Pipeline TM Flex Embolization Device with Shield Technology TM PED2-400-16Recalls & reported problems
Micro Therapeutics, Inc. · STENT PED2-400-16 V08 MDR · Class 3 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00763000449186), received between Jan 12, 2026 and Apr 24, 2026 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Activation Failure” (2 reports).
12 other recalls cover devices cleared under the same FDA submission (P100018); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Activation Failure2
- Material Deformation1
- Physical Resistance/Sticking1
- Retraction Problem1
- Structural Problem1
- Unintended Movement1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 24, 2026Retraction Problem; Structural Problem; Unintended Movement; Activation Failure; Physical Resistance/Sticking
MEDTRONIC RECEIVED A REPORT THAT THE PATIENT WAS UNDERGOING TREATMENT FOR A SACCULAR, UNRUPTURED LEFT CLINOID C5 SEGMENT ANEURYSM WITH A MAX DIAMETER OF 6.3 MM AND A 3.0 MM NECK DIAMETER. IT WAS NOTED THAT THE PATIENT'S VESSEL TORTUOSITY WAS SEVERE. THE ACCESS VESSEL WAS THE FEMORAL ARTERY, WITH 7.8MM DIAMETER. THE LANDING ZONE WAS 3.1MM DISTALLY AND 4.0MM PROXIMALLY. DUAL ANTIPLATELET TREATMENT (DAPT) WAS ADMINISTERED. THE PRU LEVEL WAS NORMAL. IT WAS REPORTED THAT AFTER THE OPERATOR ESTABLISHED ACCESS, A PHENOM 27 MICROCATHETER WAS POSITIONED IN PLACE. A SHIELD PED400-18 DEVICE WAS UNPACKED AND HYDRATED. THE DISTAL END OF THE DEVICE WAS ADVANCED THROUGH THE Y-VALVE TO THE PROXIMAL/TAIL END OF THE PHENOM 27 MICROCATHETER. WHILE THE ASSISTANT ADVANCED THE DELIVERY WIRE TO ITS FULL EXTENT, VERY LITTLE RESISTANCE WAS NOTED. THE DISTAL END OF THE DEVICE WAS UNABLE TO ENTER THE PROXIMAL/TAIL END OF THE PHENOM 27 MICROCATHETER. THE OPERATOR THEN WITHDREW THE ENTIRE DEVICE ONTO THE TABLE AND ATTEMPTED TO ADVANCE THE DELIVERY WIRE BACK AND FORTH; HOWEVER, THE STENT REMAINED STATIONARY. IT WAS SUBSEQUENTLY NOTED THAT THE DEVICE HAD BECOME DETACHED (DEVICE DETACHMENT). THE OPERATOR THEN REP
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24982287 ↗MalfunctionJan 12, 2026Material Deformation; Activation Failure
MEDTRONIC RECEIVED A REPORT THAT THE PIPELINE FLEX STENT FAILED TO OPEN IN THE DISTAL END. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF A SACCULAR, UNRUPTURED ANEURYSM IN THE OPHTHALMIC SEGMENT OF THE INTERNAL CAROTID ARTERY. IT HAD A MAX DIAMETER OF 5 MM AND A 4 MM NECK DIAMETER. THE LANDING ZONE WAS THE POSTERIOR COMMUNICATING SEGMENT OF THE INTERNAL CAROTID ARTERY, AVOIDING THE ANTERIOR AND POSTERIOR COMMUNICATING ARTERIES, AND IN THE C5 REGION. IT WAS NOTED THE PATIENT'S VESSEL TORTUOSITY WAS NORMAL. THE PATIENT HAS A MEDICAL HISTORY OF STAGE 2 HYPERTENSION. DUAL ANTIPLATELET THERAPY (DAPT) WAS ADMINISTERED AT A PLATELET REACTIVITY UNITS (PRU) LEVEL OF 100 MG ASPIRIN, 75 MG OF CLOPIDOGREL, WHICH WERE GIVE DAILY FOR 5 DAYS, PRIOR TO SURGERY. THE ANGIOGRAPHIC RESULT POST PROCEDURE WAS OF AN ANEURYSM OF THE OPHTHALMIC SEGMENT OF THE INTERNAL CAROTID ARTERY. IT WAS REPORTED THAT ON (B)(6) 2026, THE OPERATOR PERFORMED A FLOW-DIVERTING STENT IMPLANTATION FOR AN ANEURYSM AT THE OPHTHALMIC SE GMENT OF THE INTERNAL CAROTID ARTERY. A PHENOM 27 CATHETER WAS USED, AND THE PED2-400-16 STENT WAS SELECTED. A 6F NAVIEN GUIDE WIRE WAS USED TO DELIVER THE PHENOM 27 CATHETER TO THE M1 SEGM
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24036357 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P100018. One submission often covers many models, so check the product and lot details in each record.
