HyperForm 104-4715Recalls & reported problems
Micro Therapeutics, Inc. · BALLOON SYSTEM 104-4715 V20 HYPERFORM · Class 2
All models of HyperForm (7)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00763000294298), received between Aug 10, 2022 and Nov 17, 2025 — classified as 1 injury, 7 malfunctions.
The most frequently reported specific device problem is “Burst Container or Vessel” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Burst Container or Vessel4
- Inflation Problem2
- Leak/Splash2
- Deflation Problem1
- Difficult to Advance1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 17, 2025Deflation Problem; Deflation Problem
ADDITIONAL INFORMATION WAS RECEIVED. VESSEL TORTUOSITY WAS ASSESSED AS MODERATE. THE DEVICE WAS PREPARED ACCORDING TO THE IFU, INCLUDING INSPECTION AND HYDRATION. THE CAUSE OF THE EVENT IS UNKNOWN. THE PHYSICIAN TESTED THE BALLOON PRIOR TO USE, AND IT PERFORMED AS INTENDED DURING TESTING. IODINATED ALCOHOL WAS USED AS THE CONTRAST AGENT. THE GUIDEWIRE TIP WAS ADVANCED MORE THAN 3 CM OUT OF THE CATHETER TIP DURING INFLATION, AND A 1 ML SYRINGE WAS USED FOR BALLOON INFLATION AND DEFLATION. IT IS UNKNOWN WHETHER BLOOD REFLUXED INTO THE BALLOON LUMEN. NO FRICTION OR DIFFICULTY WAS ENCOUNTERED DURING INFLATION. AFTER MULTIPLE INFLATIONS, THE CATHETER AND GUIDEWIRE WERE EXAMINED FOR CLOT AT THE TIP OR INFLATION HOLES, AND NO THROMBOSIS WAS FOUND. NO EXTENSIVE GUIDEWIRE MANIPULATION OCCURRED DURING THE PROCEDURE. NO KINK OR DAMAGE WAS OBSERVED ON EITHER THE CATHETER OR THE GUIDEWIRE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23569561 ↗MalfunctionJul 9, 2025Burst Container or Vessel; Burst Container or Vessel
MEDTRONIC RECEIVED A REPORT THAT THE PATIENT WAS PLANNED TO UNDERGO EMBOLIZATION OF A RIGHT LENTICULOSTRIATE ARTERY ANEURYSM USING THIS CONSUMABLE. OCCLUSION OF THE MIDDLE CEREBRAL ARTERY WAS INTENDED TO FACILITATE THE EMBOLIZATION. DURING BALLOON INFLATION, THE BALLOON BURST. HEALTHCARE PROVIDER (HCP) REPLACED IT WITH ANOTHER BALLOON FOR USE TO COMPLETE THE PROCEDURE. THIS EVENT SERIOUSLY AFFECTED THE PROGRESS OF SURGERY. NO PATIENT SYMPTOMS OR FURTHER COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT. THE PATIENT WAS UNDERGOING SURGERY FOR VESSEL OCCLUSION TREATMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22459891 ↗MalfunctionMay 29, 2025Burst Container or Vessel; Burst Container or Vessel
ADDITIONAL INFORMATION WAS RECEIVED REPORTING THE FIRST BALLOON BURST, SO WE REPLACED IT WITH THE SECOND ONE. THE SECOND ONE BURST AGAIN, SO WE REPLACED IT WITH THE THIRD ONE. THEN THE OPERATION WAS SUCCESSFULLY COMPLETED. THE CAUSE OF THE BURSTS WAS NOT DETERMINED. BECAUSE THE BALLOON HAD BEEN DISCARDED IT WAS UNCLEAR WHERE WAS THE BALLOONS' LEAK WAS. NO CARTRIDGE SYRINGES WERE USED, AND IT WAS UNCLEAR WHAT SYRINGES WERE USED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22107003 ↗MalfunctionMar 10, 2025Burst Container or Vessel; Inflation Problem; Burst Container or Vessel; Inflation Problem
ADDITIONAL INFORMATION WAS RECEIVED STATING THAT THE CAUSE OF THE BALLOON BEING BROKEN WAS NOT DETERMINED. THE PHYSICIAN TESTED THE BALLOON PRIOR TO USE. THE BALLOON DID NOT PERFORM AS INTENDED DURING TESTING. THE PHYSICIAN SHAPED THE GUIDEWIRE TIP. THE INJECTION RATE WAS STEADY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21562386 ↗MalfunctionJun 26, 2024Leak/Splash; Leak/Splash
MEDTRONIC RECEIVED A REPORT THAT WHEN THE SURGEON WAS TESTING THE BALLOON DURING THE OPERATION, THEY FOUND THAT THE BALLOON WAS LEAKING AND COULD NOT WORK PROPERLY, SO THEY REPLACED IT WITH ANOTHER PRODUCT FOR THE OPERATION. NO PATIENT SYMPTOMS OR COMPLICATIONS WERE ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19613477 ↗MalfunctionOct 30, 2023Inflation Problem; Leak/Splash; Inflation Problem; Leak/Splash
ON (B)(6) 2023 MPXR 1111735 (REP, HCP, FOR, DIS): MEDTRONIC RECEIVED A REPORT THAT AFTER OPENING THE PACKAGE AND PERFORMING HYDRATION, IT WAS DISCOVERED THAT THE BALLOON WAS LEAKING AND COULD NOT BE INFLATED. REPLACED THE PRODUCT TO COMPLETE THE SURGERY. NO PATIENT SYMPTOMS OR FURTHER COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18029150 ↗InjuryJan 19, 2023Difficult to Advance; Difficult to Advance
