Apollo 105-5096-000Recalls & reported problems
Micro Therapeutics, Inc. · MICRO CATH 105-5096-000 V08 APOLLO 3.0CM · Class 3
All models of Apollo (2)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Micro Therapeutics Apollo microcatheter has 77 adverse event reports, with no deaths and 3 injuries. Reports mostly describe catheter parts breaking, detaching, leaking, blocking, or resisting during embolization procedures, sometimes leaving foreign material in patients. There are no recalls naming this device. Reports fell from 34 in the prior 12 months to 14 in the last 12 months.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 79
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 79 adverse event reports that list this exact device (DI 00763000446758), received between May 8, 2023 and Sep 3, 2026 — classified as 3 injuries, 76 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (39 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component39
- Physical Resistance/Sticking19
- Burst Container or Vessel15
- Leak/Splash12
- Packaging Problem12
- Break10
- Material Deformation10
- Difficult to Advance6
Patient problems
- No Clinical Signs, Symptoms or Conditions62
- Foreign Body In Patient24
- Vasoconstriction3
- Intracranial Pressure Increased1
- Paralysis1
Recent adverse event reports
The 25 most recent of 79 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 3, 2026Detachment of Device or Device Component
Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions
MAUDE report 50590811 ↗MalfunctionAug 12, 2026Difficult to Remove; Separation Failure; Physical Resistance/Sticking
Patient problems: Vasoconstriction
MAUDE report 50221258 ↗MalfunctionJun 17, 2026Material Deformation; Physical Resistance/Sticking
MEDTRONIC RECEIVED A REPORT THAT AN APOLLO CATHETER ENCOUNTERED RESISTANCE AT THE DISTAL SECTION. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF A ARTERIOVENOUS MALFORMATION FLUID EMBOLIZATION. IT WAS NOTED THE PATIENT'S ACCESS VESSEL WAS THE FEMORAL ARTERY AND THAT VESSEL TORTUOSITY WAS MODERATE. DURING LIQUID EMBOLIZATION TREATMENT FOR AN ARTERIOVENOUS MALFORMATION, AFTER ADVANCING THE MICROCATHETER TO THE DESIGNATED SITE DURING THE PROCEDURE, SIGNIFICANT RESISTANCE WAS ENCOUNTERED UPON INJECTING THE EMBOLIC AGENT, WITH MINIMAL FLOW OBSERVED FROM THE TIP. INSPECTION OF THE MICROCATHETER AND ITS ASSOCIATED SYSTEM REVEALED A KINK APPROXIMATELY 5 CM FROM THE DISTAL END OF THE MICROCATHETER; CONSEQUENTLY, THE MICROCATHETER WAS REPLACED TO SUCCESSFULLY COMPLETE THE PROCEDURE. THE KINK WAS OBSERVED AT THE DISTAL SECTION OF THE CATHETER. THE CATHETER WAS FLUSHED AND ALL DEVICES WERE PREPPED AS INDICATED IN THE IFU. NO PATIENT COMPLICATIONS WERE ASSOCIATED WITH THIS EVENT. ADDITIONAL INFORMATION RECEIVED REPORTED THAT THE CAUSE OF THE RESISTANCE WAS UNKNOWN. ADDITIONAL INFORMATION RECEIVED. FACILITY AND PHYSICIAN¿S CONTACT INFORMATION WERE PROVIDED. ADDITIONAL INFORMATION WAS RECEIV
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25536391 ↗MalfunctionApr 16, 2026Compatibility Problem
