PhenomTM 21 Catheter FG13160-0615-1S (UDI 00763000399696)Recalls & reported problems
Micro Therapeutics, Inc. · CATHETER FG13160-0615-1S V06 US · Class 2
All models of PhenomTM 21 Catheter (4)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00763000399696), received between Aug 31, 2023 and Aug 31, 2026 — classified as 8 malfunctions.
The most frequently reported specific device problem is “Physical Resistance/Sticking” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Physical Resistance/Sticking7
- Material Deformation2
- Retraction Problem2
- Break1
- Fracture1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 31, 2026Retraction Problem; Physical Resistance/Sticking
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50522620 ↗MalfunctionApr 20, 2026Retraction Problem
MEDTRONIC RECEIVED A REPORT THAT A SOLITAIRE FR STENT ENCOUNTERED RESISTANCE DURING RETRIEVAL. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF AN ISCHEMIC STROKE, LOCATED ON THE RIGHT MIDDLE CEREBRAL ARTERY (MCA). IT WAS NOTED THE PATIENT'S PARENT VESSEL WAS THE M1 WHICH HAD A 2.6MM DIAMETER AND THAT THE VESSEL TORTUOSITY WAS SEVERE. STROKE ONSET TO REPERFUSION TIME WAS 7 HOURS. THE MRS WAS 2 FOR BOTH THE BASELINE AND PROCEDURE. NIHSS SCORE WAS 10 AT BASELINE AND 4 POST PROCEDURE. THE TICI SCORE WAS 1 AT BASELINE AND 2B POST PROCEDURE. IV TPA WAS CONTRAINDICATED AND THE DEVICE HAD A TOTAL OF 2 PASSES. IT WAS REPORTED THAT ACCORDING TO THE DOCTOR, WHILE RETRIEVING THE SECOND PASS THERE WAS TOO MUCH RESISTANCE AND FORCE HAD TO BE APPLIED TO PULL THE STENT RETRIEVER. HENCE, THE SOLITAIRE AND PHENOM 21 WAS REMOVED. CASE DONE BY USING COMPETITION PRODUCT. THERE WAS RESISTANCE AT THE MIDDLE SECTION OF THE CATHETER DURING RETRIEVAL. ALL DEVICES WERE PREPARED AS INDICATED IN THE IFU. NO PATIENT COMPLICATIONS WERE ASSOCIATED WITH THIS EVENT. ANCILLARY DEVICES INCLUDE A PHENOM 21 MICRO CATHETER, RED 71 GUIDE CATHETER, NURON MAX SHEATH. ADDITIONAL EVENT DETAILS PROVIDED. THE CAUSE OF THE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24933041 ↗MalfunctionOct 15, 2025Material Deformation; Physical Resistance/Sticking; Material Deformation; Physical Resistance/Sticking
ADDITIONAL INFORMATION WAS RECEIVED. THE CAUSE OF THE EVENT WAS NOT DETERMINED. THERE WAS NO FRICTION OR DIFFICULTY DURING DELIVERY; HOWEVER, THE DEVICE COULD NOT BE RESHEATHED TO REPOSITION. ADDITIONALLY, IT WAS CLARIFIED THAT THE THIRD ATTEMPT AT TREATMENT WAS SUCCESSFUL WITH THE PHENOM 027 AND PIPELINE SHIELD, NOT WITH VANTAGE AS PREVIOUSLY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23294176 ↗MalfunctionJul 11, 2025Physical Resistance/Sticking; Physical Resistance/Sticking
MEDTRONIC RECEIVED A REPORT THAT THE PATIENT WAS UNDERGOING TREATMENT FOR MIDDLE CEREBRAL ARTERY (MCA) ISCHEMIC CEREBRAL (IC) INFARCTION. IT WAS NOTED THAT THE PATIENT'S VESSEL TORTUOSITY WAS MODERATE. THE PATIENT'S PRE-OPERATIVE TICI SCORE WAS 1, AND THE POST-OPERATIVE TICI SCORE IMPROVED TO 2B. THE PATIENT WAS TREATED WITH A THROMBECTOMY PROCEDURE USING A TPS (THROMBECTOMY PLATFORM SYSTEM). THE SOLITAIRE DEVICE WAS USED, WITH ZERO PASSES PERFORMED DURING THE PROCEDURE. THE TIME REQUIRED FROM THE ONSET OF CEREBRAL INFARCTION TO REVASCULARIZATION/REVASCULARIZE WAS 30 MINUTES. IT WAS REPORTED THAT THE PHENOM 21 AND SOLITAIRE WERE SELECTED. THE DEGREE OF VASCULAR TORTUOSITY LEADING TO THE MCA WAS MODERATE. THE OPTIMO 9FR WAS DELIVERED TO THE IC, AND THE PHENOM 21 WAS PREPARED ACCORDING TO THE INSTRUCTIONS FOR USE (IFU). THE CHIKAI 14 WIRE, WHICH WAS ALSO PREPARED ACCORDING TO THE IFU, WAS SET UP. AFTER DELIVERING THE PHENOM 21 TO THE OCCLUSION SITE, AN ATTEMPT WAS MADE TO DELIVER THE SOLITAIRE STENT, WHICH HAD BEEN PREPARED ACCORDING TO THE IFU, INTO THE PHENOM 21. HOWEVER, STRONG RESISTANCE AND STIFFNESS WERE FELT. DESPITE SENSING ABNORMALITIES IN BOTH THE PHENOM 21 AND THE SOLITAIR
