NA 106A3 (UDI 00763000486778)Recalls & reported problems
Medtronic CryoCath LP · CONSOLE 106A3 GEN V ROHS PLX · Class 3
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00763000486778), received between Dec 25, 2023 and May 14, 2025 — classified as 3 injuries.
The most frequently reported specific device problem is “Protective Measures Problem” (3 reports).
2 other recalls cover devices cleared under the same FDA submission (P020045); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem3
- Protective Measures Problem3
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryMay 14, 2025Adverse Event Without Identified Device or Use Problem; Protective Measures Problem; Adverse Event Without Identified Device or Use Problem; Protective Measures Problem
IT WAS REPORTED THAT DURING A CRYO ABLATION PROCEDURE, AN ERROR MESSAGE WAS RECEIVED INDICATING THAT THERE WAS A SERIAL PERIPHERAL INTERFACE (SPI) CIRCULAR REDUNDANCY CHECK ERROR. THE CASE WAS ABORTED WHILE THE PATIENT WAS UNDER DEEP SEDATION AND/OR GENERAL ANESTHESIA. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22022601 ↗InjuryJan 26, 2024Adverse Event Without Identified Device or Use Problem; Protective Measures Problem; Adverse Event Without Identified Device or Use Problem; Protective Measures Problem
IT WAS REPORTED THAT DURING A CRYO ABLATION PROCEDURE, A SYSTEM NOTICE WAS RECEIVED INDICATING THAT THE SAFETY SYSTEM DETECTED A HIGH LEVEL OF REFRIGERANT FLOW AND STOPPED THE INJECTION. THE ISSUE WAS UNABLE TO BE RESOLVED AND THE PROCEDURE WAS ABORTED. IT IS UNKNOWN WHETHER THE PATIENT WAS UNDER GENERAL ANESTHESIA. A HIGH LEVEL OF PHYSICIAN FRUSTRATION WAS NOTED. DURING A LATER SERVICING, AN ATYPICAL INJECTION PROPORTIONAL VALVE (MKS) FAILURE WAS OBSERVED. THE MKS AND VALVE BOARD WERE REPLACED WHICH RESOLVED THE ISSUES. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18592063 ↗InjuryDec 25, 2023Adverse Event Without Identified Device or Use Problem; Protective Measures Problem; Adverse Event Without Identified Device or Use Problem; Protective Measures Problem
IT WAS REPORTED THAT DURING A CRYO ABLATION PROCEDURE, A SYSTEM NOTICE WAS RECEIVED INDICATING THAT THE SYSTEM DETECTED A HIGH LEVEL OF REFRIGERANT FLOW DURING THE SYSTEM FLUSH, AND THE SYSTEM FLUSH WAS STOPPED. ADDITIONALLY, A SYSTEM NOTICE WAS RECEIVED INDICATING THAT THE SAFETY SYSTEM DETECTED A HIGH LEVEL OF REFRIGERANT FLOW AND STOPPED THE INJECTION. THE COAXIAL UMBILICAL CABLE WAS REPLACED AND THE ISSUE REMAINED. THE CASE WAS ABORTED AND THE PATIENT WAS UNDER GENERAL ANESTHESIA. THE FAILURE FILES WERE REVIEWED AND CONFIRMED THE SYSTEM NOTICE INDICATING THAT THE SAFETY SYSTEM DETECTED A HIGH LEVEL OF REFRIGERANT FLOW AND STOPPED THE INJECTION. IT WAS ALSO NOTED THAT A SYSTEM NOTICE WAS RECEIVED ONCE INDICATING THAT THE REFRIGERANT DELIVERY PATH WAS OBSTRUCTED. AFTER THE CASE,TEST INJECTIONS WERE PERFORMED USING A FOCAL CATHETER. A FEW SUCCESSFUL ABLATIONS WERE PERFORMED AND THEN A SYSTEM NOTICE WAS RECEIVED INDICATING THAT THE SAFETY SYSTEM DETECTED A HIGH LEVEL OF REFRIGERANT FLOW AND STOPPED THE INJECTION. A FEW TEST INJECTIONS WERE PERFORMED USING A BALLOON CATHETER. THE DATA FILES WERE ANALYZED AND CONFIRMED SYSTEM NOTICES A SYSTEM NOTICE WAS RECEIVED INDICATING THAT THE S
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18390524 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P020045. One submission often covers many models, so check the product and lot details in each record.
- Z-2777-2015 · Sep 1, 2015 · TerminatedMedtronic CryoConsole, models 106A3, 106E2, and 106A2-K For use in performing cardiac ablation procedures.
- Z-1452-06 · Aug 4, 2006 · Terminated7F Freezor Cardiac Cryoablation catheter, REF # 207F1. CryoCath Technologies Inc.
Other Cardiac ablation percutaneous catheter devices
- THERMOCOOL SMARTTOUCH SF D134805Biosense Webster Inc · Cardiac ablation percutaneous catheter1 recall176 deaths4,106 reports
- THERMOCOOL SMARTTOUCH SF D134804Biosense Webster Inc · Cardiac ablation percutaneous catheter44 deaths784 reports
- THERMOCOOL SMARTTOUCH SF D134801Biosense Webster Inc · Cardiac ablation percutaneous catheter15 deaths608 reports
- ThermoCool SmartTouch D132705Biosense Webster Inc · Cardiac ablation percutaneous catheter25 deaths574 reports
- THERMOCOOL SMARTTOUCH SF D134702Biosense Webster Inc · Cardiac ablation percutaneous catheter19 deaths366 reports
- THERMOCOOL SMARTTOUCH SF D134701Biosense Webster Inc · Cardiac ablation percutaneous catheter11 deaths278 reports
- ThermoCool SmartTouch D133602Biosense Webster Inc · Cardiac ablation percutaneous catheter8 deaths225 reports
- SMARTABLATE SAT001Biosense Webster Inc · Cardiac ablation percutaneous catheter185 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.