IT WAS REPORTED THAT DURING A PROCEDURE TO TREAT AN ARTERIOVENOUS MALFORMATION USING THE APOLLO CATHETER, THE CATHETER BECAME BLOCKED AFTER FOLLOWING THE STANDARD PREPARATION PROCESS, WHICH INCLUDED FLUSHING WITH DMSO SOLVENT AND INJECTING THE EMBOLIC AGENT. THE BLOCKAGE PREVENTED FURTHER INJECTION OF THE EMBOLIC AGENT. THE SAME ISSUE OCCURRED WHEN A SECOND APOLLO CATHETER. THE CATHETER WAS THEN REPLACED WITH A MARATHON FLOATING MICROCATHETER, ALLOWING THE PROCEDURE TO PROCEED NORMALLY. THE ACCESS VESSEL WAS THE FEMORAL ARTERY WITH A DIAMETER OF 9.5MM AND NORMAL VESSEL TORTUOSITY. THE CATHETER AND ACCESSORY DEVICES WERE PREPARED AND FLUSHED ACCORDING TO THE INSTRUCTIONS FOR USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT THE CAUSE OF THE BLOCKAGE WAS NOT DETERMINED. THE DEAD SPACE OF THE DELIVERY CATHETER WAS FILLED WITH DMSO. THERE WAS NO RESISTANCE DURING FLUSHING OR DMSO FILLING, AND NO KINK OR DAMAGE WAS OBSERVED ON THE CATHETER. ONYX REFLUX WAS PRESENT, BUT THE VOLUME OF REFLUX IS UNKNOWN. THE PHYSICIAN PAUSED DURING INJECTION, WITH WAITING TIMES BETWEEN INJECTIONS RANGING FROM 30 SECONDS TO 2 MINUTES
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24896652 ↗MalfunctionApr 7, 2026Burst Container or Vessel; Leak/Splash; Packaging Problem
MEDTRONIC RECEIVED A REPORT THAT DURING ARTERIOVENOUS FISTULA SURGERY, AFTER THE INTERMEDIATE CATHETER WAS POSITIONED, THE MICROCATHETER WAS REMOVED AND FLUSHED. LEAKAGE WAS OBSERVED DUE TO RUPTURE OF THE MICROCATHETER. THE MICROCATHETER WAS REPLACED, AND THE PROCEDURE CONTINUED. THERE WAS CATHETER LEAK IN THE DISTAL SECTION. THE CATHETER WAS INSPECTED OR TESTED PRIOR TO USE. THE LEAK WAS OBSERVED DURING PREPARATION. THERE WERE NO PATIENT SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THIS EVENT. THE DEVICE AND ANY ACCESSORIES WERE PREPARED AS INDICATED IN THE INSTRUCTIONS FOR USE (IFU). THE CATHETER WAS FLUSHED AS INDICATED IN THE IFU.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24800178 ↗MalfunctionApr 3, 2026Device Damaged by Another Device; Physical Resistance/Sticking
MEDTRONIC RECEIVED A REPORT THAT THERE WAS RESISTANCE DURING ADVANCEMENT AND THE MICRO-GUIDEWIRE PROTRUDED FROM THE SIDE OF THE MICROCATHETER TIP WALL. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF AN ARTERIOVENOUS MALFORMATION (AVM) EMBOLIZATION. IT WAS NOTED THE PATIENT'S VESSEL TORTUOSITY WAS MODERATE. THE ACCESS VESSEL WAS THE FEMORAL ARTERY WITH A DIAMETER OF 12MM. IT WAS REPORTED THAT FOR AVM EMBOLIZATION, THE OPERATOR PLANNED TO USE AN APOLLO MICROCATHETER FOR GEL EMBOLIZATION. THE GUIDEWIRE WAS INSERTED INTO THE MICROCATHETER AND ADVANCED INTO THE MIDDLE CATHETER. RESISTANCE WAS ENCOUNTERED DURING ADVANCEMENT, AND UPON REMOVAL, THE MICRO-GUIDEWIRE WAS FOUND TO HAVE PROTRUDED FROM THE SIDE OF THE MICROCATHETER TIP WALL. THE DEVICE AND ANY ACCESSORIES WERE PREPARED AS INDICATED IN THE IFU. THE CATHETER WAS FLUSHED AS INDICATED IN THE IFU. NO PATIENT SYMPTOMS OR FURTHER COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT. THE CATHETER WAS CRUSHED IN THE DISTAL SECTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24779742 ↗MalfunctionMar 18, 2026Detachment of Device or Device Component
