Onyx Frontier™ ONYXNG40038UXRecalls & reported problems

Medtronic, Inc. · STENT ONYXNG40038UX ONYX 4.00X38RX · Class 3 · Implant

All models of Onyx Frontier (71)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00763000511715), received between Oct 26, 2022 and Jul 31, 2025 — classified as 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Difficult to Remove” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Remove2
  • Material Deformation2
  • Deflation Problem1
  • Detachment of Device or Device Component1
  • Device Damaged by Another Device1
  • Device Dislodged or Dislocated1
  • Positioning Problem1

Patient problems

  • Insufficient Information2
  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 31, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT DURING A PROCEDURE INVOLVING ONE ONYX FRONTIER CORONARY DRUG ELUTING STENT, THE STENT CAME OFF THE UNINFLATED BALLOON WHILE PARTIALLY IN THE CATHETER. THE STENT WAS RETRIEVED TO THE ARTERY IN THE HAND, AND THE PATIENT SUBSEQUENTLY UNDERWENT SURGERY THE NEXT DAY TO EXTRACT THE STENT. THE EVENT LED TO AN EXTENSION OF THE PATIENT'S HOSPITALIZATION. THE DEVICE WAS INSPECTED PRIOR TO USE AND NEGATIVE PREP WAS PERFORMED, BOTH WITH NO ISSUES NOTED. EXCESSIVE FORCE WAS NOT USED DURING DELIVERY. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Insufficient Information

    MAUDE report 22678111 ↗
  • MalfunctionNov 18, 2024Difficult to Remove; Device Damaged by Another Device; Material Deformation; Positioning Problem; Difficult to Remove; Device Damaged by Another Device; Material Deformation; Positioning Problem

    IT WAS REPORTED THAT AN ATTEMPT WAS MADE TO USE ONE ONYX FRONTIER CORONARY DRUG ELUTING STENT TO TREAT A LESION WHEN STENT DEFORMATION OCCURRED IN VIVO DURING POSITIONING/ADVANCEMENT. THE DEVICE WAS INSPECTED WITH NO ISSUES. NEGATIVE PREP WAS PERFORMED WITH NO ISSUES. THE LESION WAS PRE-DILATED. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT. IT WAS DETAILED THAT THE DEVICE WAS UNABLE TO CROSS THE LESION. THE STENT BALLOON WAS BEING REMOVED TO PRE-DILATE THE LESION MORE, AND WHEN REMOVING THE STENT THROUGH THE TUOHY STENT DEFORMATION WAS NOTICED. IT WAS ALSO REPORTED THAT THERE WAS DIFFICULTY REMOVING THE DEVICE PRIOR TO STENT DEPLOYMENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER 4.0X38MM ONYX FRONTIER STENT. THE PATIENT IS ALIVE WITH NO INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20706038 ↗
  • InjuryAug 16, 2023Deflation Problem; Difficult to Remove; Detachment of Device or Device Component; Deflation Problem; Difficult to Remove; Detachment of Device or Device Component

    AN ATTEMPT WAS MADE TO USE ONE ONYX FRONTIER CORONARY DRUG ELUTING STENT TO TREAT A NON-TORTUOUS, MODERATELY CALCIFIED LESION IN THE PROXIMAL RIGHT CORONARY ARTERY (RCA). THE LESION WAS PRE-DILATED. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT. RESISTANCE WAS NOT ENCOUNTERED WHEN ADVANCING THE DEVICE. EXCESSIVE FORCE WAS NOT USED DURING DELIVERY. IT WAS REPORTED THAT BALLOON DEFLATION DIFFICULTIES OCCURRED AT THE LESION SITE. IT WAS DETAILED THAT THE BALLOON WOULD NOT DEFLATE EVEN WHEN ATTEMPTS WERE MADE TO OVER INFLATE THE BALLOON AND POPPING WITH EXTRA WIRES. THE BALLOON WAS EVENTUALLY REMOVED BY PULLING. THE PATIENT IS ALIVE WITH NO FURTHER INJURY.

    Patient problems: Insufficient Information

    MAUDE report 17560327 ↗
  • MalfunctionOct 26, 2022Material Deformation; Material Deformation

    AN ATTEMPT WAS MADE TO USE ONE ONYX FRONTIER CORONARY DRUG ELUTING STENT TO TREAT A PROXIMAL VEIN GRAFT. IT WAS REPORTED THAT THE STENT FAILED TO CROSS THE LESION AND THE STENT WAS BENT. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15672522 ↗