Skip to content

NA 2035URecalls & reported problems

Medtronic, Inc. · CABLE 2035U ELECTRICAL UMBILICAL BMED · Class 3

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00763000359355), received between Nov 18, 2025 and Nov 18, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Off-Label Use” (1 report).

2 other recalls cover devices cleared under the same FDA submission (P020045); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Off-Label Use1
  • Protective Measures Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 18, 2025Off-Label Use; Protective Measures Problem; Off-Label Use; Protective Measures Problem

    IT WAS REPORTED THAT WHEN THE ELECTRICAL CABLE FOR THE CARDIAC ABLATION CATHETER WAS CONNECTED, THE SYSTEM DID NOT RECOGNIZE THE CATHETER AND A SYSTEM NOTICE APPEARED INDICATING AN UNRECOGNIZED CATHETER. ATTEMPTS TO DISCONNECT AND RECONNECT THE CABLE DID NOT RESOLVE THE ISSUE, SO THE ELECTRICAL CABLE WAS REPLACED, WHICH RESOLVED THE PROBLEM. THE CASE WAS COMPLETED WITH CRYO. IT WAS LATER NOTED THAT THE ELECTRICAL UMBILICAL CABLE WAS USED PAST ITS USE BY DATE AND WAS EXPIRED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23579697 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P020045. One submission often covers many models, so check the product and lot details in each record.

  • Z-2777-2015 · Sep 1, 2015 · Terminated
    Medtronic CryoConsole, models 106A3, 106E2, and 106A2-K For use in performing cardiac ablation procedures.
  • Z-1452-06 · Aug 4, 2006 · Terminated
    7F Freezor Cardiac Cryoablation catheter, REF # 207F1. CryoCath Technologies Inc.

Other Cardiac ablation percutaneous catheter devices

All Cardiac ablation percutaneous catheter devices →