StrataMR 42965MRRecalls & reported problems
Medtronic Ps Medical, Inc. · VALVE 42965MR STRATAMR II REGULAR · Class 2 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00763000422714), received between Sep 10, 2024 and Aug 12, 2025 — classified as 2 injuries.
The most frequently reported specific device problem is “Infusion or Flow Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Infusion or Flow Problem1
- Mechanical Jam1
- Mechanics Altered1
Patient problems
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryAug 12, 2025Mechanical Jam; Mechanics Altered; Mechanical Jam; Mechanics Altered
MEDTRONIC RECEIVED INFORMATION REGARDING A MAGNETIC RESISTANT ADJUSTABLE VALVE. IT WAS REPORTED THE VALVE CHANGED FROM 2.0 TO 2.5 AFTER GOING THROUGH AN MRI, BUT THEN WHEN THE CUSTOMER WENT TO ADJUST THE SETTING, THEY WERE ONLY ABLE TO DO SO ON THE 11TH ATTEMPT, AND IT COULD ONLY GO FROM 2.5 TO 2.0 AND THE MAGNET HAD TO SIT DIRECTLY OVER THE VALVE. THE DEVICE WAS REPLACED WITH A NON MAGNETIC ADJUSTABLE VALVE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22790535 ↗InjurySep 10, 2024Infusion or Flow Problem; Infusion or Flow Problem
MEDTRONIC RECEIVED INFORMATION REGARDING AN ADJUSTABLE VALVE. IT WAS REPORTED THE PATIENT'S VALVE WAS SET TO 0.5, BUT THEY WERE NOT DRAINING ENOUGH/INSUFFICIENTLY. THE PATIENT CAME INTO THE HOSPITAL, AND CT SHOWED ACUTE HYDROCEPHALUS. THE VENTRICLES WERE REFILLED TO A HIGH PRESSURE. THE SITE DID NOT OBSERVE ANY FLAWS IN THE CATHETERS WHEN THEY EXPLAINED THEM, AND THE RESERVOIR ON THE VALVE WAS SOFT ENOUGH TO PUSH DOWN MEANING IT WAS NOT OBSTRUCTED WHEN THEY PALPATED IT BEFORE THE VALVE WAS EXPLANTED. IT WAS NOTED THAT THE MANUFACTURER REPRESENTATIVE WAS THERE WHEN THEY IMPLANTED IT, AND IT WAS UNDER AXIEM NAVIGATION. THEY HAD IMMEDIATE CEREBROSPINAL FLUID FLOW AFTER PLACED THE VENTRICULAR CATHETER. THE SITE DID NOT DO A CHEST ABDOMEN CT TO SEE IF THE PERITONEAL CATHETER WAS KINKED. WHEN EXPLANTING, THE SURGEON SAID THEY FLUSHED SALINE INTO THE PERITONEAL CATHETER WITHOUT RESISTANCE. AN EXTERNAL VENTRICULAR DRAIN WAS PLACED WHICH DRAINED THE PATIENT ADEQUATELY WITH AN INTRACRANIAL PRESSURE RANGING FROM 2-10. THE VALVE SETTING WAS CONFIRMED WITH THE REPORTED DEVICE HAND TOOLS OF 0.5. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY.
Patient problems: Insufficient Information
MAUDE report 20184383 ↗
Other Shunt, central nervous system and components devices
- NATUS EDS3 DRAINAGE SYSTEM NT821731CNatus Medical Incorporated · Shunt, Central Nervous System And Components1 recall208 reports
- Exacta™ 46700Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall2 deaths163 reports
- Strata® 42866Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 death160 reports
- proGAV 2.0® FX410TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components119 reports
- Becker® 46128Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall50 reports
- PROGAV 2.0 FX643TAesculap AG · Shunt, central nervous system and components100 reports
- proSA® FV701TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components82 reports
- M.BLUE PLUS FX804TAesculap AG · Shunt, central nervous system and components82 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page