Skip to content

KYPHON® Express™ II Inflatable Bone Tamp KE102Recalls & reported problems

Medtronic Sofamor Danek, Inc. · BONE TAMP KE102 10/2 EXPRESS II IBT · Class 1

All models of KYPHON Express II Inflatable Bone Tamp (6)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00763000312015), received between Sep 28, 2021 and May 19, 2025 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture3

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Extravasation1
  • Low Blood Pressure/ Hypotension1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 19, 2025Material Rupture; Material Rupture

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING PERCUTANEOUS VE RTEBROPLASTY THERAPY FOR PRIMARY OSTEOPOROSIS WITH COMPRESSION FRACTURE. LEVELS IMPLANTED WAS 12. IT WAS REPORTED THAT THE BALLOON RUPTURED DURING EXPANSION. THE DOCTOR CONCLUDED THAT IT BURST UPON HITTING HARD BONE WITHIN THE VERTEBRAL BODY. THERE WAS NO PATIENT SYMPTOM REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT. THE KE102 IS INCLUDED IN THE KEX102EB AND DISTRIBUTED IN JAPAN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22044523 ↗
  • InjurySep 20, 2023Material Rupture; Material Rupture

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING BALLOON KYPHOPLASTY FOR T12 FRACTURE. IT WAS REPORTED THAT BILATERAL KYPHOPLASTY WAS BEING PERFORMED AND AFTER THE SURGEON USED THE BALLOON TO INFLATE ONE SIDE HE USED THE SAME BALLOON ON THE OTHER SIDE AND THAT'S WHEN THE BALLOON POPPED AND DYE WENT EVERYWHERE THEN ALL A SUDDEN THE PATIENTS BLOOD PRESSURE DROPPED, THE ANESTHESIOLOGIST GAVE FLUIDS TO KEEP THE BP UP AND THE SURGEON FINISHED THE KYPHOPLASTY BUT DIDN'T PUT AS MUCH CEMENT AS HE WOULD HAVE LIKED DUE TO THE ANESTHESIA PROVIDER ASKING TO TERMINATE THE PROCEDURE SO HE COULD GET THE PATIENT'S BP UP. PATIENT WOKE UP AFTERWARDS AND THEY KEPT AN ADDITIONAL HOUR TO MONITOR. THERE WAS NO OTHER SYMPTOMS REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: Extravasation; Low Blood Pressure/ Hypotension

    MAUDE report 17779375 ↗
  • MalfunctionSep 28, 2021Material Rupture; Material Rupture

    INFORMATION WAS RECEIVED FROM A HEALTHCARE PROFESSIONAL (HCP) VIA A MEDTRONIC REPRESENTATIVE REGARDING A PATIENT IMPLANTED WITH A SPINAL PRODUCT USING IBTS IN A KYPHOPLASTY FOR A VERTEBRAE FRACTURE AT T10 AND T12. IT WAS REPORTED THAT THE HCP TRIED TO CURVE THE IBTS BY BENDING THE INNER GUIDEWIRE. WHEN FORCING THE IBTS DOWN THE PATH CREATED IN THE BONE, THE GUIDEWIRE PUSHED FORWARD AND PUNCTURED THE IBTS. THE RUPTURES OCCURRED DURING INFLATION. THERE WERE NO PATIENT SYMPTOMS OR COMPLICATIONS AS A RESULT OF THE EVENT. THERE WAS A DELAY IN OVERALL PROCEDURE TIME BY APPROX. 3 MINUTES. THERE WAS NO IN-PATIENT HOSPITALIZATION OR PROLONGATION OF EXISTING HOSPITALIZATION NECESSARY OR TREATMENT OR ADDITIONAL SURGERY PERFORMED AS A RESULT OF THIS EVENT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A NEW PRODUCT. NO FURTHER COMPLICATIONS WERE REPORTED/ ANTICIPATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12538995 ↗

Other TAMP devices

All TAMP devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page