MEDTRONIC RECEIVED A REPORT OF PREMATURE DETACHMENT OF AN APOLLO CATHETER TIP. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF AN ARTERIOVENOUS MALFORMATION (AVM). IT WAS NOTED THE PATIENT'S VESSEL TORTUOSITY WAS MODERATE. THE ACCESS VESSEL WAS THE RADIAL ARTERY WITH A 3 MM VESSEL DIAMETER. IT WAS REPORTED THAT DURING REMOVAL OF THE MICROCATHETER, THE OPERATOR NOTED THAT THE DISTAL TIP OF THE MICROCATHETER DETACHED SPONTANEOUSLY. IT WAS NOTED THAT THERE WAS TIP PREMATURE DETACHMENT. NO FRICTION OR DIFFICULTY WAS OBSERVED DURING INJECTION. THERE WAS NO FORCE APPLIED DURING REMOVAL. THE CATHETER TIP WAS NOT ENTRAPPED OR STUCK. THERE WAS NO VASOSPASM. THERE WAS NO SURGICAL OR MEDICINAL INTERVENTION REQUIRED. IT WAS NOT SPECIFIED IF THIS OCCURRED DURING ONYX INJECTION. IT WAS INDICATED THAT ALL DEVICES WERE PREPARED AS PER THE INSTRUCTIONS FOR USE (IFU), AND NO PATIENT SYMPTOMS OR FURTHER COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT. THE CATHETER WAS FLUSHED AS INDICATED IN THE IFU.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24632203 ↗MalfunctionMar 2, 2026Burst Container or Vessel
MEDTRONIC RECEIVED A REPORT THAT AN APOLLO MICROCATHETER RUPTURED DURING A SURGICAL PROCEDURE. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF A TAMPONADE MALFORMATION ANEURYSM, BY DELIVERY OF LIQUID EMBOLIZATION SYSTEM. IT WAS REPORTED THAT DURING SURGERY, THE DOCTOR USED A MICROCATHETER TO DELIVER A LIQUID EMBOLIZATION SYSTEM TO FILL AN ABNORMAL ANEURYSM. DURING THE PROCEDURE, THE MICROCATHETER RUPTURED. THE DOCTOR IMMEDIATELY MONITORED THE PATIENT'S VITAL SIGNS, WHICH SHOWED NO ABNORMAL FLUCTUATIONS. DSA EXAMINATION OF THE CEREBRAL VESSELS REVEALED LOCALIZED LEAKAGE OF EMBOLIC MATERIAL, BUT NO DAMAGE TO OTHER VESSELS. POSTOPERATIVE EXAMINATION SHOWED STABLE VITAL SIGNS AND NO BODILY DAMAGE. AFTER REMOVAL OF THE ENDOTRACHEAL TUBE AND 30 MINUTES OF OBSERVATION WITHOUT ANY SPECIAL ISSUES, THE PATIENT WAS TRANSFERRED TO THE PEDIATRIC ICU. NO PATIENT COMPLICATIONS WERE ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24483546 ↗MalfunctionFeb 3, 2026Burst Container or Vessel
MEDTRONIC RECEIVED A REPORT THAT DURING ONYX LIQUID EMBOLIC INJECTION, THE DISTAL END OF THE APOLLO MICROCATHETER RUPTURED. NO PATIENT SYMPTOMS OR COMPLICATIONS WERE ASSOCIATED WITH THIS EVENT. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF AN INTRACRANIAL CEREBROVASCULAR MALFORMATION. IT WAS NOTED THE PATIENT'S VESSEL TORTUOSITY WAS NORMAL. RIGHT FEMORAL ARTERY ACCESS VESSEL DIAMETER WAS 6.0 MM. THE REPORTED DEVICE AND ANY ACCESSORY DEVICES WERE PREPARED, AND THE CATHETER WAS FLUSHED, AS INDICATED IN THE INSTRUCTIONS FOR USE (IFU). THERE WAS NO FRICTION OR DIFFICULTY DURING DELIVERY. IT WAS UNKNOWN IF THERE WAS ANY OTHER DAMAGE TO THE CATHETER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24248095 ↗MalfunctionJan 26, 2026Detachment of Device or Device Component; Packaging Problem
