Skip to content

Onyx Frontier™ ONYXNG45018UXRecalls & reported problems

Medtronic, Inc. · STENT ONYXNG45018UX ONYX 4.50X18RX · Class 3 · Implant

All models of Onyx Frontier (71)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00763000511746), received between Aug 15, 2025 and Aug 15, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Activation, Positioning or Separation Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Activation, Positioning or Separation Problem1
  • Detachment of Device or Device Component1
  • Device Damaged by Another Device1
  • Entrapment of Device1
  • Material Deformation1

Patient problems

  • Cardiovascular Dysfunction/Insufficiency1
  • Device Embedded In Tissue or Plaque1
  • Ischemia1
  • Myocardial Infarction1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 15, 2025Entrapment of Device; Activation, Positioning or Separation Problem; Detachment of Device or Device Component; Device Damaged by Another Device; Material Deformation; Entrapment of Device; Activation, Positioning or Separation Problem; Detachment of Device or Device Component; Device Damaged by Another Device; Material Deformation

    ONE ONYX FRONTIER CORONARY DRUG ELUTING STENT WAS IMPLANTED IN A PATIENT. APPROXIMATELY 3 MONTHS LATER THE PATIENT UNDERWENT A BALLOON ANGIOPLASTY PROCEDURE OF THE LEFT MAIN CORONARY ARTERY IN TO THE LEFT ANTERIOR DESCENDING ARTERY (LAD) WITH A 5X15MM NON COMPLIANT BALLOON. THE BALLOON BECAME STUCK WITHIN THE STENT AND IT WAS NOT POSSIBLE TO MOVE THE BALLOON OR WIRE INDEPENDENTLY. ALL WIRES AND CATHETER WAS PULLED OUT AS ONE UNIT. UPON REMOVING THE BALLOON CATHETER, IT APPEARED THAT A SEGMENT OF THE ONYX FRONTIER STENT WAS STUCK TO THE BALLOON AND WAS RETRIEVED ALONG WITH THE BALLOON. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED A RESULT OF THIS EVENT.

    Patient problems: Cardiovascular Dysfunction/Insufficiency; Device Embedded In Tissue or Plaque; Ischemia; Myocardial Infarction; No Clinical Signs, Symptoms or Conditions

    MAUDE report 22813548 ↗

Other Coronary drug-eluting stent devices

All Coronary drug-eluting stent devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page