MIDAS REX STEALTH MIDAS MR8 EM800NRecalls & reported problems
Medtronic Powered Surgical Solutions · MOTOR EM800N MR8 ELECTRIC NAVIGATED S8 · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00763000430924), received between Aug 20, 2024 and Sep 4, 2025 — classified as 1 injury, 9 malfunctions.
The most frequently reported specific device problem is “Overheating of Device” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Overheating of Device7
- Smoking3
- Corroded2
- Communication or Transmission Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions9
- Chemical Exposure1
- Hemorrhage/Blood Loss/Bleeding1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 4, 2025Corroded; Corroded
REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF CORROSION. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT. REPAIR REQUEST ESCALATED TO PRODUCT EVENT DUE TO CUSTOMER INDICATION THAT THIS REPORT IS A COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22960853 ↗MalfunctionAug 13, 2025Corroded; Corroded
REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF CORROSION.IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT. REPAIR REQUEST ESCALATED TO PRODUCT EVENT DUE TO CUSTOMER INDICATION THAT THIS REPORT IS A COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22793514 ↗MalfunctionJun 10, 2025Overheating of Device; Smoking; Overheating of Device; Smoking
IT WAS REPORTED THAT DURING L3-L5 POSTERIOR DECOMPRESSION AND FUSION PROCEDURE, DRILL WAS GLOWING, COULD SMELL SMOKE AND VISIBLY SEE SMOKE COMING FROM DRILL. IT WAS REPORTED THAT THERE WAS NO INJURY TO THE PATIENT DUE TO THE EVENT AND THE PROCEDURE WAS EXTENDED FOR FIVE MINUTES. THE DRILL WAS WRAPPED IN WET LAPS AND REMOVED FROM THE ROOM AND THE PROCEDURE WAS COMPLETED WITH BACKUP PRODUCTS. ADDTIONAL INFORMATION RECEIVED FROM FACILITY STATED THAT THE REPLACEMENT DRILL ALSO EXHIBITING OVERHEATING ISSUES AND REQUIRED MULTIPLE BURRS TO COMPLETE THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22174198 ↗MalfunctionJun 10, 2025Overheating of Device; Smoking; Overheating of Device; Smoking
IT WAS REPORTED THAT DURING L3-L5 POSTERIOR DECOMPRESSION AND FUSION PROCEDURE, DRILL WAS GLOWING, COULD SMELL SMOKE AND VISIBLY SEE SMOKE COMING FROM DRILL. IT WAS REPORTED THAT THERE WAS NO INJURY TO THE PATIENT DUE TO THE EVENT AND THE PROCEDURE WAS EXTENDED FOR FIVE MINUTES. THE DRILL WAS WRAPPED IN WET LAPS AND REMOVED FROM THE ROOM AND THE PROCEDURE WAS COMPLETED WITH BACKUP PRODUCTS. ADDTIONAL INFORMATION RECEIVED FROM FACILITY STATED THAT THE REPLACEMENT DRILL ALSO EXHIBITING OVERHEATING ISSUES AND REQUIRED MULTIPLE BURRS TO COMPLETE THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22174175 ↗MalfunctionMar 18, 2025Overheating of Device; Overheating of Device
REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF OVERHEATING. IT WAS REPORTED THAT THERE WAS NO PATIENT OR STAFF IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21633535 ↗MalfunctionFeb 21, 2025Overheating of Device; Smoking; Overheating of Device; Smoking
IT WAS REPORTED THAT DEVICE WITH THE REPORT OF OVERHEATING AND SMOKE WHEN USED IN A CASE. IT WAS REPORTED THAT THERE WAS NO PATIENT OR STAFF IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21435233 ↗MalfunctionFeb 18, 2025Overheating of Device; Overheating of Device
REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF OVERHEATING. IT WAS REPORTED THAT THERE WAS NO PATIENT OR STAFF IMPACT. REPAIR REQUEST ESCALATED TO PRODUCT EVENT DUE TO CUSTOMER INDICATION THAT THIS REPORT IS A COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21401432 ↗InjuryJan 3, 2025Communication or Transmission Problem; Communication or Transmission Problem
IT WAS REPORTED THAT DURING CRANIAL RESECTION PROCEDURE THE EQUIPMENT DOES NOT SHOW THE CORRECT ANATOMICAL POINT. IT WAS REPORTED TH AT BLEEDING OCCURS WHEN USING THE DEVICE. THERE WAS A DELAY OF ONE HOUR OR LONGER AND THE PATIENT WAS UNDER ANASTHESIA DURING THE PROCEDURE. THIS LED TO EXTENDED HOSPITALIZATION FOR THE PATIENT AND THE PATIENT WAS TAKEN INTO THE INTENSIVE CARE UNIT (ICU). ADDITIONAL INFORMATION REGARDING THE PATIENT IMPACT IS BEING FOLLOWED UP FROM THE FACILITY.
Patient problems: Chemical Exposure; Hemorrhage/Blood Loss/Bleeding
MAUDE report 21068868 ↗MalfunctionDec 29, 2024Overheating of Device; Overheating of Device
REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF OVERHEATING. NO PATIENT IMPACT REPORTED. REPAIR REQUEST ESCALATED TO A PRODUCT EVENT DUE REASON FOR RETURN AND RETURN IN LESS THAN 90 DAYS FROM PURCHASE OR REPAIR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21032163 ↗MalfunctionAug 20, 2024Overheating of Device; Overheating of Device
IT WAS REPORTED THAT THE HANDPIECE WAS OVERHEATING. IT WAS ALSO REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20005111 ↗
Other Neurological stereotaxic Instrument devices
- StealthStation S8 Platform 9735665Medtronic Navigation, Inc. · Neurological stereotaxic Instrument4 deaths4,668 reports
- Stealthstation® S7™ 9733856Medtronic Navigation, Inc. · Neurological stereotaxic Instrument6 deaths3,908 reports
- StealthStation S8 Platform 9735665Medtronic Navigation, Inc. · Neurological stereotaxic Instrument1 death1,896 reports
- StealthStation S8 9735669Medtronic Navigation, Inc. · Neurological stereotaxic Instrument1,831 reports
- StealthStation S8 Platform 9735665Medtronic Navigation, Inc. · Neurological stereotaxic Instrument2 deaths1,542 reports
- Fusion™ ENT Image Guidance System 9733560XOMMedtronic Navigation, Inc. · Neurological stereotaxic Instrument1,316 reports
- Stealthstation® S7™ 9733858Medtronic Navigation, Inc. · Neurological stereotaxic Instrument8 deaths984 reports
- Stealthstation® S7™ 9733858Medtronic Navigation, Inc. · Neurological stereotaxic Instrument1 death695 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page