ADDITIONAL INFORMATION RECEIVED REPORTED THAT THE DAMAGED APOLLO CATHETER WAS NOT FLUSHED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24184719 ↗MalfunctionJan 26, 2026Break; Detachment of Device or Device Component; Physical Resistance/Sticking
ON (B)(6) 2026 MPXR (B)(4) (FOR, HCP, REP): MEDTRONIC RECEIVED A REPORT THAT THE APOLLO CATHETER PREMATURELY DETACHED. THE PATIENT WAS UNDERGOING SURGERY FOR ARTERIOVENOUS MALFORMATION TREATMENT. THE ACCESS VESSEL WAS THE RIGHT FEMORAL ARTERY. IT WAS NOTED THE PATIENT'S VESSEL TORTUOSITY WAS NORMAL. IT WAS REPORTED THAT A LIQUID EMBOLIZATION PROCEDURE FOR A CAVERNOUS SINUS ARTERIOVENOUS FISTULA WAS PERFORMED USING AN APOLLO CATHETER. PREOPERATIVE PREPARATION AND OPERATION WERE CARRIED OUT ACCORDING TO THE INSTRUCTIONS, AND ONYX WAS DELIVERED. DURING THE ESTABLISHMENT OF THE SURGICAL PATHWAY, A PREMATURE DETACHMENT EVENT OCCURRED AT THE TIP END OF THE APOLLO MICROCATHETER. THE TIP END WAS DETACHED SPONTANEOUSLY BEFORE REACHING THE FISTULA SITE. THE SURGEON ATTEMPTED TO USE A BALLOON CATHETER TO RETRIEVE THE DETACHED PORTION BUT WAS UNSUCCESSFUL. THE SL-10 MICROCATHETER WAS THEN USED AS A REPLACEMENT, POSITIONED AGAINST THE DETACHED APOLLO, TO DELIVER THE ONYX GLUE. THE DELIVERY WAS SUCCESSFUL, AND THE SURGERY WAS COMPLETED. THERE WERE NO PATIENT SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THIS EVENT. THE DEVICE AND ANY ACCESSORIES WERE PREPARED AS INDICATED IN THE INSTRUCTIONS FOR USE
Patient problems: Foreign Body In Patient
MAUDE report 24182907 ↗MalfunctionDec 11, 2025Material Deformation; Material Deformation
MEDTRONIC RECEIVED A REPORT THAT THE APOLLO CATHETER WAS DAMAGED DURING HYDRATION. IT WAS REPORTED THAT DURING THE PROCEDURE, THE ASSISTANT TOOK OUT THE APOLLO CATHETER FROM THE PACKAGING AS USUAL AND PERFORMED ROUTINE HYDRATION. DURING HYDRATION, IT WAS FOUND THAT THERE WAS NO DETACHMENT ZONE AT THE TIP END, WHICH LED TO A COMPLAINT BEING RAISED. THE DEVICE WAS THEN REPLACED WITH ANOTHER APOLLO, WHICH DID NOT EXHIBIT ANY ABNORMALITIES. THE PROCEDURE WAS COMPLETED SUCCESSFULLY, AND THE PATIENT EXPERIENCED NO ADVERSE REACTIONS POSTOPERATIVELY. THE REPORTED DEVICE AND ANY ACCESSORY DEVICES WERE PREPARED AS INDICATED IN THE INSTRUCTIONS FOR USE (IFU). THE CATHETER WAS FLUSHED PER THE IFU. NO PATIENT SYMPTOMS OR COMPLICATIONS WERE ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23782834 ↗InjuryDec 3, 2025Break; Break
ADDITIONAL INFORMATION RECEIVED REPORTED THAT THE PATIENT WAS BEING TREATED FOR A LEFT FRONTAL LOBE MALFORMATION, CEREBRAL HEMORRHAGE, 2CM IN SIZE. THERE WAS MODERATE TO SEVERE VESSEL TORTUOSITY. THE PATIENT HAD A BRAIN HERNIATION AND RIGHT-SIDED HEMIPLEGIA. THE TIP PORTION (25CM) REMAINED INSIDE THE SKULL, WHILE THE REST WAS REMOVED. THE ANEURYSM WAS NOT TREATED. NO FURTHER INTERVENTIONS OR TREATMENT WERE PERFORMED. THE CUSTOMER INDICATED THAT THE DSA EQUIPMENT IMAGING WAS NORMAL.
