Becker® 27761 (UDI 00763000333577)Recalls & reported problems
Medtronic Ps Medical, Inc. · BECKER 27761 STOP BCNDLESS PATLN 24 · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00763000333577), received between Dec 13, 2023 and Feb 26, 2024 — classified as 8 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (5 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Nov 1, 2024Open, Classified
Due to complaints received reporting cracks and/or leaks of the stopcocks associated with the external drainage and monitoring system (EDMS).
Z-0661-2025 · Root cause: Under Investigation by firm · FDA record ↗
Most reported problems
Device problems
- Fluid/Blood Leak5
- Break4
- Fracture4
Patient problems
- No Clinical Signs, Symptoms or Conditions8
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 26, 2024Break; Fracture; Break; Fracture
ADDITIONAL INFORMATION RECEIVED REPORTED THAT THE HEALTHCARE PROFESSIONAL (HCP) WENT DOWN TO THE EMERGENCY DEPARTMENT TO HELP WITH T HE SETUP OF THE DEVICE WITH THE EMERGENCY ROOM STAFF PER THE REQUEST OF THE STROKE COORDINATOR DUE TO THE FACT THAT THEY WERE ATTEMPTING TO SET UP THE SYSTEM AND IT WAS LEAKING FROM THE PRIMING OF THE SYSTEM FROM THE PROVIDER. THE NEXT THREE THAT WERE CHECKED ALSO HAD CRACKS NEAR THE SAME LOCATIONS. THE LOCATION ON THESE DEVICES WERE NEAR WHERE YOU HOOK-UP THE TRANSDUCER TO THE STOPCOCK. THIS WOULD BE THE STOPCOCK THAT LED TO THE DRAINAGE SYSTEM BAG SYSTEM-THE BOTTOM SIDE OF THE STOPCOCK. ALL THREE DEVICES HAD HAIRLINE CRACKS THROUGH THE TUBING THAT YOU COULD FEEL WITH YOUR FINGER. THE PATIENT DID HAVE A SYSTEM PLACED WITH A DIFFERENT MODEL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18776602 ↗MalfunctionFeb 8, 2024Break; Fracture; Break; Fracture
ADDITIONAL INFORMATION RECEIVED REPORTED THAT THE HEALTHCARE PROFESSIONAL (HCP) WENT DOWN TO THE EMERGENCY DEPARTMENT TO HELP WITH T HE SETUP OF THE DEVICE WITH THE EMERGENCY ROOM STAFF PER THE REQUEST OF THE STROKE COORDINATOR DUE TO THE FACT THAT THEY WERE ATTEMPTING TO SET UP THE SYSTEM AND IT WAS LEAKING FROM THE PRIMING OF THE SYSTEM FROM THE PROVIDER. THE NEXT THREE THAT WERE CHECKED ALSO HAD CRACKS NEAR THE SAME LOCATIONS. THE LOCATION ON THESE DEVICES WERE NEAR WHERE YOU HOOK-UP THE TRANSDUCER TO THE STOPCOCK. THIS WOULD BE THE STOPCOCK THAT LED TO THE DRAINAGE SYSTEM BAG SYSTEM-THE BOTTOM SIDE OF THE STOPCOCK. ALL THREE DEVICES HAD HAIRLINE CRACKS THROUGH THE TUBING THAT YOU COULD FEEL WITH YOUR FINGER. THE PATIENT DID HAVE A SYSTEM PLACED WITH A DIFFERENT MODEL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18669340 ↗MalfunctionFeb 8, 2024Break; Fracture; Break; Fracture
ADDITIONAL INFORMATION RECEIVED REPORTED THAT THE HEALTHCARE PROFESSIONAL (HCP) WENT DOWN TO THE EMERGENCY DEPARTMENT TO HELP WITH T HE SETUP OF THE DEVICE WITH THE EMERGENCY ROOM STAFF PER THE REQUEST OF THE STROKE COORDINATOR DUE TO THE FACT THAT THEY WERE ATTEMPTING TO SET UP THE SYSTEM AND IT WAS LEAKING FROM THE PRIMING OF THE SYSTEM FROM THE PROVIDER. THE NEXT THREE THAT WERE CHECKED ALSO HAD CRACKS NEAR THE SAME LOCATIONS. THE LOCATION ON THESE DEVICES WERE NEAR WHERE YOU HOOK-UP THE TRANSDUCER TO THE STOPCOCK. THIS WOULD BE THE STOPCOCK THAT LED TO THE DRAINAGE SYSTEM BAG SYSTEM-THE BOTTOM SIDE OF THE STOPCOCK. ALL THREE DEVICES HAD HAIRLINE CRACKS THROUGH THE TUBING THAT YOU COULD FEEL WITH YOUR FINGER. THE PATIENT DID HAVE A SYSTEM PLACED WITH A DIFFERENT MODEL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18669325 ↗MalfunctionFeb 8, 2024Break; Fluid/Blood Leak; Fracture; Break; Fluid/Blood Leak; Fracture
