Becker 46128Recalls & reported problems
Medtronic Ps Medical, Inc. · EDMS 46128 BECKER II BLUE PT LN · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00763000406073), received between Sep 9, 2024 and Apr 20, 2026 — classified as 10 malfunctions.
The most frequently reported specific device problem is “Break” (5 reports).
1 FDA recall record name this device's identifier, 1 still open.
6 other recalls cover devices cleared under the same FDA submission (K984053); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Nov 1, 2024Open, Classified
Due to complaints received reporting cracks and/or leaks of the stopcocks associated with the external drainage and monitoring system (EDMS).
Z-0661-2025 · Root cause: Under Investigation by firm · FDA record ↗
Most reported problems
Device problems
- Break5
- Fluid/Blood Leak5
- Fracture5
- Disconnection1
- Mechanical Jam1
Patient problems
- No Clinical Signs, Symptoms or Conditions10
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 20, 2026Break; Fracture
MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THAT THE PRODUCT RUPTURED. NO PATIENT PRESENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24931962 ↗MalfunctionAug 25, 2025Break; Fracture; Break; Fracture
MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THAT 3 DEVICES IN A ROW FAILED IN THE OPERATING ROOM AS THE CONNECTOR WAS CRACKED. A 4TH DEVICE WAS OPENED AND SUCCESSFULLY USED. NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22882646 ↗MalfunctionAug 25, 2025Break; Fracture; Break; Fracture
MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THAT 3 DEVICES IN A ROW FAILED IN THE OPERATING ROOM AS THE CONNECTOR WAS CRACKED. A 4TH DEVICE WAS OPENED AND SUCCESSFULLY USED. NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22882640 ↗MalfunctionFeb 19, 2025Break; Fracture; Break; Fracture
MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THAT THE LUMBAR DRAIN WAS FOUND SNAPPED OFF AT THE TUBING PROXIMAL TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21410439 ↗MalfunctionJan 14, 2025Fluid/Blood Leak; Fluid/Blood Leak
ADDITIONAL INFORMATION RECEIVED REPORTED THAT THE DRAIN WAS SWITCHED OUT DUE TO SAFETY RISK INCLUDING RISK OF HERNIATION FROM EXCESSIVE CSF LEAK AND INFECTION. THE PATIENT'S CURRENT STATUS WAS DECEASED, HOWEVER, THE FAMILY MADE THE DECISION TO WITHDRAW LIFE-SUSTAINING MEASURES PER THE PATIENT'S ADVANCED DIRECTIVE, AND THEIR COMPLEX DIAGNOSES (ANEURYSM RUPTURE, MULTI ORGAN FAILURE) WERE NOT RELATED TO THE MEDICAL DEVICE AND/OR THERAPY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21147547 ↗MalfunctionOct 18, 2024Break; Fracture; Mechanical Jam; Break; Fracture; Mechanical Jam
MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THAT THE COLLECTION BAG ON THE LUMBAR DRAIN WAS DUE FOR A CHANGE. THE LUER LOCK WHERE THE COLLECTION BAG CONNECTED TO THE DRAIN TUBING WAS STUCK, AND THE BAG SNAPPED OFF THE TUBING WITH THE LUER LOCK STILL STUCK. THE SITE WAS UNABLE TO DETACH THE LUER LOCK FROM THE REST OF THE SYSTEM. THE DEVICE WAS REMOVED. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20480497 ↗MalfunctionSep 10, 2024Fluid/Blood Leak; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THERE WAS WATER LEAKAGE JUST BELOW THE DRIP CHAMBER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20181485 ↗MalfunctionSep 10, 2024Fluid/Blood Leak; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THERE WAS WATER LEAKAGE BETWEEN THE DRIP CHAMBER AND DRAINAGE BAG. NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20181037 ↗MalfunctionSep 10, 2024Fluid/Blood Leak; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THERE WAS WATER LEAKAGE. NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20180913 ↗MalfunctionSep 9, 2024Disconnection; Fluid/Blood Leak; Disconnection; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THAT THERE WERE 2 CLAMPS BETWEEN THE FLOW CHAMBER AND THE BAG. TOP OF THE CLAMP CAME OUT AND DISCONNECTED. THE NURSE CLEANED THE BOTTOM OF THE TUBE AND RECONNECTED AND ESCALATED TO THE NEUROSURGERY RESIDENT ON CALL. THE RESIDENT ON CALL STATED IT WAS OKAY NOW BUT TO MENTION TO THE DOCTOR ON ROUNDS. THEN, LEAKING WAS NOTICED AT THE DISTAL ACCESS SITE BY THE DAY SHIFT NURSE BETWEEN THE BURETROL AND BAD. THE NEUROSURGERY TEAM WAS NOTIFIED, AND THE SYSTEM WAS CHANGED. THERE WAS A PROCEDURE DELAY OF 30 MINUTES. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20168770 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K984053. One submission often covers many models, so check the product and lot details in each record.
- Z-2412-2023 · Jun 28, 2023 · Open, ClassifiedMedtronic REF 46915 Duet External Drainage and Monitoring system, Interlink Injection sites, Ventricular Catheter, STERILE EO, Rx Only
- Z-1924-2014 · Jun 9, 2014 · TerminatedMedtronic Duet External Drainage and Monitoring System, Interlink Injection Sites, Lumbar Catheter, Catalog No. 46917. Intended to allow drainage and measurement of cerebrospinal
- Z-1921-2014 · Jun 9, 2014 · TerminatedMedtronic Duet External Drainage and Monitoring System, SmartSite Injection Sites, Catalog No. 46914. Intended to allow drainage and measurement of cerebrospinal fluid (CSF) and o
- Z-1922-2014 · Jun 9, 2014 · TerminatedMedtronic Duet External Drainage and Monitoring System, Interlink Injection Sites, Ventricular Catheter, Catalog No. 46915. Intended to allow drainage and measurement of cerebrosp
- Z-1923-2014 · Jun 9, 2014 · TerminatedMedtronic Duet External Drainage and Monitoring System, SmartSite Injection Sites, Ventricular Catheter, Catalog No. 46916. Intended to allow drainage and measurement of cerebrosp
- Z-1920-2014 · Jun 9, 2014 · TerminatedMedtronic Duet External Drainage and Monitoring System, Interlink Injection Sites, Catalog No. 46913. Intended to allow drainage and measurement of cerebrospinal fluid (CSF) and o
Other Shunt, central nervous system and components devices
- NATUS EDS3 DRAINAGE SYSTEM NT821731CNatus Medical Incorporated · Shunt, Central Nervous System And Components1 recall208 reports
- Exacta™ 46700Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall2 deaths163 reports
- Strata® 42866Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 death160 reports
- proGAV 2.0® FX410TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components119 reports
- PROGAV 2.0 FX643TAesculap AG · Shunt, central nervous system and components100 reports
- Becker® 46128Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall50 reports
- M.BLUE PLUS FX804TAesculap AG · Shunt, central nervous system and components82 reports
- proSA® FV701TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components82 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.