Skip to content

Becker® 27761 (UDI 00763000333577)Recalls & reported problems

Medtronic Ps Medical, Inc. · BECKER 27761 STOP BCNDLESS PATLN 24 · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
8
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00763000333577), received between Dec 13, 2023 and Feb 26, 2024 — classified as 8 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (5 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 4 reportsJan 2024: 0 reportsFeb 2024: 4 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 1, 2024Open, Classified

    Due to complaints received reporting cracks and/or leaks of the stopcocks associated with the external drainage and monitoring system (EDMS).

    Z-0661-2025 · Root cause: Under Investigation by firm · FDA record ↗

Most reported problems

Device problems

  • Fluid/Blood Leak5
  • Break4
  • Fracture4

Patient problems

  • No Clinical Signs, Symptoms or Conditions8

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 26, 2024Break; Fracture; Break; Fracture

    ADDITIONAL INFORMATION RECEIVED REPORTED THAT THE HEALTHCARE PROFESSIONAL (HCP) WENT DOWN TO THE EMERGENCY DEPARTMENT TO HELP WITH T HE SETUP OF THE DEVICE WITH THE EMERGENCY ROOM STAFF PER THE REQUEST OF THE STROKE COORDINATOR DUE TO THE FACT THAT THEY WERE ATTEMPTING TO SET UP THE SYSTEM AND IT WAS LEAKING FROM THE PRIMING OF THE SYSTEM FROM THE PROVIDER. THE NEXT THREE THAT WERE CHECKED ALSO HAD CRACKS NEAR THE SAME LOCATIONS. THE LOCATION ON THESE DEVICES WERE NEAR WHERE YOU HOOK-UP THE TRANSDUCER TO THE STOPCOCK. THIS WOULD BE THE STOPCOCK THAT LED TO THE DRAINAGE SYSTEM BAG SYSTEM-THE BOTTOM SIDE OF THE STOPCOCK. ALL THREE DEVICES HAD HAIRLINE CRACKS THROUGH THE TUBING THAT YOU COULD FEEL WITH YOUR FINGER. THE PATIENT DID HAVE A SYSTEM PLACED WITH A DIFFERENT MODEL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18776602 ↗
  • MalfunctionFeb 8, 2024Break; Fracture; Break; Fracture

    ADDITIONAL INFORMATION RECEIVED REPORTED THAT THE HEALTHCARE PROFESSIONAL (HCP) WENT DOWN TO THE EMERGENCY DEPARTMENT TO HELP WITH T HE SETUP OF THE DEVICE WITH THE EMERGENCY ROOM STAFF PER THE REQUEST OF THE STROKE COORDINATOR DUE TO THE FACT THAT THEY WERE ATTEMPTING TO SET UP THE SYSTEM AND IT WAS LEAKING FROM THE PRIMING OF THE SYSTEM FROM THE PROVIDER. THE NEXT THREE THAT WERE CHECKED ALSO HAD CRACKS NEAR THE SAME LOCATIONS. THE LOCATION ON THESE DEVICES WERE NEAR WHERE YOU HOOK-UP THE TRANSDUCER TO THE STOPCOCK. THIS WOULD BE THE STOPCOCK THAT LED TO THE DRAINAGE SYSTEM BAG SYSTEM-THE BOTTOM SIDE OF THE STOPCOCK. ALL THREE DEVICES HAD HAIRLINE CRACKS THROUGH THE TUBING THAT YOU COULD FEEL WITH YOUR FINGER. THE PATIENT DID HAVE A SYSTEM PLACED WITH A DIFFERENT MODEL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18669340 ↗
  • MalfunctionFeb 8, 2024Break; Fracture; Break; Fracture

    ADDITIONAL INFORMATION RECEIVED REPORTED THAT THE HEALTHCARE PROFESSIONAL (HCP) WENT DOWN TO THE EMERGENCY DEPARTMENT TO HELP WITH T HE SETUP OF THE DEVICE WITH THE EMERGENCY ROOM STAFF PER THE REQUEST OF THE STROKE COORDINATOR DUE TO THE FACT THAT THEY WERE ATTEMPTING TO SET UP THE SYSTEM AND IT WAS LEAKING FROM THE PRIMING OF THE SYSTEM FROM THE PROVIDER. THE NEXT THREE THAT WERE CHECKED ALSO HAD CRACKS NEAR THE SAME LOCATIONS. THE LOCATION ON THESE DEVICES WERE NEAR WHERE YOU HOOK-UP THE TRANSDUCER TO THE STOPCOCK. THIS WOULD BE THE STOPCOCK THAT LED TO THE DRAINAGE SYSTEM BAG SYSTEM-THE BOTTOM SIDE OF THE STOPCOCK. ALL THREE DEVICES HAD HAIRLINE CRACKS THROUGH THE TUBING THAT YOU COULD FEEL WITH YOUR FINGER. THE PATIENT DID HAVE A SYSTEM PLACED WITH A DIFFERENT MODEL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18669325 ↗
  • MalfunctionFeb 8, 2024Break; Fluid/Blood Leak; Fracture; Break; Fluid/Blood Leak; Fracture

    ADDITIONAL INFORMATION RECEIVED REPORTED THAT THE HEALTHCARE PROFESSIONAL (HCP) WENT DOWN TO THE EMERGENCY DEPARTMENT TO HELP WITH T HE SETUP OF THE DEVICE WITH THE EMERGENCY ROOM STAFF PER THE REQUEST OF THE STROKE COORDINATOR DUE TO THE FACT THAT THEY WERE ATTEMPTING TO SET UP THE SYSTEM AND IT WAS LEAKING FROM THE PRIMING OF THE SYSTEM FROM THE PROVIDER. THE NEXT THREE THAT WERE CHECKED ALSO HAD CRACKS NEAR THE SAME LOCATIONS. THE LOCATION ON THESE DEVICES WERE NEAR WHERE YOU HOOK-UP THE TRANSDUCER TO THE STOPCOCK. THIS WOULD BE THE STOPCOCK THAT LED TO THE DRAINAGE SYSTEM BAG SYSTEM-THE BOTTOM SIDE OF THE STOPCOCK. ALL THREE DEVICES HAD HAIRLINE CRACKS THROUGH THE TUBING THAT YOU COULD FEEL WITH YOUR FINGER. THE PATIENT DID HAVE A SYSTEM PLACED WITH A DIFFERENT MODEL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18669280 ↗
  • MalfunctionDec 13, 2023Fluid/Blood Leak; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THAT THE SYSTEM'S STOPCOCK WAS LEAKING. THIS ISSUE WAS DISCOVERED DURING SET UP SO THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18318763 ↗
  • MalfunctionDec 13, 2023Fluid/Blood Leak; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THAT THE SYSTEM'S STOPCOCK WAS LEAKING. THIS ISSUE WAS DISCOVERED DURING SET UP SO THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18318734 ↗
  • MalfunctionDec 13, 2023Fluid/Blood Leak; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THAT THE SYSTEM'S STOPCOCK WAS LEAKING. THIS ISSUE WAS DISCOVERED DURING SET UP SO THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18318688 ↗
  • MalfunctionDec 13, 2023Fluid/Blood Leak; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THAT THERE WAS LEAKAGE FROM THE PRECONNECTED STOPCOCK. THE PATIENT WAS NOT IMPACTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18318453 ↗

Other Shunt, central nervous system and components devices

All Shunt, central nervous system and components devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.