ADDITIONAL INFORMATION RECEIVED REPORTED THAT THE HEALTHCARE PROFESSIONAL (HCP) WENT DOWN TO THE EMERGENCY DEPARTMENT TO HELP WITH T HE SETUP OF THE DEVICE WITH THE EMERGENCY ROOM STAFF PER THE REQUEST OF THE STROKE COORDINATOR DUE TO THE FACT THAT THEY WERE ATTEMPTING TO SET UP THE SYSTEM AND IT WAS LEAKING FROM THE PRIMING OF THE SYSTEM FROM THE PROVIDER. THE NEXT THREE THAT WERE CHECKED ALSO HAD CRACKS NEAR THE SAME LOCATIONS. THE LOCATION ON THESE DEVICES WERE NEAR WHERE YOU HOOK-UP THE TRANSDUCER TO THE STOPCOCK. THIS WOULD BE THE STOPCOCK THAT LED TO THE DRAINAGE SYSTEM BAG SYSTEM-THE BOTTOM SIDE OF THE STOPCOCK. ALL THREE DEVICES HAD HAIRLINE CRACKS THROUGH THE TUBING THAT YOU COULD FEEL WITH YOUR FINGER. THE PATIENT DID HAVE A SYSTEM PLACED WITH A DIFFERENT MODEL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18669280 ↗MalfunctionDec 13, 2023Fluid/Blood Leak; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THAT THE SYSTEM'S STOPCOCK WAS LEAKING. THIS ISSUE WAS DISCOVERED DURING SET UP SO THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18318763 ↗MalfunctionDec 13, 2023Fluid/Blood Leak; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THAT THE SYSTEM'S STOPCOCK WAS LEAKING. THIS ISSUE WAS DISCOVERED DURING SET UP SO THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18318734 ↗MalfunctionDec 13, 2023Fluid/Blood Leak; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THAT THE SYSTEM'S STOPCOCK WAS LEAKING. THIS ISSUE WAS DISCOVERED DURING SET UP SO THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18318688 ↗MalfunctionDec 13, 2023Fluid/Blood Leak; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THAT THERE WAS LEAKAGE FROM THE PRECONNECTED STOPCOCK. THE PATIENT WAS NOT IMPACTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18318453 ↗
Other Shunt, central nervous system and components devices
- NATUS EDS3 DRAINAGE SYSTEM NT821731CNatus Medical Incorporated · Shunt, Central Nervous System And Components1 recall208 reports
- Exacta™ 46700Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall2 deaths163 reports
- Strata® 42866Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 death160 reports
- proGAV 2.0® FX410TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components119 reports
- PROGAV 2.0 FX643TAesculap AG · Shunt, central nervous system and components100 reports
- Becker® 46128Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall50 reports
- M.BLUE PLUS FX804TAesculap AG · Shunt, central nervous system and components82 reports
- proSA® FV701TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components82 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.