Skip to content

HyperForm 104-4770 (UDI 00763000294328)Recalls & reported problems

Micro Therapeutics, Inc. · BALLOON SYSTEM 104-4770 V20 HYPERFORM · Class 2

All models of HyperForm (7)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00763000294328), received between Jan 6, 2023 and Jun 17, 2024 — classified as 1 injury, 10 malfunctions.

The most frequently reported specific device problem is “Leak/Splash” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 2 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Leak/Splash6
  • Burst Container or Vessel4
  • Inflation Problem4
  • Material Deformation1
  • Physical Resistance/Sticking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions11
  • Vascular Dissection1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 17, 2024Leak/Splash; Leak/Splash

    MEDTRONIC RECEIVED A REPORT THAT THERE WAS LIQUID LEAKAGE WITH A BALLOON, TWO GUIDEWIRES COULD NOT EXIT THE DISTAL END OF THE BALLOON, ONE COIL WAS STRETCHED AND REPLACED WITH A PRODUCT OF THE SAME MODEL AND IMPLANTED SUCCESSFULLY, AND ANOTHER COIL COULD NOT BE DETACHED AND WAS SUCCESSFULLY IMPLANTED AFTER REPLACEMENT. THE REPORTED DEVICES AND ANY ACCESSORY DEVICES WERE PREPARED AS INDICATED IN THE INSTRUCTIONS FOR USE (IFU). NO PATIENT SYMPTOMS OR COMPLICATIONS WERE ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19546963 ↗
  • MalfunctionJun 13, 2024Leak/Splash; Leak/Splash

    MEDTRONIC RECEIVED A REPORT THAT ASSISTED EMBOLIZATION ANEURYSM SURGERY, CONTRAST AGENT LEAKED OUT DURING INFLATION. THE BALLOON WAS REPLACED AND THE SURGERY WENT SMOOTHLY. NO PATIENT SYMPTOMS OR FURTHER COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19528682 ↗
  • MalfunctionApr 2, 2024Leak/Splash; Leak/Splash

    MEDTRONIC RECEIVED A REPORT THAT ANEURYSM EMBOLIZATION WAS PERFORMED. AFTER THE BALLOON WAS IN PLACE, IT WAS INFLATED. LEAKAGE OF THE BALLOON WAS FOUND, ANOTHER BALLOON WAS REPLACED, FILLED SMOOTHLY, AND THE OPERATION WAS COMPLETED NO PATIENT SYMPTOMS OR FURTHER COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT. ADDITIONAL INFORMATION RECEIVED THAT THE BALLOON DID NOT INFLATE. THE LEAK WAS OBSERVED INSIDE THE BODY. IT WAS CLARIFIED THAT LIQUID WAS LEAKING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19020224 ↗
  • MalfunctionMar 22, 2024Burst Container or Vessel; Material Deformation; Burst Container or Vessel; Material Deformation

    MEDTRONIC RECEIVED A REPORT THAT THE HYPERFORM BALLOON WAS FOUND TO BE DAMAGED DURING EXTRACORPOREAL INFLATING. NO PATIENT SYMPTOMS OR FURTHER COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT THE BALLOON WAS PREPARED AS INDICATED IN THE INSTRUCTIONS FOR USE (IFU). THE PROCEDURE WAS COMPLETED BY REPLACING WITH ANOTHER BALLOON. THE PROCEDURE THAT WAS BEING PERFORMED WAS A BALLOON-ASSISTED COIL EMBOLIZATION. LEAKAGE WAS DISCOVERED WHEN THE BALLOON WAS INFLATED DURING THE SYSTEM PREPARATION IN VITRO. THE CAUSE OF THE DAMAGE WAS THE BALLOON WAS DAMAGED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18961229 ↗
  • MalfunctionJan 11, 2024Inflation Problem; Inflation Problem

