Skip to content

TurboHawk™ Plus THP-LXRecalls & reported problems

Medtronic, Inc. · ATHERECTOMY THP-LX TH PLUS 7F EXT TIP US · Class 2

All models of TurboHawk Plus (4)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00763000402358), received between Feb 28, 2022 and Apr 17, 2026 — classified as 1 injury, 6 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (4 reports).

2 other recalls cover devices cleared under the same FDA submission (K212027); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component4
  • Adverse Event Without Identified Device or Use Problem1
  • Delivered as Unsterile Product1
  • Device Difficult to Setup or Prepare1
  • Material Deformation1
  • Physical Resistance/Sticking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • Perforation1
  • Vascular Dissection1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 17, 2026Device Difficult to Setup or Prepare; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING THE CLEANING PROCESS WITH THE ATHERECTOMY TURBOHAWK PLUS, THE DISTAL FLUSH TOOL BECAME STUCK ON THE DISTAL TIP AND COULD NOT BE ADVANCED BACK TO THE PROXIMAL POSITION. THE DEVICE WAS SAFELY REMOVED FROM THE PATIENT. NO HEALTH CONSEQUENCES OR IMPACT WERE REPORTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. WHEN THE DEVICE WAS RETURNED IT WAS NOTED THAT THE DEVICE WAS DETACHED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24911702 ↗
  • MalfunctionNov 13, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING A PROCEDURE TO TREAT PLAQUE IN THE PROXIMAL AND MID SEGMENTS OF THE LEFT SUPERFICIAL FEMORAL ARTERY IN A PATIENT, MODERATE RESISTANCE WAS ENCOUNTERED WHEN ADVANCING THE TURBOHAWK DEVICE. TIP DAMAGE AND DETACHMENT OCCURRED IN VIVO DURING THE PROCEDURE, SPECIFICALLY AT THE HINGE PIN. THE DETACHED TIP WAS REMOVED WITHIN THE SHEATH. THE DEVICE WAS SAFELY REMOVED FROM THE PATIENT, AND THE PROCEDURE WAS COMPLETED USING A NEW DEVICE OF THE SAME MAKE AND MODEL. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23543173 ↗
  • MalfunctionSep 10, 2025Detachment of Device or Device Component; Physical Resistance/Sticking; Detachment of Device or Device Component; Physical Resistance/Sticking

    IT WAS REPORTED THAT DURING TREATMENT OF A FIBROUS PLAQUE LESION IN THE LEFT MID SUPERFICIAL FEMORAL ARTERY USING THE ATHERECTOMY TH PLUS 7F EXT TIP, THERE WAS A DELAY IN SURGERY OF FIVE MINUTES DUE TO THE PRODUCT NOT PROPERLY ENGAGING THE OPPOSITE SIDE OF THE VESSEL AND THE JOG NOT WORKING PROPERLY. THE ARTERY DIAMETER WAS 6MM. A 7FR NON-MEDTRONIC SHEATH AND A SPIDER FILTER WIRE FOR EMBOLIC PROTECTION WERE USED. THE VESSEL WAS NOT PRE-DILATED BUT WAS POST-DILATED. THE DEVICE WAS SAFELY REMOVED AND THE PROCEDURE WAS COMPLETED USING A NEW HAWKONE LX ATHERECTOMY DEVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. WHEN THE DEVICE WAS RETURNED FOR EVALUATION IT WAS NOTED THAT THE CUTTER WAS DETACHED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23013766 ↗
  • MalfunctionJan 9, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    PHYSICIAN ATTEMPTED TO USE A TURBOHAWK PLUS ATHERECTOMY DEVICE FOR PATIENT TREATMENT IN THE LEFT PROX/MID SUPERFICIAL FEMORAL ARTERY WITH CALCIFICATION (CTO). NO ABNORMALITIES REPORTED IN RELATION TO ANATOMY. A NON-MEDTRONIC SHEATH (7F COOK) WAS USED. IFU WAS FOLLOWED. VESSEL WAS PREDILATED. IT IS REPORTED THAT TIP DETACHED / TIP DAMAGE OCCURRED. THERE WAS A BREAK BETWEEN THE BLADE AND THE WIRE THAT DROVE THE BLADE. THE SURGERY WAS CONTINUED BY REPLACING WITH ANOTHER DEVICE AND WITH THE COOPERATION OF SPIDER FX TO COMPLETE THE SURGERY SUCCESSFULLY. NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21111333 ↗
  • MalfunctionMay 15, 2024Delivered as Unsterile Product; Delivered as Unsterile Product

