Skip to content

NA 2037A (UDI 00763000306496)Recalls & reported problems

Medtronic, Inc. · BOX 2037A AUTO CONNECTION BOX MDR · Class 3

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00763000306496), received between Apr 8, 2026 and Apr 8, 2026 — classified as 1 injury.

The most frequently reported specific device problem is “Protective Measures Problem” (1 report).

2 other recalls cover devices cleared under the same FDA submission (P020045); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • No Apparent Adverse Event1
  • Protective Measures Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 8, 2026Protective Measures Problem; No Apparent Adverse Event

    IT WAS REPORTED THAT DURING A CRYO ABLATION PROCEDURE, N-002 SYSTEM NOTICE OCCURRED. THE USER WAS UNABLE TO REMOVE THE COAX CAP AT THE TIME OF THE CALL, WHICH PREVENTED VERIFICATION OF A POSSIBLE COAX PORT OBSTRUCTION. THE CASE WAS ABORTED WHILE THE PATIENT WAS UNDER GENERAL ANESTHESIA. IT WAS REPORTED THAT 12 OF 14 ABLATION APPLICATIONS WERE ¿PERFECT,¿ AND THE N-002 OCCURRED ON APPLICATIONS 13 AND 14. A SUBSEQUENT INSPECTION FOUND THE COAX PORT TO BE CLEAR OF OBSTRUCTIONS, AND PHM RECORDS WERE RECEIVED AND REVIEWED. TEST ABLATIONS WERE REQUESTED/PERFORMED, DURING WHICH A FAULTY AUTO CONNECTION BOX WAS REPORTED AND THE CATHETER WAS NOT RECOGNIZED; REPLACING THE AUTO CONNECTION BOX RESOLVED THE CATHETER RECOGNITION ISSUE, AND TEST ABLATIONS WERE PROCEEDING. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24820959 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P020045. One submission often covers many models, so check the product and lot details in each record.

  • Z-2777-2015 · Sep 1, 2015 · Terminated
    Medtronic CryoConsole, models 106A3, 106E2, and 106A2-K For use in performing cardiac ablation procedures.