MEDTRONIC RECEIVED A REPORT THAT DURING BALLOON-ASSISTED EMBOLIZATION OF ANEURYSM, THE OCCLUSION BALLOON CATHETER SYSTEM COULD NOT PASS THROUGH THE PARENT ARTERY. THE ADAPTIVE GUIDE WIRE PASSED THROUGH THE BEND OF CAVERNOUS SINUS BLOOD VESSEL, AND THE GUIDE WIRE FORMED A LOOP IN THE VASCULAR LUMEN. MANY ATTEMPTS FAILED, WHICH LED TO THE FAILURE OF BALLOON-ASSISTED EMBOLIZATION. NO PATIENT SYMPTOMS OR FURTHER COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16204135 ↗MalfunctionAug 10, 2022Burst Container or Vessel; Burst Container or Vessel
MEDTRONIC RECEIVED INFORMATION REGARDING A RUPTURED HYPERFORM BALLOON. IT WAS REPORTED THAT THE PATIENT WAS BEING TREATED FOR A LEFT INTERNAL CAROTID ARTERY (ICA) TUMOR AND THE HYPERFORM BALLOON WAS BEING USED IN A BALLOON TEST OCCLUSION. THE HYPERFORM BALLOON CATHETER WAS PREPARED PER THE INSTRUCTIONS FOR USE (IFU). THE HYPERFORM WAS THEN INSERTED THROUGH THE SHEATH AND GUIDE CATHETER AND INFLATED USING A MEDTRONIC CADENCE SYRINGE TO 0.20CC (5.7MM BALLOON SIZE). THE BALLOON WAS NOTED TO BE LOSING PRESSURE SO THE DOCTOR REMOVED IT AND NOTED CONTRAST EXITING THE DISTAL TIP. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15204167 ↗
FDA clearance and classification
Cleared through 510(k) K101570 on Jul 13, 2010 (decision: Substantially Equivalent). Submitted by Micro Therapeutics Dba Ev3 Neurovascular as “Hyperglide And Hyperform Occlusion Balloon Systems”. The FDA review took 39 days. FDA record ↗
Cleared through 510(k) K091458 on Oct 15, 2009 (decision: Substantially Equivalent). Submitted by Ev3 Neurovascular as “Hyperglide Occlusion Balloon System, Hyperform Occlusion Balloon System, Model 104-4112, 104-4113, 104-4127,104-4132,104”. The FDA review took 150 days. FDA record ↗
The FDA regulates product code MJN (catheter, intravascular occluding, temporary) under 21 CFR 870.4450 as a Class II device. 21 CFR 870.4450 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the HyperForm 104-4715
Has the HyperForm 104-4715 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00763000294298), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the HyperForm 104-4715?
The FDA's MAUDE database lists 8 adverse event reports for this device, received between Aug 10, 2022 and Nov 17, 2025. In those reports, the most common specific device problems are “Burst Container or Vessel” (4), “Inflation Problem” (2) and “Leak/Splash” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the HyperForm 104-4715?
None of the reports are classified as deaths, 1 as an injury and 7 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the HyperForm 104-4715 FDA approved?
It was cleared through 510(k) K091458 on Oct 15, 2009; its record lists 2 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the HyperForm 104-4715, and is it still sold?
The FDA's device database lists Micro Therapeutics, Inc. as the company responsible for it (brand HyperForm, model 104-4715). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Catheter, intravascular occluding, temporary devices
- Pruitt F3 Carotid Shunt 2012-10Lemaitre Vascular, Inc. · Catheter, Intravascular Occluding, Temporary1 death158 reports
- GORE Molding Occlusion Balloon Catheter MOB37W. L. Gore & Associates, Inc. · Catheter, Intravascular Occluding, Temporary6 deaths152 reports
- Bridge Occlusion Balloon 590-001Spectranetics Corporation · Catheter, Intravascular Occluding, Temporary4 deaths57 reports
- Pruitt F3 Carotid Shunt e2012-10Lemaitre Vascular, Inc. · Catheter, Intravascular Occluding, Temporary53 reports
- Pruitt F3 Carotid Shunt 2013-10Lemaitre Vascular, Inc. · Catheter, Intravascular Occluding, Temporary46 reports
- Equalizer™ M001171100Boston Scientific Corporation · Catheter, intravascular occluding, temporary18 reports
- Pruitt F3 Carotid Shunt e2013-10Lemaitre Vascular, Inc. · Catheter, Intravascular Occluding, Temporary17 reports
- TRANSFORM M003SSC07070Stryker Corporation · Catheter, intravascular occluding, temporary17 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page