N/A 9732353 (UDI 00763000340179)Recalls & reported problems

Medtronic Navigation, Inc. · FRAME 9732353 PERC REF CROSS PIN · Class 2

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00763000340179), received between Sep 2, 2021 and Mar 10, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Mechanical Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Problem2
  • Activation, Positioning or Separation Problem1
  • Device Sensing Problem1
  • Incorrect, Inadequate or Imprecise Result or Readings1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 10, 2026Mechanical Problem; Incorrect, Inadequate or Imprecise Result or Readings; Activation, Positioning or Separation Problem

    MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED DURING A SACROILIAC AND THORACOLUMBAR PROCEDURE. IT WAS REPORTED THAT WHEN THE FRAME WAS PLACED ON THE PERCUTANEOUS PIN AND IMAGE WAS TAKEN, THE FRAME ROTATED AWAY BECAUSE THE NOTCH HAD BROKEN OFF. IT WAS UNKNOWN WHEN THE FRAME BROKE. IT WAS REPORTED THAT THE FRAME MOVED SINCE THE PERC PIN WAS NOT HOLDING THE FRAME IN PLACE. THE MOVEMENT CAUSED AN INACCURACY OF APPROXIMATELY 50MM. THE PROCEDURE WAS COMPLETED BY OPENING A NEW TRAY, PLACING A NEW STERILE REFERENCE FRAME ON THE PERC PIN AND COMPLETING A NEW IMAGING SYSTEM SPIN. THERE WAS A SURGICAL DELAY OF LESS THAN 1 HOUR. THERE WAS NO IMPACT ON PATIENT OUTCOME. 2026-FEB-26 E1 (REP): ADDITIONAL INFORMATION WAS RECEIVED. IT WAS REPORTED THAT THE FRAME MOVED SINCE THE PERC PIN WAS NOT HOLDING THE FRAME IN PLACE. THE MOVEMENT CAUSED AN INACCURACY OF APPROXIMATELY 50MM. THE PROCEDURE WAS COMPLETED BY OPENING A NEW TRAY, PLACING A NEW STERILE REFERENCE FRAME ON THE PERC PIN AND COMPLETING A NEW IMAGING SYSTEM SPIN. 2026-MAR-06 E2 (REP): NO NEW INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24559064 ↗
  • MalfunctionFeb 10, 2026Mechanical Problem

    MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT THE PERCUTANEOUS (PERC) PIN REFERENCE FRAME WAS OBSERVED TO HAVE DAMAGE TO THE LOCKING MECHANISM, PREVENTING IT FROM LOCKING EFFECTIVELY TO THE PERC PIN. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24307504 ↗
  • MalfunctionSep 2, 2021Device Sensing Problem; Device Sensing Problem

    ADDITIONAL INFORMATION REGARDING A NAVIGATION SYSTEM. TROUBLESHOOTING WAS PERFORMED BUT WAS UNSUCCESSFUL TO RESOLVE THE ISSUE. THE SYSTEM WAS RESTARTED, CAMERA REPOSITIONED AND THE SITE WIPED DOWN THE SPHERES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12413619 ↗