Exacta™ 27666Recalls & reported problems

Medtronic Ps Medical, Inc. · EDM 27666 EXCTA W GLUE STPCK GR PT LN · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Medtronic Exacta is a device used to monitor intracranial pressure and drain cerebrospinal fluid. Most of the 124 reports describe fluid leaking from the system, and some describe cracks near the stopcock. The reports include no deaths and no injuries. There is 1 open recall, from November 2024, over cracks and leaks at the drainage system's stopcocks.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
124
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 124 adverse event reports that list this exact device (DI 00763000333515), received between Jul 19, 2023 and Dec 9, 2024 — classified as 124 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (121 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 117 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 2 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports202420252026117
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 1, 2024Open, Classified

    Due to complaints received reporting cracks and/or leaks of the stopcocks associated with the external drainage and monitoring system (EDMS).

    Z-0662-2025 · Root cause: Under Investigation by firm · FDA record ↗

Most reported problems

Device problems

  • Fluid/Blood Leak121
  • Fracture6
  • Break4

Patient problems

  • No Clinical Signs, Symptoms or Conditions124

Recent adverse event reports

The 25 most recent of 124 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 9, 2024Break; Fracture; Break; Fracture

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THAT THE DRAINAGE SYSTEM WAS FRACTURING; SPECIFICALLY, THE PORTION NEAR THE STOPCOCK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20881813 ↗
  • MalfunctionDec 9, 2024Break; Fracture; Break; Fracture

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THAT THE DRAINAGE SYSTEM WAS FR ACTURING; SPECIFICALLY, THE PORTION NEAR THE STOPCOCK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20881000 ↗
  • MalfunctionSep 5, 2024Break; Fracture; Break; Fracture

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THAT THE SMARTSITE INJECTION SITE ON THE DRAINAGE SYSTEM WHERE CEREBROSPINAL FLUID (CSF) SAMPLE WAS EXTRACTED WAS PUSHED IN. THE SYSTEM USED A BAXTER INTERLINK NEEDLELESS INJECTION SITE AND DID NOT HAVE A SMARTSITE INJECTION SITE. THERE WAS NO INFORMATION PROVIDED REGARDING A SYRINGE USED FOR OBTAINING THE SAMPLE AND IF IT WAS APPROPRIATE FOR THE SYSTEM. THE SYSTEM WAS STILL IN USE. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20145358 ↗
  • MalfunctionFeb 7, 2024Fluid/Blood Leak; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THERE WAS A LEAK AT A CONNECTION POINT. THE DEVICE WAS SWITCHED OUT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18658400 ↗
  • MalfunctionJan 11, 2024Break; Fluid/Blood Leak; Fracture; Break; Fluid/Blood Leak; Fracture

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THAT THE DRAINAGE SYSTEM (BURETROL) WAS LEAKING CEREBROSPINAL FLUID (CSF) ON THE THREE WAY STOPCOCK. A NEW DRAIN WAS PLACED, AND THE SITE THOUGHT THERE WAS A SLIGHT CRACK IN THE STOPCOCK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18491567 ↗
  • MalfunctionNov 30, 2023Fluid/Blood Leak; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THE DEVICE WAS DEFECTIVE/DAMAGED DUE TO LEAKING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18242934 ↗
  • MalfunctionNov 30, 2023Fluid/Blood Leak; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THE DEVICE WAS DEFECTIVE/DAMAGED DUE TO LEAKING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18242686 ↗
  • MalfunctionNov 30, 2023Fluid/Blood Leak; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THE DEVICE WAS DEFECTIVE/DAMAGED DUE TO LEAKING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18242675 ↗
  • MalfunctionNov 30, 2023Fluid/Blood Leak; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THE DEVICE WAS DEFECTIVE/DAMAGED DUE TO LEAKING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18242670 ↗
  • MalfunctionNov 30, 2023Fluid/Blood Leak; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THE DEVICE WAS DEFECTIVE/DAMAGED DUE TO LEAKING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18242655 ↗
  • MalfunctionNov 30, 2023Fluid/Blood Leak; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THE DEVICE WAS DEFECTIVE/DAMAGED DUE TO LEAKING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18242082 ↗
  • MalfunctionNov 30, 2023Fluid/Blood Leak; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THE DEVICE WAS DEFECTIVE/DAMAGED DUE TO LEAKING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18241998 ↗
  • MalfunctionNov 30, 2023Fluid/Blood Leak; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THE DEVICE WAS DEFECTIVE/DAMAGED DUE TO LEAKING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18241518 ↗
  • MalfunctionNov 30, 2023Fluid/Blood Leak; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THE DEVICE WAS DEFECTIVE/DAMAGED DUE TO LEAKING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18241361 ↗
  • MalfunctionNov 30, 2023Fluid/Blood Leak; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THE DEVICE WAS DEFECTIVE/DAMAGED DUE TO LEAKING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18241353 ↗
  • MalfunctionNov 30, 2023Fluid/Blood Leak; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THE DEVICE WAS DEFECTIVE/DAMAGED DUE TO LEAKING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18241166 ↗
  • MalfunctionNov 30, 2023Fluid/Blood Leak; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THE DEVICE WAS DEFECTIVE/DAMAGED DUE TO LEAKING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18241000 ↗
  • MalfunctionNov 30, 2023Fluid/Blood Leak; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THE DEVICE WAS DEFECTIVE/DAMAGED DUE TO LEAKING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18240882 ↗
  • MalfunctionNov 30, 2023Fluid/Blood Leak; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THE DEVICE WAS DEFECTIVE/DAMAGED DUE TO LEAKING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18240675 ↗
  • MalfunctionNov 30, 2023Fluid/Blood Leak; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THE DEVICE WAS DEFECTIVE/DAMAGED DUE TO LEAKING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18240530 ↗
  • MalfunctionNov 30, 2023Fluid/Blood Leak; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THE DEVICE WAS DEFECTIVE/DAMAGED DUE TO LEAKING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18240333 ↗
  • MalfunctionNov 30, 2023Fluid/Blood Leak; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THE DEVICE WAS DEFECTIVE/DAMAGED DUE TO LEAKING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18240262 ↗
  • MalfunctionNov 30, 2023Fluid/Blood Leak; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THE DEVICE WAS DEFECTIVE/DAMAGED DUE TO LEAKING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18239583 ↗
  • MalfunctionNov 30, 2023Fluid/Blood Leak; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THE DEVICE WAS DEFECTIVE/DAMAGED DUE TO LEAKING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18239552 ↗
  • MalfunctionNov 30, 2023Fluid/Blood Leak; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THE DEVICE WAS DEFECTIVE/DAMAGED DUE TO LEAKING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18239214 ↗

Other Device, monitoring, intracranial pressure devices

All Device, monitoring, intracranial pressure devices →