Pipeline TM Flex Embolization Device with Shield Technology TM PED2-500-30Recalls & reported problems
Micro Therapeutics, Inc. · STENT PED2-500-30 V08 MDR · Class 3 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00763000449582), received between Nov 21, 2025 and Sep 9, 2026 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Activation Failure” (5 reports).
12 other recalls cover devices cleared under the same FDA submission (P100018); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Activation Failure5
- Material Deformation2
- Physical Resistance/Sticking1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Vascular Dissection1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 9, 2026Activation Failure
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50690516 ↗MalfunctionAug 26, 2026Material Deformation; Activation Failure
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50457076 ↗MalfunctionMar 25, 2026Activation Failure; Physical Resistance/Sticking
ADDITIONAL INFORMATION WAS RECEIVED. NO CAUSES OR CONTRIBUTING FACTORS FOR THE EVENT WERE IDENTIFIED. RESISTANCE WAS NOTED IN THE DISTAL SECTION OF THE CATHETER. THE CATHETER WAS FLUSHED PER IFU DURING THE PROCEDURE. THE PIPELINE WAS NOT PLACED IN A VESSEL BEND WHEN IT FAILED TO OPEN, AND NO ADDITIONAL STEPS OR OTHER DEVICES (SUCH AS RESHEATHING, WIRE OR CATHETER MANIPULATION, OR BALLOON USE) WERE ATTEMPTED TO OPEN THE PIPELINE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24686762 ↗MalfunctionJan 29, 2026Activation Failure
MEDTRONIC RECEIVED INFORMATION REGARDING PIPELINE FLEX WITH SHIELD EMBOLIZATION DEVICE THAT FAILED TO OPEN. THE PATIENT WAS UNDERGOING A PROCEDURE FOR FLOW DIVERTER STENT TREATMENT OF AN INTERNAL CAROTID ARTERY (ICA) C4 SEGMENT UNRUPTURED AMORPHOUS ANEURYSM. THE ANEURYSM MAX DIAMETER WAS 10MM AND THE NECK DIAMETER WAS 5.3MM. THE DISTAL LANDING ZONE WAS 4.5MM; THE PROXIMAL LANDING ZONE WAS 4.9MM. VESSEL TORTUOSITY WAS MODERATE. IT WAS REPORTED THAT THE PIPELINE AND ALL ACCESSORY DEVICES WERE PREPARED AS INDICATED IN THE INSTRUCTIONS FOR USE (IFU). AFTER THE CATHETER DELIVERY SYSTEM WAS POSITIONED IN LACE, THE PIPELINE WAS HYDRATED AND DELIVERED TO THE MIDDLE CEREBRAL ARTERY (MCA) STRAIGHT SEGMENT. THE MICROCATHETER WAS RETRACTED TO START DEPLOYMENT OF THE PIPELINE STENT. HOWEVER, THE TIP OF THE STENT DID NOT OPEN. DEPLOYMENT WAS CONTINUED WITH REPEATED PUSHING AND PULLING BUT THE TIP STILL DID NOT OPEN. THE PIPELINE WAS FULLY RESHEATHED IN THE MICROCATHETER AND REPOSITIONED AT THE PROXIMAL END OF THE POSTERIOR COMMUNICATING ARTERY (PCOM) FOR DEPLOYMENT BUT THE PIPELINE STILL DID NOT OPEN. THE DEVICE WAS REMOVED AND REPLACED TO COMPLETE THE PROCEDURE SUCCESSFULLY. THERE WAS NO HARM O
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24209703 ↗MalfunctionNov 21, 2025Material Deformation; Activation Failure; Material Deformation; Activation Failure
ADDITIONAL INFORMATION WAS RECEIVED. THE CAUSE OF THE PIPELINE FAILURE TO OPEN WAS NOT DETERMINED. THE PIPELINE HAD NOT BEEN PLACED IN A VESSEL BEND WHEN IT FAILED TO OPEN, AND IT WAS NOT RESHEATHED DURING THE PROCEDURE. NO VISIBLE DAMAGE WAS OBSERVED ON THE PHENOM 27 MICROCATHETER
Patient problems: Vascular Dissection
MAUDE report 23620836 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P100018. One submission often covers many models, so check the product and lot details in each record.
- Z-1245-2025 · Jan 30, 2025 · Open, ClassifiedMedtronic Pipeline Vantage Embolization Device with Shield Technology, REF: PED3-027-350-12, PED3-027-350-14, PED3-027-350-16, PED3-027-350-20, PED3-027-350-25, PED3-027-400-12, PE
- Z-1246-2025 · Jan 30, 2025 · Open, ClassifiedMedtronic Pipeline Vantage Embolization Device with Shield Technology, REF: PED3-021-350-16, PED3-021-250-14, PED3-021-350-20, PED3-021-250-20, PED3-021-300-16, PED3-021-300-20, PE
- Z-2402-2021 · Jul 13, 2021 · Open, ClassifiedPipeline Flex Embolization Device
- Z-2403-2021 · Jul 13, 2021 · Open, ClassifiedEV3 Pipeline Flex Embolization Device with Flex Shield Technology
- Z-1527-2020 · Feb 14, 2020 · Terminatedev3 Pipeline Flex Embolization Device. For neurological endovascular use.
- Z-0307-2017 · Oct 5, 2016 · TerminatedPipeline Embolization Device (Pipeline Classic) The device is indicated for the endovascular treatment of adults (22 year of age or older) with large or giant widenecked intracrani
- Z-0308-2017 · Oct 5, 2016 · TerminatedAlligator Retrieval Device The device is intended for use in the peripheral and neurovasculature for foreign body removal. Catalog No. FA-88810-20V04 FA-88810-20V06 FA-88810-30V04
- Z-0309-2017 · Oct 5, 2016 · TerminatedMarathon Flow Directed Micro Catheter, UltraFlow HPC Flow Directed Micro Catheter Marathon Flow: The device is intended to access periphera and neurovasculature for the controlled
- Z-0310-2017 · Oct 5, 2016 · TerminatedX-Celerator Hydrophilic Exchange Guidewire Catalog No. 103-0601-300 103-0601-350V10 103-0602-350V06 103-0601-300V01 103-0602-300 103-0602-350V09 103-0601-300V02 103-0602-300V01 10
- Z-0120-2016 · Sep 4, 2015 · Terminated-Pipeline Embolization Device -Pipeline Flex Embolization Device The Pipeline and Pipeline Flex Embolization Devices consist of a permanent implant combined with a guidewire based
- Z-1453-2014 · Apr 1, 2014 · TerminatedPipeline Embolization Device (PED). Used endovascular treatment of adults with intracranial aneurysms in the internal carotid artery.
- Z-2045-2013 · Jul 19, 2013 · TerminatedPipeline Embolization Device (PED) The Pipeline Embolization Device is indicated for the endovascular treatment of adults (22 years of age and older) with large or giant wide-neck
Other Intracranial aneurysm flow diverter devices
- Pipeline TM Flex Embolization Device with Shield Technology TM PED2-300-12Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter2 recalls1 report
- PIPELINE PED-500-20Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall31 reports
- PIPELINE PED-500-35Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall1 death29 reports
- PIPELINE PED-500-16Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall2 deaths27 reports
- PIPELINE PED-500-25Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall1 death23 reports
- PIPELINE PED-500-30Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall21 reports
- PIPELINE PED-450-25Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall20 reports
- PIPELINE PED-475-30Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall1 death19 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page