Onyx Frontier™ ONYXNG22530UXRecalls & reported problems

Medtronic, Inc. · STENT ONYXNG22530UX ONYX 2.25X30RX · Class 3 · Implant

All models of Onyx Frontier (71)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00763000511241), received between Dec 22, 2022 and Feb 13, 2026 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated2
  • Deformation Due to Compressive Stress1
  • Detachment of Device or Device Component1
  • Fracture1
  • Positioning Problem1

Patient problems

  • Insufficient Information2
  • Device Embedded In Tissue or Plaque1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 13, 2026Fracture; Deformation Due to Compressive Stress; Detachment of Device or Device Component; Positioning Problem

    IT WAS REPORTED THAT DURING A PROCEDURE INVOLVING TWO ONYX FRONTIER CORONARY DRUG ELUTING STENTS, BOTH DEVICES DETACHED IN VIVO DURING DELIVERY THROUGH A LESION IN THE PROXIMAL CIRCUMFLEX (CX) ARTERY WHICH WAS HEAVILY CALCIFIED. THE PROXIMAL CX ARTERY WAS THEN BALLOONED WITH A 4.0X15MM NC EUPHORA AND THE STENT DEVICES PASSED SUCCESSFULLY. THE DISTAL OBTUSE MARGINAL (OM) ARTERY WAS SUCCESSFULLY STENTED WITH NO ISSUES. THE TARGET LESION WAS LOCATED IN THE DISTAL OM ARTERY WHICH EXHIBITED MODERATE TORTUOSITY, SEVERE CALCIFICATION AND CHRONIC TOTAL OCCLUSION (CTO) 100%. BOTH DEVICES WERE INSPECTED PRIOR TO USE AND NEGATIVE PREP WAS PERFORMED WITH NO ISSUES. THE LESION WAS NOT PRE-DILATED. THE DEVICES DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT. RESISTANCE WAS ENCOUNTERED WHEN ADVANCING BOTH DEVICES. THE DETACHED PORTIONS OF THE DEVICES DO NOT REMAIN IN THE PATIENT AND WERE REMOVED THROUGH THE GUIDE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24330497 ↗
  • InjuryJul 19, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated

    AN ATTEMPT WAS MADE TO USE ONE 2.25X30MM ONYX FRONTIER CORONARY DRUG ELUTING STENT TO TREAT A MILDLY TORTUOUS LESION WITH 90% STENOSIS IN THE MID/PROXIMAL LEFT ANTERIOR DESCENDING (LAD) ARTERY AND LEFT MAIN (LM) CORONARY ARTERY. THE DEVICE WAS INSPECTED WITH NO ISSUES NOTED. NEGATIVE PREP WAS NOT PERFORMED. THE LESION WAS PRE-DILATED. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT. EXCESSIVE FORCE WAS NOT USED DURING DELIVERY. IT WAS REPORTED THAT STENT DISLODGEMENT OCCURRED DURING DELIVERY TO/AT LESION. THE DISLODGED STENT WAS NOT REMOVED FROM THE PATIENT. IT WAS DETAILED THAT THE STENT WAS DEPLOYED USING A 1.2MM BALLOON IN SEQUENTIAL FASHION UP TO 2.5MM. NO FURTHER PATIENT INJURY WAS REPORTED.

    Patient problems: Device Embedded In Tissue or Plaque; Insufficient Information

    MAUDE report 17356609 ↗
  • InjuryDec 22, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated

    AN ATTEMPT WAS MADE TO USE ONE ONYX FRONTIER CORONARY DRUG ELUTING STENT TO TREAT A CALCIFIED LESION IN THE PROXIMAL DIAGONAL ARTERY. IT WAS REPORTED THAT THE STENT FAILED TO CROSS THE LESION AND THE STENT CAME OFF THE BALLOON BEFORE IT WAS DEPLOYED. A SNARE WAS USED TO REMOVE THE STENT OUT OF THE LEFT MAIN ARTERY. NO FURTHER INJURY WAS REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 16031952 ↗