Onyx Frontier™ ONYXNG20026UXRecalls & reported problems

Medtronic, Inc. · STENT ONYXNG20026UX ONYX 2.00X26RX · Class 3 · Implant

All models of Onyx Frontier (71)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00763000511166), received between Dec 6, 2022 and Feb 12, 2026 — classified as 4 injuries, 5 malfunctions.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated6
  • Positioning Problem4
  • Material Deformation3
  • Detachment of Device or Device Component1
  • Device Damaged by Another Device1
  • Difficult to Remove1
  • Entrapment of Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Insufficient Information4
  • Device Embedded In Tissue or Plaque1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 12, 2026Material Deformation

    IT WAS REPORTED THAT DURING A PROCEDURE INVOLVING ONE ONYX FRONTIER CORONARY DRUG ELUTING STENT, STENT DEFORMATION OCCURRED IN VIVO DURING POSITIONING WITHIN A HEAVILY CALCIFIED RIGHT CORONARY ARTERY. THE DEVICE WAS NOT INSPECTED PRIOR TO USE. NEGATIVE PREP WAS PERFORMED WITH NO ISSUES. THE LESION WAS PRE-DILATED. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY-DEPLOYED STENT. RESISTANCE WAS ENCOUNTERED WHEN ADVANCING THE DEVICE, AND EXCESSIVE FORCE WAS USED DURING DELIVERY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24321861 ↗
  • MalfunctionJun 12, 2025Material Deformation; Positioning Problem; Material Deformation; Positioning Problem

    IT WAS REPORTED THAT DURING A PROCEDURE INVOLVING ONE ONYX FRONTIER CORONARY DRUG ELUTING STENT, STENT DEFORMATION OCCURRED IN VIVO DURING POSITIONING/ADVANCEMENT. THE LESION BEING TREATED WAS A MODERATELY TORTUOUS, MODERATELY CALCIFIED LESION WITH 90% STENOSIS IN THE PROXIMAL RIGHT CORONARY ARTERY (RCA). THE DEVICE WAS INSPECTED PRIOR TO USE WITH NO ISSUES NOTED. NEGATIVE PREP WAS PERFORMED WITH NO ISSUES NOTED. THE LESION WAS PRE-DILATED USING A NUMBER OF NON-MEDTRONIC SEMI-COMPLIANT AND NON-COMPLIANT BALLOONS. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT. RESISTANCE WAS ENCOUNTERED WHEN ADVANCING THE DEVICE. EXCESSIVE FORCE WAS NOT USED DURING DELIVERY. THERE WERE TWO ATTEMPTS MADE TO DELIVER THE STENT BUT THEY WERE UNSUCCESSFUL. UPON REMOVAL OF THE STENT AFTER THE SECOND ATTEMPT, THE STENT FOUND TO BE MANGLED AT THE TIP AREA. THE PROCEDURE WAS COMPLETED WITH TWO SMALLER ONYX FRONTIER STENTS BEING IMPLANTED. THE PATIENT WAS REPORTED TO BE FINE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22194033 ↗
  • MalfunctionOct 25, 2024Device Dislodged or Dislocated; Material Deformation; Positioning Problem; Device Dislodged or Dislocated; Material Deformation; Positioning Problem

    AN ATTEMPT WAS MADE TO USE ONE ONYX FRONTIER CORONARY DRUG ELUTING STENT TO TREAT A MODERATELY TORTUOUS, MILDLY CALCIFIED LESION WITH 90% STENOSIS IN THE MID CIRCUMFLEX (CX) ARTERY. THE DEVICE WAS INSPECTED WITH NO ISSUES. NEGATIVE PREP WAS NOT PERFORMED. THE LESION WAS PRE-DILATED. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT. RESISTANCE WAS ENCOUNTERED DURING THE DELIVERY OF THE DEVICE. EXCESSIVE FORCE WAS NOT USED DURING DELIVERY. IT WAS REPORTED THAT STENT DISLODGEMENT OCCURRED AFTER REMOVAL FROM THE PATIENT. THE PATIENT IS ALIVE WITH NO INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20532373 ↗
  • InjurySep 17, 2024Difficult to Remove; Device Dislodged or Dislocated; Positioning Problem; Difficult to Remove; Device Dislodged or Dislocated; Positioning Problem