Patient problems: Foreign Body In Patient; Intracranial Pressure Increased; Paralysis
MAUDE report 23705238 ↗MalfunctionNov 17, 2025Break; Leak/Splash; Break; Leak/Splash
MEDTRONIC RECEIVED A REPORT THAT THE PATIENT WAS UNDERGOING TREATMENT FOR LEFT VERTEBRAL ARTERY VENOUS MALFORMATION EMBOLIZATION. IT WAS NOTED THAT THE PATIENT'S VESSEL TORTUOSITY WAS MINIMAL. THE ACCESS VESSEL WAS THE RIGHT FEMORAL ARTERY, WITH 6.3MM DIAMETER. IT WAS REPORTED THAT AFTER THE APOLLO MICROCATHETER WAS POSITIONED, GLUE INJECTION WAS INITIATED. IT WAS OBSERVED THAT THE GLUE WAS LEAKING FROM THE PROXIMAL END OF THE MICROCATHETER TIP. UPON WITHDRAWING THE MICROCATHETER, A RUPTURE WAS FOUND IN THE DISTAL SHAFT OF THE MICROCATHETER. THERE WAS NO FRICTION OR DIFFICULTY DURING DELIVERY. ASIDE FROM RUPTURE, THERE NO OTHER DAMAGE TO THE CATHETER. THE INJECTION RATE WAS 0.25ML/MIN. NO THE REPORTED DEVICE AND ANY ACCESSORY DEVICES WERE PREPARED AND FLUSHED AS INDICATED IN THE INSTRUCTIONS FOR USE (IFU). NO PATIENT SYMPTOMS OR COMPLICATIONS WERE ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23576644 ↗MalfunctionNov 13, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
ADDITIONAL INFORMATION WAS RECEIVED. THE PATIENT WAS TREATED FOR EMBOLIZATION OF THE EXTERNAL CAROTID ARTERY. NO ACTIONS OR INTERVENTIONS WERE TAKEN TO RESOLVE THE TIP PREMATURE DETACHMENT, AS THE DETACHED TIP REMAINS IN THE PATIENT AND WILL NOT BE RETURNED FOR ANALYSIS. THE CAUSE OF THE PREMATURE DETACHMENT WAS NOT DETERMINED, AND IT OCCURRED DURING DELIVERY WITHIN THE VESSEL. THE DETACHABLE SEGMENT AT THE TIP OF THE MICROCATHETER DETACHED WITHIN THE VESSEL, SPECIFICALLY IN THE EXTRACRANIAL SEGMENT OF THE CAROTID ARTERY. THERE WAS NO RESISTANCE WHEN ADVANCING OR RETRIEVING THE APOLLO, AND THE TIP IS NOT EMBEDDED IN THE ONYX CAST. THERE ARE NO FUTURE PLANS TO RETRIEVE THE CATHETER TIP. THE SUBJECT IS CURRENTLY RECOVERING WITHOUT ABNORMALITIES, AND NO VESSEL CALCIFICATION WAS NOTED. THE MODEL AND LOT NUMBER OF THE GUIDE CATHETER ARE UNKNOWN, WHILE THE GUIDE WIRE USED WAS AN ASAHI-10, WHICH WAS HYDRATED AS INDICATED IN THE IFU. NO KINK OR DAMAGE WAS NOTICED ON ANY OF THE DEVICES, AND A CONTINUOUS FLUSH WAS MAINTAINED THROUGH THE GUIDE CATHETER DURING THE PROCEDURE.
Patient problems: Foreign Body In Patient
MAUDE report 23543277 ↗MalfunctionNov 5, 2025Break; Packaging Problem; Break; Packaging Problem
ADDITIONAL INFORMATION WAS RECEIVED STATING THAT THE LESION WAS AN ARTERIOVENOUS FISTULA. THE CAUSE OF THE DAMAGE WAS NOT DETERMINED. THE DEVICE PACKAGING WAS NOT DAMAGED ON DELIVERY. THERE WERE NO ISSUES THAT RESULTED IN THE DAMAGE THAT WAS FOUND.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23479878 ↗MalfunctionOct 27, 2025Material Deformation; Physical Resistance/Sticking; Material Deformation; Physical Resistance/Sticking
