GUIDEWIRE COULD NOT PASS THROUGH THE ECHELON MICROCATHETER, AND THE OPERATION WAS SUCCESSFUL AFTER IT WAS REPLACED; THE GUIDEWIRE WAS OF ANOTHER BRAND. DURING HYPERFORM BALLOON PREPARATION, AIR LEAKAGE WAS DETECTED. THE REPORTED DEVICE AND ANY ACCESSORY DEVICES WERE PREPARED AS INDICATED IN THE INSTRUCTIONS FOR USE (IFU). THE CATHETER WAS FLUSHED PER THE IFU. NO PATIENT SYMPTOMS OR COMPLICATIONS WERE ASSOCIATED WITH THIS EVENT. ANCILLARY DEVICES INCLUDE 2 ECHELON CATHETERS, A HYPER FORM BALLOON AND AN ADDITIONAL AXIUM COIL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23525971 ↗MalfunctionJul 21, 2025Inflation Problem; Leak/Splash; Inflation Problem; Leak/Splash
MEDTRONIC RECEIVED A REPORT THAT THE BALLOON WAS PRE-FILLED IN VITRO, LEAKED AIR, AND THE OPERATION WAS SUCCESSFULLY COMPLETED AFTER REPLACEMENT. IT WAS UNKNOWN WHETHER THE REPORTED DEVICE AND ANY ACCESSORY DEVICES WERE PREPARED AS INDICATED IN THE INSTRUCTIONS FOR USE (IFU). NO PATIENT SYMPTOMS OR COMPLICATIONS WERE ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22555804 ↗MalfunctionSep 13, 2024Burst Container or Vessel; Burst Container or Vessel
ADDITIONAL INFORMATION WAS RECEIVED THAT THE PATIENT WAS UNDERGOING TREATMENT FOR A RUPTURED ANEURYSM OF THE LEFT ANTERIOR COMMUNICATING ARTERY, 3.8-2.6MM. VESSEL TORTUOSITY WAS MODERATE. THE DEVICE WAS PREPARED AS INDICATED IN THE IFU. THE CAUSE OF THE BALLOON BURST WAS NOT DETERMINED. THE PHYSICIAN DID NOT TEST THE BALLOON PRIOR TO USE. BALLOON LEAK OCCURRED IN THE MIDDLE SECTION. THE PHYSICIAN DID NOT SHAPE THE GUIDEWIRE TIP. THE BALLOON WAS INFLATED AND DEFLATED 2 TIMES. THE INJECTION RATE WAS STEADY. A CARTRIDGE SYRINGE WAS USED DURING INFLATION/DEFLATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20211947 ↗MalfunctionJun 19, 2024Inflation Problem; Leak/Splash; Inflation Problem; Leak/Splash
MEDTRONIC RECEIVED A REPORT THAT WHEN USING THE HYPERFORM BALLOON, AIR LEAKAGE WAS FOUND WHEN VENTING OUTSIDE THE BODY. NO PATIENT SYMPTOMS OR FURTHER COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT. ADDITIONAL INFORMATION RECEIVED THAT WHEN THE BALLOON WAS DEFLATED OUTSIDE THE BODY, CONTRAST AGENT WAS FOUND TO HAVE LEAKED OUT. THERE WAS NO GUIDEWIRE OPERATIONS. AN XPEDION GUIDEWIRE WAS USED WITH THE BALLOON. DURING THE INFLATION, THE GUIDEWIRE TIP WAS ADVANCED OUT OF THE CATHETER TIP ABOUT 1 CM. THE PHYSICIAN DID SHAPE THE GUIDEWIRE TIP. CONTRAST AGENT AND SALINE 1:1 RATIO. THE INJECTION RATE WAS STEADY. THE METHOD USED TO DEFLATE THE BALLOON WAS WITHDRAWAL WITH A SYRINGE. THERE WERE NOT MULTIPLE INFLATIONS, AND CONTRAST AGENT LEAKAGE WAS FOUND DURING THE FIRST INFLATION. BLOOD DID NOT ENTER THE CATHETER LUMEN AS THE BALLOON DID NOT ENTER THE PATIENT. CONTRAST WAS OBSERVED LEAKING DURING THE INFLATION ATTEMPT. THERE WERE NO KINKS ON THE CATHETER DURING USE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19570525 ↗MalfunctionJul 31, 2023Leak/Splash; Physical Resistance/Sticking; Leak/Splash; Physical Resistance/Sticking
ADDITIONAL INFORMATION RECEIVED REPORTED LESION CHARACTERISTICS: GIANT MALFORMATION AVM FAR FROM M2. VESSEL TORTUOSITY WAS MODERATE. THE DEVICE WAS PREPARED AS INDICATED IN THE IFU. THE PROCEDURE COMPLETED WITH A NEW BALLOON. THE GUIDEWIRE TIP WAS ADVANCED OUT OF THE CATHETER TIP DURING THE INFLATION. THE PHYSICIAN SHAPED THE GUIDEWIRE TIP. THERE WAS NO KINK OR DAMAGE ON THE GUIDEWIRE. THE GUIDEWIRE WAS ABLE TO PASS THE CATHETER HUB.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17424135 ↗