    AN ATTEMPT WAS MADE TO USE ONE ONYX FRONTIER CORONARY DRUG-ELUTING STENT TO TREAT A MILDLY TORTUOUS LESION IN THE MID DIAGONAL BRANCH. THE DEVICE WAS INSPECTED WITH NO ISSUES NOTED. NEGATIVE PREP WAS PERFORMED WITH NO ISSUES NOTED. THE LESION WAS NOT PRE-DILATED. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT. EXCESSIVE FORCE WAS NOT USED DURING DELIVERY. IT WAS REPORTED THAT STENT DISLODGEMENT OCCURRED DURING REMOVAL FOLLOWING A FAILED DELIVERY. AFTER THE STENT COULD NOT BE DELIVERED, THE DELIVERY SYSTEM WAS PULLED BACK FROM THE LESION BUT THE STENT CAME OFF AND WAS IN BETWEEN THE GUIDE AND THE AORTA. A SNARE WAS USED TO TRY TO REMOVE THE STENT. WHEN TRYING TO POSITION THE SNARE, THE DISLODGED STENT MOVED BACK INTO THE SUPERFICIAL FEMORAL ARTERY (SFA) GROIN AREA BUT WAS ABLE TO BE RETRIEVED AND REMOVED. NO STENT WAS USED TO TREAT POTENTIAL LESION IN SECOND DIAGONAL. THE PATIENT IS ALIVE WITH NO FURTHER INJURY.

    Patient problems: Insufficient Information

    MAUDE report 20234713 ↗
  • InjuryMay 24, 2024Entrapment of Device; Device Damaged by Another Device; Device Dislodged or Dislocated; Positioning Problem; Entrapment of Device; Device Damaged by Another Device; Device Dislodged or Dislocated; Positioning Problem

    AN ATTEMPT WAS MADE TO USE ONE ONYX FRONTIER CORONARY DRUG ELUTING STENT TO TREAT A MODERATELY TORTUOUS, MODERATELY CALCIFIED LESION IN THE MID LEFT MAIN (LM) CORONARY ARTERY. THE PATIENT HAD TORTUOUS VESSELS. THE DEVICE WAS INSPECTED WITH NO ISSUES. NEGATIVE PREP WAS NOT PERFORMED. THE LESION WAS PRE-DILATED. THE DEVICE DID NOT DEVICE PASS THROUGH A PREVIOUSLY DEPLOYED STENT. RESISTANCE WAS ENCOUNTERED WHEN ADVANCING THE DEVICE. EXCESSIVE FORCE WAS NOT USED DURING DELIVERY. IT WAS REPORTED THAT STENT DISLODGEMENT OCCURRED DURING REMOVAL FOLLOWING FAILED DELIVERY. THE DISLODGED STENT WAS NOT REMOVED. AFTER REVIEW OF THE PROCEDURAL FILMS, IT SEEMS LIKELY THAT THE STENT CAUGHT ON THE PREVIOUSLY IMPLANTED MEDTRONIC TAVR VALVE, OR ON THE PREVIOUSLY IMPLANTED PROXIMAL LAD STENT WHEN THE DELIVERY SYSTEM WAS ATTEMPTED TO PASS THROUGH PREVIOUS STENT. THE STENT WOULD NOT CROSS, WAS PULLED BACK AND DISLODGED IN THE LM. THE STENT WAS CRUSHED IN THE LM WITH AN ADDITIONAL STENT. THE PATIENT IS ALIVE WITH NO FURTHER INJURY.

    Patient problems: Device Embedded In Tissue or Plaque; Insufficient Information

    MAUDE report 19391297 ↗
  • MalfunctionFeb 20, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    AN ATTEMPT WAS MADE TO USE ONE ONYX FRONTIER CORONARY DRUG ELUTING STENT TO TREAT A CALCIFIED LESION IN THE MID CIRCUMFLEX (CX) ARTERY. THERE WERE NO ISSUES REMOVING DEVICE FROM HOOP. THERE WERE NO ISSUES DURING PREP. THERE WAS SOME RESISTANCE ENCOUNTERED WHILE TRYING TO DELIVER THE STENT TO THE LESION. NO EXCESSIVE FORCE WAS USED WHILE DELIVERING THE STENT. IT WAS REPORTED THAT THE SHAFT OF THE STENT SNAPPED AND BROKE IN HALF WHEN PLACING THE STENT DOWN THE VESSEL. IT WAS STATED THAT THE SHAFT SNAPPED OUTSIDE THE PATIENT SO IT WAS EASY TO REMOVE. ANOTHER SAME SIZE STENT WAS THEN USED. NO INJURY TO THE PATIENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18740944 ↗
  • InjuryJan 9, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated

