HyperGlide 104-4112Recalls & reported problems

Micro Therapeutics, Inc. · BALLOON SYSTEM 104-4112 V20 HYPERGLIDE · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00763000294359), received between Jul 10, 2023 and Jun 11, 2025 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Inflation Problem” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Inflation Problem3
  • Leak/Splash3
  • Burst Container or Vessel1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 11, 2025Burst Container or Vessel; Inflation Problem; Burst Container or Vessel; Inflation Problem

    MEDTRONIC RECEIVED A REPORT THAT THE PHYSICIAN TESTED THE BALLOON INFLATION BEFORE USING IT ON THE PATIENT AND FOUND THAT THE BALLOON WAS LEAKING AND COULD NOT BE FULLY INFLATED. NO PATIENT SYMPTOMS OR COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22183975 ↗
  • MalfunctionFeb 18, 2025Inflation Problem; Leak/Splash; Inflation Problem; Leak/Splash

    ADDITIONAL INFORMATION RECEIVED REPORTED THE BALLOON LEAK WAS OBSERVED DURING INTRAOPERATIVE USE, EXTRACORPOREAL INFLATION. THE PATIENT WAS BEING TREATED FOR AN IRREGUALR ANEURSYM. VESSEL TORTUOSITY WAS MILD. THE DEVICES WERE PREPARED AS INDICATED IN THE INSTRUCTIOSN FOR USE (IFU. THE BALLOON WAS REPLACED TO COMPLETE THE PROCEDURE. THERE WAS NOT EXTENISVE GUIDEWIRE MANIPULATION. DURING INFLATION THE GUIDEWIRE TIP WAS NOT ADVANCED OUT OF THE CATHETER TIP. THE INJECTION RATE WAS SLOW. AFTER MULTIPLE INFLATIONS THE CATHETER AND GUIDEWIRE WAS REMOVED AND INSPECTED. THE HEALTHCARE PROVIDER (HCP) DID NOT VERIFY THE PERFORMANCE OF THE BALLOON AND INSPECT FOR PRESENCE OF CLOT AT THE TIP OR INFLATION HOLES. BLOOD DID NOT ENTER THE CATHETER LUMEN. CONTAST WAS NOT OBSERVED LEAKING DURING THE INFLATION ATTEMPT. THERE WERE NO KINKS ON THE CATHETER DURING USE. THE HCP TESTED THE BALLOON PRIOR TO USE. THE BALLOON DID NOT PERFORM AS INTENDED DURING TESTING. THE BALLOON LEAK WAS AT THE TIP. A 1ML SYRINGE WAS USED DURING INFLATION/DEFLATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21393275 ↗
  • MalfunctionAug 23, 2024Leak/Splash; Leak/Splash

    MEDTRONIC RECEIVED INFORMATION THAT A HYPERGLUIDE BALLOON WAS LEAKING. THE BALLOON LEAKED LIQUID. AFTER REPLACING IT, IT WAS USED SMOOTHLY. NO PATIENT SYMPTOMS OR COMPLICATIONS WERE REPORTED. THE DEVICE PREPARED AS INDICATED IN THE IFU. THE CAUSE OF THE BALLOON LEAK WAS NOT DETERMINED. THE MATCHING GUIDEWIRE WAS USED WITH THE BALLOON. THE GUIDEWIRE TIP WAS UNSHAPED. THE GUIDEWIRE WAS ABOUT 3CM FROM THE CATHETER TIP DURING INFLATION/DEFLATION. THE PHYSICIAN DID TEST THE BALLOON PRIOR TO USE. THE BALLOON DID PERFORM AS INTENDED DURING TESTING. THE CONTRAST RATIO WAS 50/50. THE INJECTION RATE WAS STEADY. A CARTRIDGE SYRINGE WAS USED DURING INFLATION/DEFLATION OF THE BALLOON.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20051214 ↗
  • MalfunctionJan 12, 2024Inflation Problem; Inflation Problem

    MEDTRONIC RECEIVED A REPORT THAT FOR A LARGE ANEURYSM OF THE C4 SEGMENT OF THE INTERNAL CAROTID ARTERY, 18-20MM, THE PHYSICIAN USED THE OCCLUDING BALLOON CATHETER SYSTEM TO ASSIST EMBOLIZATION. AFTER THE OCCLUDING BALLOON CATHETER SYSTEM WAS PLACED IN PLACE, THE BALLOON COULD NOT BE INFLATED AND THEN WITHDREW FROM THE BODY. IT ALSO COULD NOT BE INFLATED OUTSIDE THE BODY.  THE BALLOON WAS REPLACED AND THE SURGERY WENT SMOOTHLY. NO PATIENT SYMPTOMS OR FURTHER COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18502796 ↗
  • MalfunctionJul 10, 2023Leak/Splash; Leak/Splash

    ADDITIONAL INFORMATION WAS RECEIVED THAT THE CAUSE OF THE LEAK WAS UNKNOWN. THERE WAS NO OBVIOUS DAMAGE OR EVIDENCE OF TAMPERING TO DEVICE PACKAGING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17290604 ↗

FDA clearance and classification

Cleared through 510(k) K101570 on Jul 13, 2010 (decision: Substantially Equivalent). Submitted by Micro Therapeutics Dba Ev3 Neurovascular as “Hyperglide And Hyperform Occlusion Balloon Systems”. The FDA review took 39 days. FDA record ↗

Cleared through 510(k) K091458 on Oct 15, 2009 (decision: Substantially Equivalent). Submitted by Ev3 Neurovascular as “Hyperglide Occlusion Balloon System, Hyperform Occlusion Balloon System, Model 104-4112, 104-4113, 104-4127,104-4132,104”. The FDA review took 150 days. FDA record ↗

The FDA regulates product code MJN (catheter, intravascular occluding, temporary) under 21 CFR 870.4450 as a Class II device. 21 CFR 870.4450 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the HyperGlide 104-4112

Has the HyperGlide 104-4112 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00763000294359), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the HyperGlide 104-4112?

The FDA's MAUDE database lists 5 adverse event reports for this device, received between Jul 10, 2023 and Jun 11, 2025. In those reports, the most common specific device problems are “Inflation Problem” (3), “Leak/Splash” (3) and “Burst Container or Vessel” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the HyperGlide 104-4112?

None of the reports are classified as deaths, 0 as injuries and 5 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the HyperGlide 104-4112 FDA approved?

It was cleared through 510(k) K091458 on Oct 15, 2009; its record lists 2 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the HyperGlide 104-4112, and is it still sold?

The FDA's device database lists Micro Therapeutics, Inc. as the company responsible for it (brand HyperGlide, model 104-4112). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Catheter, intravascular occluding, temporary devices

All Catheter, intravascular occluding, temporary devices →