HyperForm 104-4715Recalls & reported problems

Micro Therapeutics, Inc. · BALLOON SYSTEM 104-4715 V20 HYPERFORM · Class 2

Adverse event reports
8
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00763000294298), received between Aug 10, 2022 and Nov 17, 2025 — classified as 1 injury, 7 malfunctions.

The most frequently reported specific device problem is “Burst Container or Vessel” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Burst Container or Vessel4
  • Inflation Problem2
  • Leak/Splash2
  • Deflation Problem1
  • Difficult to Advance1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 17, 2025Deflation Problem; Deflation Problem

    ADDITIONAL INFORMATION WAS RECEIVED. VESSEL TORTUOSITY WAS ASSESSED AS MODERATE. THE DEVICE WAS PREPARED ACCORDING TO THE IFU, INCLUDING INSPECTION AND HYDRATION. THE CAUSE OF THE EVENT IS UNKNOWN. THE PHYSICIAN TESTED THE BALLOON PRIOR TO USE, AND IT PERFORMED AS INTENDED DURING TESTING. IODINATED ALCOHOL WAS USED AS THE CONTRAST AGENT. THE GUIDEWIRE TIP WAS ADVANCED MORE THAN 3 CM OUT OF THE CATHETER TIP DURING INFLATION, AND A 1 ML SYRINGE WAS USED FOR BALLOON INFLATION AND DEFLATION. IT IS UNKNOWN WHETHER BLOOD REFLUXED INTO THE BALLOON LUMEN. NO FRICTION OR DIFFICULTY WAS ENCOUNTERED DURING INFLATION. AFTER MULTIPLE INFLATIONS, THE CATHETER AND GUIDEWIRE WERE EXAMINED FOR CLOT AT THE TIP OR INFLATION HOLES, AND NO THROMBOSIS WAS FOUND. NO EXTENSIVE GUIDEWIRE MANIPULATION OCCURRED DURING THE PROCEDURE. NO KINK OR DAMAGE WAS OBSERVED ON EITHER THE CATHETER OR THE GUIDEWIRE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23569561 ↗
  • MalfunctionJul 9, 2025Burst Container or Vessel; Burst Container or Vessel

    MEDTRONIC RECEIVED A REPORT THAT THE PATIENT WAS PLANNED TO UNDERGO EMBOLIZATION OF A RIGHT LENTICULOSTRIATE ARTERY ANEURYSM USING THIS CONSUMABLE. OCCLUSION OF THE MIDDLE CEREBRAL ARTERY WAS INTENDED TO FACILITATE THE EMBOLIZATION. DURING BALLOON INFLATION, THE BALLOON BURST. HEALTHCARE PROVIDER (HCP) REPLACED IT WITH ANOTHER BALLOON FOR USE TO COMPLETE THE PROCEDURE. THIS EVENT SERIOUSLY AFFECTED THE PROGRESS OF SURGERY. NO PATIENT SYMPTOMS OR FURTHER COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT. THE PATIENT WAS UNDERGOING SURGERY FOR VESSEL OCCLUSION TREATMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22459891 ↗
  • MalfunctionMay 29, 2025Burst Container or Vessel; Burst Container or Vessel

    ADDITIONAL INFORMATION WAS RECEIVED REPORTING THE FIRST BALLOON BURST, SO WE REPLACED IT WITH THE SECOND ONE. THE SECOND ONE BURST AGAIN, SO WE REPLACED IT WITH THE THIRD ONE. THEN THE OPERATION WAS SUCCESSFULLY COMPLETED. THE CAUSE OF THE BURSTS WAS NOT DETERMINED. BECAUSE THE BALLOON HAD BEEN DISCARDED IT WAS UNCLEAR WHERE WAS THE BALLOONS' LEAK WAS. NO CARTRIDGE SYRINGES WERE USED, AND IT WAS UNCLEAR WHAT SYRINGES WERE USED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22107003 ↗
  • MalfunctionMar 10, 2025Burst Container or Vessel; Inflation Problem; Burst Container or Vessel; Inflation Problem

    ADDITIONAL INFORMATION WAS RECEIVED STATING THAT THE CAUSE OF THE BALLOON BEING BROKEN WAS NOT DETERMINED. THE PHYSICIAN TESTED THE BALLOON PRIOR TO USE. THE BALLOON DID NOT PERFORM AS INTENDED DURING TESTING. THE PHYSICIAN SHAPED THE GUIDEWIRE TIP. THE INJECTION RATE WAS STEADY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21562386 ↗
  • MalfunctionJun 26, 2024Leak/Splash; Leak/Splash

    MEDTRONIC RECEIVED A REPORT THAT WHEN THE SURGEON WAS TESTING THE BALLOON DURING THE OPERATION, THEY FOUND THAT THE BALLOON WAS LEAKING AND COULD NOT WORK PROPERLY, SO THEY REPLACED IT WITH ANOTHER PRODUCT FOR THE OPERATION. NO PATIENT SYMPTOMS OR COMPLICATIONS WERE ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19613477 ↗
  • MalfunctionOct 30, 2023Inflation Problem; Leak/Splash; Inflation Problem; Leak/Splash

    ON (B)(6) 2023 MPXR 1111735 (REP, HCP, FOR, DIS): MEDTRONIC RECEIVED A REPORT THAT AFTER OPENING THE PACKAGE AND PERFORMING HYDRATION, IT WAS DISCOVERED THAT THE BALLOON WAS LEAKING AND COULD NOT BE INFLATED. REPLACED THE PRODUCT TO COMPLETE THE SURGERY. NO PATIENT SYMPTOMS OR FURTHER COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18029150 ↗
  • InjuryJan 19, 2023Difficult to Advance; Difficult to Advance

    MEDTRONIC RECEIVED A REPORT THAT DURING BALLOON-ASSISTED EMBOLIZATION OF ANEURYSM, THE OCCLUSION BALLOON CATHETER SYSTEM COULD NOT PASS THROUGH THE PARENT ARTERY. THE ADAPTIVE GUIDE WIRE PASSED THROUGH THE BEND OF CAVERNOUS SINUS BLOOD VESSEL, AND THE GUIDE WIRE FORMED A LOOP IN THE VASCULAR LUMEN. MANY ATTEMPTS FAILED, WHICH LED TO THE FAILURE OF BALLOON-ASSISTED EMBOLIZATION. NO PATIENT SYMPTOMS OR FURTHER COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16204135 ↗
  • MalfunctionAug 10, 2022Burst Container or Vessel; Burst Container or Vessel

    MEDTRONIC RECEIVED INFORMATION REGARDING A RUPTURED HYPERFORM BALLOON. IT WAS REPORTED THAT THE PATIENT WAS BEING TREATED FOR A LEFT INTERNAL CAROTID ARTERY (ICA) TUMOR AND THE HYPERFORM BALLOON WAS BEING USED IN A BALLOON TEST OCCLUSION. THE HYPERFORM BALLOON CATHETER WAS PREPARED PER THE INSTRUCTIONS FOR USE (IFU). THE HYPERFORM WAS THEN INSERTED THROUGH THE SHEATH AND GUIDE CATHETER AND INFLATED USING A MEDTRONIC CADENCE SYRINGE TO 0.20CC (5.7MM BALLOON SIZE). THE BALLOON WAS NOTED TO BE LOSING PRESSURE SO THE DOCTOR REMOVED IT AND NOTED CONTRAST EXITING THE DISTAL TIP. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15204167 ↗