    MEDTRONIC RECEIVED A REPORT THAT AFTER PREPARING THE HYPERFORM BALLOON, AND INSERTED IT INTO THE PATIENT, THE BALLOON DOES NOT INFLATE AS EXPECTED. NO PATIENT SYMPTOMS OR FURTHER COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT. ADDITIONAL INFORMATION RECEIVED REPORTED THERE WAS NOT EXTENSIVE GUIDEWIRE MANIPULATION. A SYNCHRO GUIDEWIRE WAS USED. THE GUIDEWIRE TIP WAS REASONABLY ADVANCED OUT OF THE CATHETER TIP DURING INFLATION AND DEFLATION. THE PHYSICIAN DID NOT SHAPE THE GUIDEWIRE TIP. 50:50 CONTRAST WAS USED. A STEADY INJECTION RATE WAS USED. A SYRINGE STEADY INJECTION WAS USED TO DEFLATE THE BALLOON. AFTER MULTIPLE ATTEMPTS THE CATHETER AND GUIDEWIRE WERE REMOVED AND INSPECTED. NO CLOT WAS FOUND. NO BLOOD ENTERED THE CATHETER LUMEN. NO CONTRAST WAS OBSERVED LEAKING DURING INFLATION ATTEMPT. THERE WERE NO KINKS ON THE CATHETER DURING USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18491685 ↗
  • MalfunctionSep 25, 2023Burst Container or Vessel; Burst Container or Vessel

    ADDITIONAL INFORMATION RECEIVED REPORTED THAT THE PHYSICIAN TESTED THE BALLOON PRIOR TO USE. THE BALLOON DID NOT PERFORM AS INTENDED DURING TESTING. THE GUIDEWIRE TIP ADVANCED ABOUT 4 CM OUT OF THE CATHETER TIP DURING THE INFLATION. DURING EXTERNAL INFLATING, WATER LEAKAGE WAS FOUND AT THE DISTAL MARK. THE PHYSICIAN SHAPED THE GUIDEWIRE TIP. THE BALLOON WAS INFLATED TWICE, BUT IT STILL LEAKED. THE INJECTION RATE WAS SLOW. AN ORDINARY SYRINGE (1 MM) WAS USED DURING INFLATION/DEFLATION OF THE BALLOON.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17807768 ↗
  • MalfunctionJun 23, 2023Inflation Problem; Leak/Splash; Inflation Problem; Leak/Splash

    ADDITIONAL INFORMATION RECEIVED REPORTED THE LESION WAS IN THE OPHTHALMIC ARTERY. VESSEL TORTUOSITY WAS NORMAL. THE DEVICES WERE PRE PARED AS INDICATED IN THE INSTRUCTIONS FOR USE (IFU).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17190905 ↗
  • MalfunctionJun 6, 2023Inflation Problem; Leak/Splash; Inflation Problem; Leak/Splash

    MEDTRONIC RECEIVED A REPORT THAT THE HYPERFORM BALLOON WAS OPENED LIKE NORMAL. THE DOCTOR PREPPED THE DEVICE AND ATTACHED THE RHV <(>&<)> 3-WAY COMBO. THE CATHETER WAS ALSO FLUSHED BEFORE INSERTING THE WIRE. AFTER INSERTING THE WIRE AND WHILE ATTEMPTING TO PURGE THE BALLOON OF AIR, THE DOCTOR NOTICED THAT EVEN AFTER FLUSHING ROUGHLY 8CC OF CONTRAST MIX THE BALLOON WOULD NOT INFLATE. WE DETERMINED THAT THE DEVICE WAS DEFECTIVE AND WE REMOVED THE DEVICE FROM THE TABLE. WE OPENED ANOTHER HYPERFORM BALLOON AND FLUSHED THE DEVICE IN THE SAME MANNER AND THE NEW DEVICE WORKED AS INTENDED. ADDITIONAL INFORMATION WAS RECEIVED THAT THE CATHETER WAS NEVER INSERTED INTO THE PATIENT AS THE BALLOON WAS NOT WORKING PROPERLY. THERE WAS NO EXTENSIVE GUIDEWIRE MANIPULATION, ONLY THE WIRE THAT COMES WITH THE BALLOON WAS PLACED IN THE DEVICE. THE WIRE WAS ESTIMATED TO BE ADVANCED APPROXIMATELY 2-3CM OUT THE TIP OF THE CATHETER. THE WIRE WAS NOT SHAPED YET AND THE CONTRAST RATIO WAS 50:50. THE INJECTION RATE WAS SLOW AND STEADY. THE BALLOON WAS NEVER INFLATED. BLOOD DID NOT ENTER THE CATHETER LUMEN. CONTRAST WAS OBSERVED LEAKING AND THE BALLOON WAS NEVER INFLATED. THERE WERE NO KINKS ON THE CATHETER DU

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17073665 ↗
  • InjuryMay 3, 2023Burst Container or Vessel; Burst Container or Vessel

    MEDTRONIC RECEIVED A REPORT THAT THE HYPERFORM BALLOON POPPED UPON INFLATION IN THE INTERNAL CAROTID ARTERY. NO PATIENT SYMPTOMS OR FURTHER COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Vascular Dissection