- Z-1246-2025 · Jan 30, 2025 · Open, ClassifiedMedtronic Pipeline Vantage Embolization Device with Shield Technology, REF: PED3-021-350-16, PED3-021-250-14, PED3-021-350-20, PED3-021-250-20, PED3-021-300-16, PED3-021-300-20, PE
- Z-1245-2025 · Jan 30, 2025 · Open, ClassifiedMedtronic Pipeline Vantage Embolization Device with Shield Technology, REF: PED3-027-350-12, PED3-027-350-14, PED3-027-350-16, PED3-027-350-20, PED3-027-350-25, PED3-027-400-12, PE
- Z-2403-2021 · Jul 13, 2021 · Open, ClassifiedEV3 Pipeline Flex Embolization Device with Flex Shield Technology
- Z-2402-2021 · Jul 13, 2021 · Open, ClassifiedPipeline Flex Embolization Device
- Z-1527-2020 · Feb 14, 2020 · Terminatedev3 Pipeline Flex Embolization Device. For neurological endovascular use.
- Z-0308-2017 · Oct 5, 2016 · TerminatedAlligator Retrieval Device The device is intended for use in the peripheral and neurovasculature for foreign body removal. Catalog No. FA-88810-20V04 FA-88810-20V06 FA-88810-30V04
- Z-0310-2017 · Oct 5, 2016 · TerminatedX-Celerator Hydrophilic Exchange Guidewire Catalog No. 103-0601-300 103-0601-350V10 103-0602-350V06 103-0601-300V01 103-0602-300 103-0602-350V09 103-0601-300V02 103-0602-300V01 10
- Z-0309-2017 · Oct 5, 2016 · TerminatedMarathon Flow Directed Micro Catheter, UltraFlow HPC Flow Directed Micro Catheter Marathon Flow: The device is intended to access periphera and neurovasculature for the controlled
- Z-0307-2017 · Oct 5, 2016 · TerminatedPipeline Embolization Device (Pipeline Classic) The device is indicated for the endovascular treatment of adults (22 year of age or older) with large or giant widenecked intracrani
- Z-0120-2016 · Sep 4, 2015 · Terminated-Pipeline Embolization Device -Pipeline Flex Embolization Device The Pipeline and Pipeline Flex Embolization Devices consist of a permanent implant combined with a guidewire based
- Z-1453-2014 · Apr 1, 2014 · TerminatedPipeline Embolization Device (PED). Used endovascular treatment of adults with intracranial aneurysms in the internal carotid artery.
- Z-2045-2013 · Jul 19, 2013 · TerminatedPipeline Embolization Device (PED) The Pipeline Embolization Device is indicated for the endovascular treatment of adults (22 years of age and older) with large or giant wide-neck
Other Intracranial aneurysm flow diverter devices
- Pipeline TM Flex Embolization Device with Shield Technology TM PED2-300-12Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter2 recalls1 report
- PIPELINE PED-500-20Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall31 reports
- PIPELINE PED-500-35Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall1 death29 reports
- PIPELINE PED-500-16Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall2 deaths27 reports
- PIPELINE PED-500-25Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall1 death23 reports
- PIPELINE PED-500-30Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall21 reports
- PIPELINE PED-450-25Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall20 reports
- PIPELINE PED-475-30Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall1 death19 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page