MEDTRONIC RECEIVED A REPORT THAT THE PATIENT WAS DIAGNOSED WITH AN INTRACRANIAL ARTERIOVENOUS MALFORMATION (AVM). THE SURGEON USED A .010 MICRO GUIDEWIRE TO SUPER-SELECT THE DISTAL MALFORMED VESSELS, SUCCESSFULLY ADVANCING THE GUIDEWIRE DISTALLY. HOWEVER, THE APOLLO MICROCATHETER COULD NOT TRACK AND BE PUSHED TO GO UPPER DESPITE MULTIPLE ATTEMPTS. THE SURGEON INDICATED THAT THERE MIGHT BE AN ISSUE WITH THE INNER DIAMETER OF THE MICROCATHETER. AFTER REPLACING TO A MARATHON MICROCATHETER, THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THE DEVICE AND ANY ACCESSORIES WERE PREPARED AS INDICATED IN THE IFU. THE CATHETER WAS FLUSHED AS INDICATED IN THE IFU. THERE WAS RESISTANCE IN THE DISTAL SECTION OF THE CATHETER. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF AN ARTERIOVENOUS MALFORMATION (AVM). IT WAS NOTED THE PATIENT'S VESSEL TORTUOSITY WAS MODERATE. THE ACCESS VESSEL WAS THE FEMORAL ARTERY WITH A DIAMETER OF 10 MM. THE PATIENT IS ALIVE WITH NO INJURY. ADDITIONAL INFORMATION RECEIVED CLARIFIED THAT THE MICROGUIDEWIRE COULD NOT COME OUT, AND THE SURGEON CONSIDERED WHETHER THE MICROCATHETER'S INNER DIAMETER WAS TOO SMALL TO COME OUT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT A
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23397326 ↗MalfunctionOct 15, 2025Burst Container or Vessel; Leak/Splash; Burst Container or Vessel; Leak/Splash
ADDITIONAL INFORMATION WAS RECEIVED. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF A FISTULA ANEURYSM. VESSEL TORTUOSITY WAS MODERATE. THE PATIENT IS CURRENTLY RECOVERING WELL. NO SYMPTOMS RELATED TO THE LEAK. THE CAUSE OF THE LEAK IS UNKNOWN. APOLLO RUPTURE WAS CONFIRMED DURING HYDRATION; ONYX WAS NOT INJECTED. THE LIQUID EMBOLIC AGENT WAS NOT IMPLANTED, SO NO MEDICAL INTERVENTION WAS REQUIRED. THE ANATOMICAL LOCATION OF THE APOLLO TIP WAS IN THE DETACHABLE ZONE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23300822 ↗MalfunctionAug 18, 2025Detachment of Device or Device Component; Packaging Problem; Detachment of Device or Device Component; Packaging Problem
ADDITIONAL INFORMATION RECEIVED REPORTED THAT THERE WAS NO DAMAGE OR EVIDENCE OF TAMPERING TO THE DEVICE PACKAGING. THE CAUSE OF THE DAMAGE WAS NOT DETERMINED, AND THERE WERE NO ISSUES THAT RESULTED IN THE DEVICE DAMAGE THAT WAS FOUND. NO PREP WAS PERFORMED PRIOR TO THE DAMAGE BEING SEEN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22833389 ↗MalfunctionAug 15, 2025Difficult to Remove; Detachment of Device or Device Component; Difficult to Remove; Detachment of Device or Device Component
ADDITIONAL INFORMATION WAS RECEIVED. ONYX WAS USED AS THE LIQUID EMBOLIC DURING THE PROCEDURE. THE CAUSE OF THE EVENT WAS NOT DETERMINED. THE TIP DETACHMENT OCCURRED DURING NAVIGATION THROUGH THE GUIDE CATHETER. NO KINK OR DAMAGE WAS NOTICED ON ANY OF THE DEVICES DURING THE PROCEDURE.
Patient problems: Foreign Body In Patient; Vasoconstriction
MAUDE report 22813578 ↗MalfunctionAug 12, 2025Detachment of Device or Device Component; Difficult to Advance; Physical Resistance/Sticking; Detachment of Device or Device Component; Difficult to Advance; Physical Resistance/Sticking
MEDTRONIC RECEIVED A REPORT THAT DURING THE INJECTION PROCESS, THE CATHETER'S DETACHMENT SECTION ACCIDENTALLY DETACHED. NO PATIENT SYMPTOMS OR FURTHER COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT. THE DEVICE AND ANY ACCESSORIES WERE PREPARED AS INDICATED IN THE IFU. THE CATHETER WAS FLUSHED AS INDICATED IN THE IFU. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF AN ARTERIOVENOUS MALFORMATION. IT WAS NOTED THE PATIENT'S VESSEL TORTUOSITY WAS MINIMAL. THE ACCESS VESSEL WAS THE FEMORAL ARTERY WITH A DIAMETER OF 10MM.
Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions
MAUDE report 22785063 ↗MalfunctionJul 30, 2025Physical Resistance/Sticking; Physical Resistance/Sticking
MEDTRONIC RECEIVED A REPORT THAT THE APOLLO CATHETER ENCOUNTERED RESISTANCE WITH ONYX. THE PATIENT WAS UNDERGOING A TREATMENT OF AN ARTERIOVENOUS MALFORMATION. IT WAS NOTED THE PATIENT'S VESSEL TORTUOSITY WAS MINIMAL. THE ACCESS VESSEL WAS THE RADIAL ARTERY WITH 2MM DIAMETER. IT WAS REPORTED THAT WHILE USING THE APOLLO CATHETER TO DELIVER ONYX FOR TREATING AN ARTERIOVENOUS MALFORMATION IN THE P2 SEGMENT OF THE POSTERIOR CEREBRAL ARTERY, THE SURGEON ENCOUNTERED RESISTANCE INSIDE THE CATHETER AFTER INJECTING A PORTION OF THE LIQUID EMBOLIC AGENT. THE SYRINGE COULD NOT BE ADVANCED FURTHER. TO ENSURE SURGICAL SAFETY, THE APOLLO CATHETER WAS WITHDRAWN AND REPLACED WITH A NEW CATHETER, ALLOWING THE PROCEDURE TO CONTINUE. THE REPORTED DEVICE AND ANY ACCESSORY DEVICES WERE PREPARED AND THE CATHETER WAS FLUSHED AS INDICATED IN THE INSTRUCTIONS FOR USE (IFU). NO PATIENT SYMPTOMS OR COMPLICATIONS WERE ASSOCIATED WITH THIS EVENT. ADDITIONAL INFORMATION WAS RECEIVED THAT THE AVM WAS LOCATED IN THE OCCIPITAL LOBE. THE CAUSE OF THE RESISTANCE WAS NOT DETERMINED. IT WAS UNABLE TO BE DETERMINED IF THE ONYX HAD PREMATURE SOLIDIFICATION. THERE WAS NO KINK OR DAMAGE NOTICED ON THE CATHETER. IT WAS NOT
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22662148 ↗MalfunctionJul 8, 2025Compatibility Problem; Compatibility Problem
MEDTRONIC RECEIVED A REPORT THAT AN APOLLO CATHETER WAS OCCLUDED DURING ONYX INJECTION. AT THE BEGINNING OF THE OPERATION, THE NURSE TOOK OUT THE GLUE AND STARTED TO VIBRATE IT. AFTER THE APOLLO WAS IN PLACE, THE SURGEON PREFILLED THE APOLLO WITH DMSO SOLVENT AND STARTED TO INJECT GLUE INTO THE GLUE INTERFACE. THEY PUSHED THE SYRINGE AND FOUND THAT THE SYRINGE COULD NOT BE PUSHED, AND THE SCREEN WAS DEVELOPED. THEN THEY TOOK IT OUT FOR IN VITRO TESTING AND FOUND THAT THE MICROCATHETER WAS BLOCKED AND GLUE COULD NOT BE INJECTED. TO ENSURE THE OPERATION, THE SURGEON REPLACED IT WITH A NEW APOLLO AND USED THIS ONE NORMALLY. THE REPORTED DEVICE AND ANY ACCESSORY DEVICES WERE PREPARED AND THE CATHETER WAS FLUSHED AS INDICATED IN THE INSTRUCTIONS FOR USE (IFU). NO PATIENT SYMPTOMS OR COMPLICATIONS WERE ASSOCIATED WITH THIS EVENT. IT WAS NOTED THAT THE SURGEON WAS AN EXPERIENCED CHIEF CEREBROVASCULAR PHYSICIAN, WHO TREATED 20-30 CASES OF RELATED MALFORMATIONS AND FISTULAS EVERY YEAR. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF THE RIGHT FRONTAL ARTERIOVENOUS MALFORMATION. IT WAS NOTED THE PATIENT'S VESSEL TORTUOSITY WAS MINIMAL. THE ACCESS VESSEL WAS THE FEMORAL ARTERY WITH AN 8MM
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22446127 ↗MalfunctionJun 27, 2025Detachment of Device or Device Component; Physical Resistance/Sticking; Detachment of Device or Device Component; Physical Resistance/Sticking
MEDTRONIC RECEIVED A REPORT THAT THE PATIENT WAS UNDERGOING TREATMENT FOR A VASCULAR MALFORMATION AND INJECTION OF GLUE. IT WAS NOTED THAT THE PATIENT'S VESSEL TORTUOSITY WAS MODERATE. THE ACCESS VESSEL WAS THE FEMORAL ARTERY, WITH 6.9MM DIAMETER. IT WAS REPORTED THAT IN THE PROCESS OF INJECTING GLUE, THE APOLLO CATHETER END TIP FELL OFF. IT WAS NOTED THAT THE CATHETER ENCOUNTERED RESISTANCE WITHIN THE DISTAL SECTION. THE REPORTED DEVICE AND ANY ACCESSORY DEVICES WERE PREPARED AS INDICATED IN THE INSTRUCTIONS FOR USE (IFU). NO PATIENT SYMPTOMS OR COMPLICATIONS WERE ASSOCIATED WITH THIS EVENT.