    MAUDE report 16859890 ↗
  • MalfunctionFeb 17, 2023Burst Container or Vessel; Physical Resistance/Sticking; Burst Container or Vessel; Physical Resistance/Sticking

    ADDITIONAL INFORMATION RECEIVED CLARIFIED THAT THE LEAKING WAS FOUND DURING THE PROCEDURE, BUT THEY WERE UNSURE WHERE THE LEAK POINT WAS. THE LEAK WAS VERIFIED OUTSIDE THE PATIENT. THERE WAS NO EXTENSION GUIDEWIRE MANIPULATION, AND A BALANCE MIDDLE WEIGHT GUIDEWIRE WAS USED. THE GUIDEWIRE WAS SAID TO BE FAR ENOUGH OUT FROM THE CATHETER TIP. THE GUIDEWIRE TIP WAS SHAPED. A 50:50 CONTRAST RATIO WAS USED. THE INJECTION RATE WAS SLOW AND STEADY. BLOOD DID NOT ENTER THE CATHETER LUMEN, AND THERE WAS NO CLOT AT THE TIP OF THE INFLATION HOLES. THE LEAK WAS SEEN ON THE BLANK ROADMAP, AND CONFIRMED ON THE REAL TIME FLUORO. THERE WAS NO KINK OR DAMAGE OBSERVED ON THE CATHETER OR GUIDEWIRE DURING USE. THE PHYSICIAN DID TEST THE BALLOON PRIOR TO USE, AND THERE WAS DIFFICULTY GETTING THE GUIDEWIRE OUT OF THE TIP. THERE WAS NO FRICTION OR DIFFICULTY DURING INFLATION. A 1 CC MEDALLION SYRINGE WAS USED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16401528 ↗
  • MalfunctionJan 6, 2023Inflation Problem; Leak/Splash; Inflation Problem; Leak/Splash

    MEDTRONIC RECEIVED REPORT THAT DURING THE PROCEDURE, WHEN THE DOCTOR PUT THE X-PEDION GUIDEWIRE THROUGH THE HYPERFORM BALLOON AND ATTEMPTED TO INFLATE, THE BALLOON DID NOT INFLATE AND DRIPS CAME OUT THE END OF THE BALLOON. THE DEVICE WAS REPLACED WITH ANOTHER HYPERFORM OF A DIFFERENT LOT TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16112878 ↗

FDA clearance and classification

Cleared through 510(k) K101570 on Jul 13, 2010 (decision: Substantially Equivalent). Submitted by Micro Therapeutics Dba Ev3 Neurovascular as “Hyperglide And Hyperform Occlusion Balloon Systems”. The FDA review took 39 days. FDA record ↗

Cleared through 510(k) K091458 on Oct 15, 2009 (decision: Substantially Equivalent). Submitted by Ev3 Neurovascular as “Hyperglide Occlusion Balloon System, Hyperform Occlusion Balloon System, Model 104-4112, 104-4113, 104-4127,104-4132,104”. The FDA review took 150 days. FDA record ↗

Cleared through Special 510(k) K011656 on Jun 20, 2001 (decision: Substantially Equivalent). Submitted by Micro Therapeutics, Inc. as “Hyperform Occlusion Balloon System, Model 104-4470, 104-4471, 104-4770, 104-4771”. The FDA review took 22 days. FDA record ↗

The FDA regulates product code MJN (catheter, intravascular occluding, temporary) under 21 CFR 870.4450 as a Class II device. 21 CFR 870.4450 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the HyperForm 104-4770

Has the HyperForm 104-4770 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00763000294328), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the HyperForm 104-4770?

The FDA's MAUDE database lists 11 adverse event reports for this device, received between Jan 6, 2023 and Jun 17, 2024. In those reports, the most common specific device problems are “Leak/Splash” (6), “Burst Container or Vessel” (4) and “Inflation Problem” (4); the most common patient problem is “Vascular Dissection” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the HyperForm 104-4770?

None of the reports are classified as deaths, 1 as an injury and 10 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the HyperForm 104-4770 FDA approved?

It was cleared through Special 510(k) K011656 on Jun 20, 2001; its record lists 3 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the HyperForm 104-4770, and is it still sold?

The FDA's device database lists Micro Therapeutics, Inc. as the company responsible for it (brand HyperForm, model 104-4770). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Catheter, intravascular occluding, temporary devices

All Catheter, intravascular occluding, temporary devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page