FDA clearance and classification
Cleared through 510(k) K101570 on Jul 13, 2010 (decision: Substantially Equivalent). Submitted by Micro Therapeutics Dba Ev3 Neurovascular as “Hyperglide And Hyperform Occlusion Balloon Systems”. The FDA review took 39 days. FDA record ↗
Cleared through 510(k) K091458 on Oct 15, 2009 (decision: Substantially Equivalent). Submitted by Ev3 Neurovascular as “Hyperglide Occlusion Balloon System, Hyperform Occlusion Balloon System, Model 104-4112, 104-4113, 104-4127,104-4132,104”. The FDA review took 150 days. FDA record ↗
Cleared through Special 510(k) K011656 on Jun 20, 2001 (decision: Substantially Equivalent). Submitted by Micro Therapeutics, Inc. as “Hyperform Occlusion Balloon System, Model 104-4470, 104-4471, 104-4770, 104-4771”. The FDA review took 22 days. FDA record ↗
The FDA regulates product code MJN (catheter, intravascular occluding, temporary) under 21 CFR 870.4450 as a Class II device. 21 CFR 870.4450 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the HyperForm 104-4470
Has the HyperForm 104-4470 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00763000294267), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the HyperForm 104-4470?
The FDA's MAUDE database lists 6 adverse event reports for this device, received between Jul 31, 2023 and Nov 19, 2025. In those reports, the most common specific device problems are “Inflation Problem” (3), “Leak/Splash” (3) and “Burst Container or Vessel” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the HyperForm 104-4470?
None of the reports are classified as deaths, 0 as injuries and 6 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the HyperForm 104-4470 FDA approved?
It was cleared through Special 510(k) K011656 on Jun 20, 2001; its record lists 3 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the HyperForm 104-4470, and is it still sold?
The FDA's device database lists Micro Therapeutics, Inc. as the company responsible for it (brand HyperForm, model 104-4470). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Catheter, intravascular occluding, temporary devices
- Pruitt F3 Carotid Shunt 2012-10Lemaitre Vascular, Inc. · Catheter, Intravascular Occluding, Temporary1 death158 reports
- GORE Molding Occlusion Balloon Catheter MOB37W. L. Gore & Associates, Inc. · Catheter, Intravascular Occluding, Temporary6 deaths152 reports
- Bridge Occlusion Balloon 590-001Spectranetics Corporation · Catheter, Intravascular Occluding, Temporary4 deaths57 reports
- Pruitt F3 Carotid Shunt e2012-10Lemaitre Vascular, Inc. · Catheter, Intravascular Occluding, Temporary53 reports
- Pruitt F3 Carotid Shunt 2013-10Lemaitre Vascular, Inc. · Catheter, Intravascular Occluding, Temporary46 reports
- Equalizer™ M001171100Boston Scientific Corporation · Catheter, intravascular occluding, temporary18 reports
- Pruitt F3 Carotid Shunt e2013-10Lemaitre Vascular, Inc. · Catheter, Intravascular Occluding, Temporary17 reports
- TRANSFORM M003SSC07070Stryker Corporation · Catheter, intravascular occluding, temporary17 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page