    PHYSICIAN INTENDED TO USE A TURBOHAWK PLUS LX ATHERECTOMY DEVICE DURING PATIENT TREATMENT. IT IS REPORTED THAT THE STERILE PACKAGING WAS NOT INTACT, THE GLUE WAS NOT INTACT AT CORNER OF INSIDE PACKAGING. THE ISSUE WAS NOTED PRIOR TO USE. THERE WAS NO DAMAGE NOTED TO OUTER PACKAGING. NO LABELLING ISSUES. NO DAMAGE NOTED TO THE DEVICE. A REPLACEMENT MEDTRONIC DEVICE WAS USED TO COMPLETE PROCEDURE. NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19325741 ↗
  • MalfunctionApr 5, 2024Material Deformation; Material Deformation

    PHYSICIAN WAS ATTEMPTING TO USE A TURBOHAWK PLUS 7F DIRECTIONAL ATHERECTOMY DEVICE DURING TREATMENT OF A CALCIFIED LESION IN THE PATIENTS SUPERFICIAL FEMORAL ARTERY (SFA). THERE WERE NO ABNORMALITIES REPORTED IN RELATION TO ANATOMY. EMBOLIC PROTECTION WAS NOT USED. THERE WAS NO DAMAGE NOTED TO PACKAGING, I.E. SHELF CARTON, HOOP/TRAY. THERE WERE NO ISSUES NOTED WHEN REMOVING THE DEVICE FROM THE HOOP/TRAY. THE IFU (INSTRUCTION FOR USE) WAS FOLLOWED DURING PREPARATION, PROCEDURE, POST-PROCEDURE WITH NO ISSUES NOTED. THE VESSEL WAS PRE-DILATED. THE DEVICE WAS NOT PASSED THROUGH A PREVIOUSLY DEPLOYED STENT. NO RESISTANCE WAS ENCOUNTERED WHEN ADVANCING THE DEVICE. A BULGE IN THE NOSECONE WAS REPORTED. THE TECOTHANE JACKET WAS INTACT. THE DEVICE WAS SAFELY REMOVED FROM PATIENT. A REPLACEMENT MEDTRONIC DEVICE WAS USED TO COMPLETE PROCEDURE. NO PATIENT INJURY REPORTED. IMAGE REVIEW POTENTIALLY SHOWS THAT THE TECOTHANE JACKET MAY HAVE BEEN COMPROMISED

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19048911 ↗
  • InjuryFeb 28, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PHYSICIAN WAS ATTEMPTING TO USE THE TURBOHAWK PLUS ATHERECTOMY DEVICE DURING PROCEDURE TO TREAT LESION IN THE MID SUPERFICIAL FEMORAL ARTERY (SFA). THE VESSEL HAD NO TORTUOSITY. THE DEVICE WAS INSPECTED WITH NO ISSUES NOTED. THE DEVICE WAS PREPPED PER THE IFU WAS NO ISSUES IDENTIFIED. IT WAS REPORTED THAT VESSEL DAMAGE/INJURY OCCURRED. IT WAS REPORTED THAT DISSECTION/PERFORATION OCCURRED. NO ISSUE WITH TURBOHAWK DEVICE DURING USE. UNKNOWN IF DEVICE CONTRIBUTED TO THE VESSEL DISSECTION/PERFORATION/INJURY. NO RESISTANCE ENCOUNTERED DURING ADVANCEMENT/WITHDRAWAL OF THE DEVICE. DEVICE WAS SAFELY REMOVED FROM PATIENT. A COVERED STENT WAS USED TO FIX THE ISSUE. NO FURTHER PATIENT INJURY REPORTED.

    Patient problems: Perforation; Vascular Dissection

    MAUDE report 13618746 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K212027. One submission often covers many models, so check the product and lot details in each record.

  • Z-0694-2022 · Feb 4, 2022 · Terminated
    Medtronic TurboHawk Plus Directional Atherectomy System 6F, Multi-vessel, Catalog number THP-M. For use in peripheral vasculature.
  • Z-0695-2022 · Feb 4, 2022 · Terminated
    Medtronic TurboHawk Plus Directional Atherectomy System 6F, Small Vessel, Catalog number THP-S. For use in peripheral vasculature.

Other Catheter, peripheral, atherectomy devices