N/A 9732353 (UDI 00763000340179)Recalls & reported problems
Medtronic Navigation, Inc. · FRAME 9732353 PERC REF CROSS PIN · Class 2
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00763000340179), received between Sep 2, 2021 and Mar 10, 2026 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Mechanical Problem” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Mechanical Problem2
- Activation, Positioning or Separation Problem1
- Device Sensing Problem1
- Incorrect, Inadequate or Imprecise Result or Readings1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 10, 2026Mechanical Problem; Incorrect, Inadequate or Imprecise Result or Readings; Activation, Positioning or Separation Problem
MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED DURING A SACROILIAC AND THORACOLUMBAR PROCEDURE. IT WAS REPORTED THAT WHEN THE FRAME WAS PLACED ON THE PERCUTANEOUS PIN AND IMAGE WAS TAKEN, THE FRAME ROTATED AWAY BECAUSE THE NOTCH HAD BROKEN OFF. IT WAS UNKNOWN WHEN THE FRAME BROKE. IT WAS REPORTED THAT THE FRAME MOVED SINCE THE PERC PIN WAS NOT HOLDING THE FRAME IN PLACE. THE MOVEMENT CAUSED AN INACCURACY OF APPROXIMATELY 50MM. THE PROCEDURE WAS COMPLETED BY OPENING A NEW TRAY, PLACING A NEW STERILE REFERENCE FRAME ON THE PERC PIN AND COMPLETING A NEW IMAGING SYSTEM SPIN. THERE WAS A SURGICAL DELAY OF LESS THAN 1 HOUR. THERE WAS NO IMPACT ON PATIENT OUTCOME. 2026-FEB-26 E1 (REP): ADDITIONAL INFORMATION WAS RECEIVED. IT WAS REPORTED THAT THE FRAME MOVED SINCE THE PERC PIN WAS NOT HOLDING THE FRAME IN PLACE. THE MOVEMENT CAUSED AN INACCURACY OF APPROXIMATELY 50MM. THE PROCEDURE WAS COMPLETED BY OPENING A NEW TRAY, PLACING A NEW STERILE REFERENCE FRAME ON THE PERC PIN AND COMPLETING A NEW IMAGING SYSTEM SPIN. 2026-MAR-06 E2 (REP): NO NEW INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24559064 ↗MalfunctionFeb 10, 2026Mechanical Problem
MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT THE PERCUTANEOUS (PERC) PIN REFERENCE FRAME WAS OBSERVED TO HAVE DAMAGE TO THE LOCKING MECHANISM, PREVENTING IT FROM LOCKING EFFECTIVELY TO THE PERC PIN. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24307504 ↗MalfunctionSep 2, 2021Device Sensing Problem; Device Sensing Problem
ADDITIONAL INFORMATION REGARDING A NAVIGATION SYSTEM. TROUBLESHOOTING WAS PERFORMED BUT WAS UNSUCCESSFUL TO RESOLVE THE ISSUE. THE SYSTEM WAS RESTARTED, CAMERA REPOSITIONED AND THE SITE WIPED DOWN THE SPHERES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12413619 ↗
Other Intervertebral fusion device with bone graft, lumbar devices
- T-PAL 03.812.003Synthes GmbH · Intervertebral fusion device with bone graft, lumbar152 reports
- T-PAL 03.812.004Synthes GmbH · Intervertebral fusion device with bone graft, lumbar120 reports
- Catalyft PL Expandable Interbody System 6068073Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall66 reports
- Catalyft PL Expandable Interbody System 6069076Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall51 reports
- Catalyft PL Expandable Interbody System 6068076Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall46 reports
- EIT TLIF Inserter TFT30101EIT Emerging Implant Technologies GmbH · Intervertebral Fusion Device With Bone Graft, Lumbar88 reports
- T-PAL 03.812.001Synthes GmbH · Intervertebral fusion device with bone graft, lumbar76 reports
- Catalyft PL Expandable Interbody System 6069096Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall25 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.