Onyx Frontier™ ONYXNG27538UXRecalls & reported problems

Medtronic, Inc. · STENT ONYXNG27538UX ONYX 2.75X38RX · Class 3 · Implant

All models of Onyx Frontier (71)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00763000511449), received between Nov 1, 2022 and Aug 26, 2024 — classified as 3 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Material Deformation” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 2 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Deformation3
  • Device Dislodged or Dislocated2
  • Positioning Problem2
  • Adverse Event Without Identified Device or Use Problem1
  • Deformation Due to Compressive Stress1
  • Difficult to Advance1

Patient problems

  • Insufficient Information2
  • No Clinical Signs, Symptoms or Conditions2
  • Perforation1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 26, 2024Deformation Due to Compressive Stress; Material Deformation; Positioning Problem; Deformation Due to Compressive Stress; Material Deformation; Positioning Problem

    AN ATTEMPT WAS MADE TO USE ONE ONYX FRONTIER CORONARY DRUG ELUTING STENT TO TREAT A MILDLY TORTUOUS, SEVERELY CALCIFIED LESION IN THE MID LEFT ANTERIOR DESCENDING (LAD) ARTERY. THE DEVICE WAS INSPECTED WITH NO ISSUES. NEGATIVE PREP WAS NOT PERFORMED. THE LESION WAS PRE-DILATED. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT. RESISTANCE WAS ENCOUNTERED WHEN ADVANCING THE DEVICE. EXCESSIVE FORCE WAS NOT USED. IT WAS REPORTED THAT STENT DEFORMATION OCCURRED. DELIVERY OF THE STENT WAS ATTEMPTED BUT WAS NOT POSSIBLE DUE TO THE CALCIFICATION. THE STENT WAS REMOVED AND IT WAS FOUND THAT A STENT STRUT HAD FRAYED. THE PATIENT IS ALIVE WITH NO INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20063503 ↗
  • InjuryApr 3, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    AN ATTEMPT WAS MADE TO USE ONE ONYX FRONTIER CORONARY DRUG ELUTING STENT TO TREAT A LESION. IT WAS REPORTED THAT STENT DISLODGEMENT OCCURRED IN VIVO DURING DELIVERY TO THE LESION. THE DISLODGED STENT WAS REMOVED USING A SNARE. NO FURTHER PATIENT INJURY WAS REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 19031398 ↗
  • MalfunctionFeb 13, 2024Difficult to Advance; Material Deformation; Difficult to Advance; Material Deformation

    AN ATTEMPT WAS MADE TO USE ONE ONYX FRONTIER CORONARY DRUG ELUTING STENT TO TREAT A MILDLY TORTUOUS, SEVERELY CALCIFIED LESION IN THE MID LEFT ANTERIOR DESCENDING (LAD). THE DEVICE WAS INSPECTED WITH NO ISSUES NOTED. NEGATIVE PREP WAS PERFORMED WITH NO ISSUES. THE LESION WAS PRE-DILATED WITH A BALLOON. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT. RESISTANCE WAS NOTED WHILE ADVANCING THE DEVICE. EXCESSIVE FORCE WAS NOT USED. IT WAS REPORTED THAT STENT DEFORMATION OCCURRED IN VIVO DURING POSITIONING/ADVANCEMENT. IT WAS DETAILED THAT AS THE STENT WAS ATTEMPTED TO BE ADVANCED, RESISTANCE WAS FELT. UPON REMOVAL OF THE STENT, IT WAS NOTED TO BE DEFORMED. THE PROCEDURE WAS COMPLETED WITH A 2.75 X 38MM NON-MEDTRONIC STENT. THE PATIENT IS ALIVE WITH NO INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18693106 ↗
  • InjuryFeb 8, 2024Device Dislodged or Dislocated; Material Deformation; Positioning Problem; Device Dislodged or Dislocated; Material Deformation; Positioning Problem

    AN ATTEMPT WAS MADE TO USE ONE ONYX FRONTIER DRUG ELUTING STENT TO TREAT A LESION. IT WAS REPORTED THAT STENT DISLODGEMENT OCCURRED DURING REMOVAL FOLLOWING A FAILED DELIVERY. IT WAS DETAILED THAT THE STENT APPEARED TO GET STRIPPED OFF AS IT WAS BEING RETRIEVED BACK INTO THE GUIDE CATHETER AFTER IT FAILED TO CROSS THE LESION. THE DISLODGED STENT WAS REMOVED WITH A SNARE AS IT DISLODGED HALFWAY IN THE VESSEL AND THE GUIDE CATHETER. NO FURTHER PATIENT INJURY WAS REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 18672360 ↗
  • InjuryNov 1, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AN ATTEMPT WAS MADE TO USE ONE ONYX FRONTIER CORONARY DRUG ELUTING STENT TO TREAT A MILDLY TORTUOUS, MILDLY CALCIFIED LESION IN THE DISTAL LEFT ANTERIOR DESCENDING (LAD) ARTERY WITH 85% STENOSIS. THE DEVICE WAS INSPECTED WITH NO ISSUES NOTED. NEGATIVE PREP WAS NOT PERFORMED. THE LESION WAS PRE-DILATED. THE DEVICE PASSED THROUGH A PREVIOUSLY-DEPLOYED STENT. RESISTANCE WAS NOT ENCOUNTERED WHEN ADVANCING THE DEVICE. EXCESSIVE FORCE WAS NOT USED DURING DELIVERY. IT WAS REPORTED THAT THE LEFT ANTERIOR DESCENDING (LAD) WAS PERFORATED POST STENT DEPLOYMENT. IT WAS DETAILED THAT A 3.0X38MM ONYX FRONTIER STENT WAS DEPLOYED IN THE MID LAD. A DECISION WAS THEN MADE TO STENT DISTAL TO THE MID LAD. THE 2.75 X 38MM ONYX FRONTIER DEVICE WAS EASILY PLACED DISTALLY AND DEPLOYED. NO POST DILATION BALLOON WAS USED. A DISTAL PERFORATION OCCURRED, AND A 2.5 X 15MM BALLOON WAS USED TO SEAL OFF THE PERFORATION THE PATIENT IS REPOR TED TO BE ALIVE WITH NO INJURY.

    Patient problems: Perforation

    MAUDE report 15706377 ↗