Patient problems: Foreign Body In Patient
MAUDE report 22349694 ↗MalfunctionJun 26, 2025Physical Resistance/Sticking; Physical Resistance/Sticking
MEDTRONIC RECEIVED A REPORT THAT 2 APOLLO CATHETERS HAD RESISTANCE IN THEIR DISTAL ENDS. THE PATIENT HAD INTRACRANIAL ARTERIOVENOUS MALFORMATION. THE SURGEON USED A 0.010 MICRO-GUIDEWIRE TO GUIDE THE APOLLO MICRO-CATHETER INTO POSITION BUT FOUND THAT WHEN IT ROSE TO THE MIDDLE CEREBRAL ARTERY, THE MICRO-GUIDEWIRE COULD NOT COME OUT AND THE MICROCATHETER COULD NOT BE GUIDED FURTHER. AFTER REPEATED ATTEMPTS, THE CATHETER COULD NOT BE BROUGHT INTO POSITION. AFTER REPLACING IT WITH ANOTHER APOLLO MICROCATHETER, A SIMILAR SITUATION STILL OCCURRED. AFTER REPLACING IT WITH MARATHON, THE SURGERY WAS SUCCESSFULLY COMPLETED. THE SURGEON SAID THAT THE PREVIOUS TWO APOLLO MICROCATHETERS WOULD NOT BE CHARGED. THE REPORTED DEVICES AND ANY ACCESSORY DEVICES WERE PREPARED AND THE CATHETERS WERE FLUSHED AS INDICATED IN THE IFU AS INDICATED IN THE INSTRUCTIONS FOR USE (IFU). IT WAS UNKNOWN IF ANY PATIENT SYMPTOMS OR COMPLICATIONS WERE ASSOCIATED WITH THIS EVENT. THE PATIENT WAS UNDERGOING AVM (ARTERIOVENOUS MALFORMATION) EMBOLIZATION. IT WAS NOTED THE PATIENT'S VESSEL TORTUOSITY WAS SEVERE. T HE ACCESS VESSEL WAS THE FEMORAL ARTERY WITH A 10MM DIAMETER. ADDITIONAL INFORMATION WAS RECEIVED STATING TH
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22340889 ↗
FDA clearance and classification
Approved through premarket approval (PMA) P030004 on Jul 21, 2005. Submitted by Ev3 Neurovascular as “Onyx Liquid Embolic System”. The FDA review took 856 days. The FDA has since approved 41 supplements (changes to the device or its labeling), most recently on Oct 14, 2026. FDA record ↗
Devices of this type need premarket approval (PMA), the FDA's most demanding review, usually with clinical data.
Common questions about the Apollo 105-5096-000
Has the Apollo 105-5096-000 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00763000446758), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Apollo 105-5096-000?
The FDA's MAUDE database lists 79 adverse event reports for this device, received between May 8, 2023 and Sep 3, 2026. In those reports, the most common specific device problems are “Detachment of Device or Device Component” (39), “Physical Resistance/Sticking” (19) and “Burst Container or Vessel” (15); the most common patient problems are “Foreign Body In Patient” (24), “Vasoconstriction” (3) and “Intracranial Pressure Increased” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Apollo 105-5096-000?
None of the reports are classified as deaths, 3 as injuries and 76 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Apollo 105-5096-000 FDA approved?
It was approved through premarket approval (PMA) P030004 on Jul 21, 2005.
Is the Apollo 105-5096-000 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the Apollo 105-5096-000, and is it still sold?
The FDA's device database lists Micro Therapeutics, Inc. as the company responsible for it (brand Apollo, model 105-5096-000). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Agent, injectable, embolic devices
- ApolloTM 105-5096-000Micro Therapeutics, Inc. · Agent, injectable, embolic64 reports
- ONYX 18 105-7100-060Micro Therapeutics, Inc. · Agent, injectable, embolic1 death43 reports
- ApolloTM 105-5095-000Micro Therapeutics, Inc. · Agent, injectable, embolic34 reports
- Apollo 105-5095-000Micro Therapeutics, Inc. · Agent, injectable, embolic17 reports
- Onyx 18 105-7100-060Micro Therapeutics, Inc. · Agent, injectable, embolic16 reports
- ONYX 34 105-7100-080Micro Therapeutics, Inc. · Agent, injectable, embolic10 reports
- Onyx 34 105-7100-080Micro Therapeutics, Inc. · Agent, injectable, embolic1 report
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page