    AN ATTEMPT WAS MADE TO USE ONE ONYX FRONTIER CORONARY DRUG ELUTING STENT TO TREAT A LESION. IT WAS REPORTED THAT THE STENT CAME OFF THE BALLOON BEFORE IT WAS DEPLOYED. A WIRE AND A NON MDT GUIDE EXTENSION CATHETER WERE ALSO USED DOWN THE VESSEL. THE STENT WAS ABLE TO BE DEPLOYED WITH NEW BALLOON. NO FURTHER PATIENT INJURY WAS REPORTED. NOTE: THE PROCEDURES PERFORMED WERE RIGHT FEMORAL SHEATH SIDE-PORT ANGIOGRAPHY; BALLOON ANGIOPLASTY; PERCUTANEOUS INTERVENTION ON THE 90% DE NOVO STENOSIS IN THE PROXIMAL LAD, INTERVENTIONAL IVUS EXAMINATION, STENT PLACEMENT, BALLOON ANGIOPLASTY; PERCUTANEOUS INTERVENTION ON THE 75% DE NOVO STENOSIS IN THE PROXIMAL 1ST OBTUSE MARGINAL, BALLOON ANGIOPLASTY; PERCUTANEOUS INTERVENTION ON THE 80% DE NOVO STENOSIS IN THE PROXIMAL LEFT CIRCUMFLEX, STENT PLACEMENT, BALLOON ANGIOPLASTY.

    Patient problems: Insufficient Information

    MAUDE report 16120572 ↗
  • InjuryDec 15, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated

    AN ATTEMPT WAS MADE TO USE ONE ONYX FRONTIER CORONARY DRUG ELUTING STENT TO TREAT A SEVERELY TORTUOUS, SEVERELY CALCIFIED LESION WITH 95% STENOSIS, IN THE DISTAL CIRCUMFLEX (CX) ARTERY. THE DEVICE WAS INSPECTED WITH NO ISSUES. NEGATIVE PREP WAS NOT PERFORMED. THE LESION WAS PRE-DILATED. THE DEVICE DID NOT PASS THOUGH A PREVIOUSLY DEPLOYED STENT. RESISTANCE WAS ENCOUNTERED WHEN ADVANCING THE DEVICE. EXCESSIVE FORCE WAS USED DURING DELIVERY. IT WAS REPORTED THAT STENT DISLODGEMENT OCCURRED DURING REMOVAL FOLLOWING FAILED DELIVERY. THE STENT WAS ATTEMPTED TO BE DELIVERED THREE TIMES INTO THE SEVERELY TORTUOUS AND CALCIFIED VESSEL. THE STENT DISLODGED FROM THE CATHETER AFTER THE THIRD ATTEMPT AND WAS RETRIEVED BY A SNARE. THE DISLODGED STENT WAS REMOVED FROM THE PATIENT. THE PATIENT IS ALIVE WITH NO FURTHER INJURY.

    Patient problems: Insufficient Information

    MAUDE report 15985186 ↗
  • MalfunctionDec 6, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated

    AN ATTEMPT WAS MADE TO USE ONE ONYX FRONTIER DEVICE TO TREAT A SEVERELY TORTUOUS, MODERATELY CALCIFIED LESION WITH 95% STENOSIS IN THE DISTAL RIGHT CORONARY ARTERY (RCA). THE DEVICE WAS INSPECTED WITH NO ISSUES NOTED, NEGATIVE PREP WAS PERFORMED WITH NO ISSUES NOTED. THE LESION WAS PRE-DILATED USING A 1.0 NON MEDTRONIC BALLOON DILATION CATHETER, A 2.0 EUPHORA DEVICE AND ALSO A 2.25 NC EUPHORA DEVICE. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT. RESISTANCE WAS ENCOUNTERED WHEN ADVANCING THE DEVICE HOWEVER EXCESSIVE FORCE WAS NOT USED DURING DELIVERY. IT WAS REPORTED THAT STENT DISLODGEMENT OCCURRED DURING DELIVERY TO/AT THE LESION. THE STENT WAS FOUND IN THE GUIDE CATHETER HUB AND WAS REMOVED. THE PATIENT IS REPORTED TO BE ALIVE WITH NO INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15924084 ↗

Other Coronary drug-eluting stent devices

All Coronary drug-eluting stent devices →