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22486207 ↗MalfunctionMay 16, 2025Physical Resistance/Sticking; Physical Resistance/Sticking
MEDTRONIC RECEIVED A REPORT THAT A SOLITAIREX6-40 ENCOUNTERED RESISTANCE IN THE DISTAL SHAFT OF THE PHENOM. STENT DISCONNECTION WAS ALSO NOTED AS THE SOLITAIRE DELIVERY PUSHER BECAME DISCONNECTED WHEN IT WAS PULLED INTO THE PHENOM OUTSIDE THE BODY. THE PATIENT WAS UNDERGOING THROMBECTOMY FOR TREATMENT OF AN ISCHEMIC CEREBRAL INFARCTION IN THE RIGHT M1. IT WAS NOTED THE PATIENT'S VESSEL TORTUOSITY WAS MODERATE. THE ACCESS VESSEL WAS THE MIDDLE CEREBRAL ARTERY (MCA)/INTERNAL CAROTID ARTERY (ICA) WITH A 2-5MM DIAMETER. TICI (POSTOPERATIVE): NO PROBLEMS. THE PATIENT WAS TREATED WITH A TPS. THERE WERE 2 PASSES FOR THE SOLITAIRE. TIME REQUIRED FROM ONSET OF CEREBRAL INFARCTION TO REVASCULARIZATION/REVASCULARIZE WAS ABOUT 1 HOUR. IT WAS REPORTED THAT AFTER THE PRODUCT WAS DEPLOYED TO THE RIGHT M1 AND SUCCESSFUL REVASCULARIZATION WAS ACHIEVED, IT WAS DEPLOYED AT THE OCCLUSION OF THE RIGHT ICA AND THE SOLITAIREX WAS COLLECTED INTO THE PHENOM 21. HOWEVER, WHEN THE SOLITAIREX WAS ABOUT HALFWAY INSERTED INTO THE PHENOM 21, IT BECAME IMPOSSIBLE TO RETRACT IT. DURING SURGERY, IT WAS COLLECTED AS IS AND THE PROCEDURE WAS COMPLETED, BUT WHEN IT WAS PULLED INTO THE PHENOM 21 OUTSIDE THE BODY, THE S
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22035949 ↗MalfunctionMay 21, 2024Material Deformation; Physical Resistance/Sticking; Material Deformation; Physical Resistance/Sticking
MEDTRONIC RECEIVED INFORMATION THAT A PED3 PIPELINE WAS STUCK IN A PHENON 21 CATHETER. THE PATIENT WAS UNDERGOING ANEURYSM EMBOLIZATION WITH A FLOW DIVERTER PIPELINE. DURING THE PASSAGE OF THE FIRST STENT IN THE PHENOM 21 CATHETER THE PASSAGE WAS OBSERVED TO BE BLOCKED, MAKING IT IMPOSSIBLE TO CONTINUE PROGRESSION IN THE CATHETER. IT WAS COMPLETELY LOCKED IN THE CATHETER MAKING IT IMPOSSIBLE TO BE REMOVED EVEN USING EXTRA FORCE. IT WAS DECIDED TO REMOVE THE CATHETER AND STENT. THE PROCEDURE WAS RESTARTED WITH A NEW PHENOM 21 CATHETER AND STENT IF THE SAME SIZE WITH PERFECT PASSAGE WITHOUT FRICTION. THE CATHETER RESISTANCE WAS IN THE MIDDLE OF THE CATHETER. THE CATHETER WAS DAMAGED IN THE MIDDLE SECTION. THE PUSHWIRE WAS NOT DAMAGED. THE DEIVCES WERE PREPARED ACCORDING TO THE INSTRUCTIONS FOR USE (IFU). THE PATIENT WAS BEING TREATED FOR AN UNRUPTURED, SACCULAR ANUERYSM IN THE ACM LOCATION. HE MAX DIAMETER WAS 4MM AND THE NECK DIAMETER WAS 7MM. THE DISTAL LANDING ZONE WAS 2.3MM AND THE PROXIMAL LANDING ZONE WAS 2.2MM. VESSEL TORTUOSITY WAS NORMAL. THE ACCESS VESSEL DIAMETER WAS 2.5MM. NO PATIENT SYMPTOMS OR COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19364600 ↗MalfunctionMar 14, 2024Fracture; Physical Resistance/Sticking; Fracture; Physical Resistance/Sticking
MEDTRONIC RECEIVED A REPORT THAT THE SOLITAIRE X EXPERIENCED RESISTANCE AT THE PHENOM 21 MICROCATHETER HUB DURING PREPARATION. THEY TRIED SEVERAL TIMES SLOWLY, BUT THE DEVICE COULD NOT ENTER THE CATHETER. THE SOLITAIRE WAS DAMAGED IN THE PHENOM CATHETER HUB, AND THE HUB OF THE CATHETER WAS CRACKED. THE PATIENT DID NOT EXPERIENCE ANY INJURY OR COMPLICATIONS. THE DEVICES WERE PREPARED ACCORDING TO THE INSTRUCTIONS FOR USE (IFU). ADDITIONAL INFORMATION WAS RECEIVED. IT WAS CLARIFIED THAT THE DISTAL PART OF SOLITAIRE WAS DAMAGED. FOR THE CAUSE OF THE EVENT IT WAS REPORTED THAT, "WE GUESS IT'S A PHENOM PROBLEM."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18909431 ↗MalfunctionAug 31, 2023Break; Physical Resistance/Sticking; Break; Physical Resistance/Sticking
ADDITIONAL INFORMATION WAS RECEIVED THAT CATHETER DELAMINATION OCCURRED DURING IN-HOUSE SIMULATED USE BENCHTOP TESTING. CATHETER DELAMINATION WAS OBSERVED DURING IN-HOUSE BENCHTOP TESTING BY MEDTRONIC ENGINEERS. THERE WERE NO PATENTS OR HCPS INVOLVED IN THE EVENT. THERE WERE NO KINKS ON THE PUSHWIRES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17667555 ↗
FDA clearance and classification
Cleared through Special 510(k) K210230 on Feb 25, 2021 (decision: Substantially Equivalent). Submitted by Micro Therapeutics Lnc. D/B/A Ev3 Neurovascular as “Phenom Catheters”. The FDA review took 28 days. FDA record ↗
The FDA regulates product code KRA (catheter, continuous flush) under 21 CFR 870.1210 as a Class II device. 21 CFR 870.1210 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the PhenomTM 21 Catheter FG13160-0615-1S
Has the PhenomTM 21 Catheter FG13160-0615-1S been recalled?
No FDA recall record in our data names this exact device identifier (DI 00763000399696), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the PhenomTM 21 Catheter FG13160-0615-1S?
The FDA's MAUDE database lists 8 adverse event reports for this device, received between Aug 31, 2023 and Aug 31, 2026. In those reports, the most common specific device problems are “Physical Resistance/Sticking” (7), “Material Deformation” (2) and “Retraction Problem” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the PhenomTM 21 Catheter FG13160-0615-1S?
None of the reports are classified as deaths, 0 as injuries and 8 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the PhenomTM 21 Catheter FG13160-0615-1S FDA approved?
It was cleared through Special 510(k) K210230 on Feb 25, 2021. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the PhenomTM 21 Catheter FG13160-0615-1S MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the PhenomTM 21 Catheter FG13160-0615-1S, and is it still sold?
The FDA's device database lists Micro Therapeutics, Inc. as the company responsible for it (brand PhenomTM 21 Catheter, model FG13160-0615-1S). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Catheter, continuous flush devices
- PROWLER SELECT 606-S255XMedos International Sàrl · Catheter, continuous flush2 deaths973 reports
- PhenomTM 27 Catheter FG15150-0615-1SMicro Therapeutics, Inc. · Catheter, continuous flush2 deaths139 reports
- MarathonTM 105-5056Micro Therapeutics, Inc. · Catheter, continuous flush124 reports
- Renegade™ HI-FLO™ Kit M001183020Boston Scientific Corporation · Catheter, continuous flush109 reports
- MarksmanTM FA-55150-1030Micro Therapeutics, Inc. · Catheter, continuous flush109 reports
- EkoSonic® 6779-002Boston Scientific Corporation · Catheter, Continuous Flush1 recall4 deaths55 reports
- Chameleon CH08-40-75USMozarc Medical US LLC · Catheter, continuous flush2 recalls4 reports
- RebarTM 105-5081-153Micro Therapeutics, Inc. · Catheter, continuous flush2 deaths102 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page