Skip to content

Quadcut 1884006EMRecalls & reported problems

Medtronic Xomed, Inc. · BLADE 1884006EM RAD40 4MM M4 ROTATE ROHS · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00763000579227), received between Mar 10, 2023 and Mar 20, 2026 — classified as 9 malfunctions.

The most frequently reported specific device problem is “Break” (9 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 2 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break9
  • Noise, Audible2
  • Overheating of Device2
  • Material Deformation1
  • Mechanical Jam1

Patient problems

  • No Clinical Signs, Symptoms or Conditions9

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 20, 2026Break; Overheating of Device

    IT WAS REPORTED THAT 2 BLADES BROKE DURING THE PROCEDURE. ONE OF THEM, THE TIP BROKE OFF AND SPIRAL WRAP CAME OFF INSIDE THE PATIENT AS THE PHYSICIAN WAS USING IT AND HAD TO RETRIEVE IT USING TWEEZERS. THE OTHER ONE BROKE, THE ENTIRE SHAFT SEEMED TO GET TO WARM, AND THE SPIRAL SHAFT BEGAN TO COME OFF, BUT WAS REMOVED BEFORE IT BROKE OFF COMPLETELY AND CAN SEE WHERE IT IS ABOUT TO BREAK OFF. TIP OF FIRST BLADE WAS VERY HOT AND WAS ONLY USED A FEW MINUTES. WHILE USING THE SECOND BLADE, THE BLADE AND HANDPIECE BOTH BECAME HOT. THE BLADES WERE BEING USED IN OSCILLATION MODE AT NORMAL SPEED. THE PROCEDURE WAS COMPLETED WITH A THIRD BLADE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24647154 ↗
  • MalfunctionMar 20, 2026Break; Overheating of Device

    ADDITIONAL INFORMATION RECEIVED, THE HANDPIECE STARTED TO HEAT WITH IN A MINUTE PROBABLY 30 SECONDS OR LESS, IRRIGATION WAS USED (NORMAL SALINE) AND THE FLOW RATE WAS AT 3 CC/MIN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24646838 ↗
  • MalfunctionJul 1, 2024Break; Mechanical Jam; Break; Mechanical Jam

    IT WAS REPORTED THAT DURING STANDARD ENDOSCOPIC SINUS SURGERY, THE BLADE SOUNDED WEIRD AND THE TEETH WERE NOT LINING UP TO THE INSTRUMENT, POSSIBLY BENT. PRODUCT COULD NOT FUNCTION PROPERLY. THE 1ST SOUNDED WEIRD, AND THE TEETH WERE NOT LINING UP TO THE INSTRUMENT LIKE IT WAS BENT AND THE SECOND ONE WOULD NOT ROTATE. THE PROCEDURE WAS EXTENDED BY 10 MINUTES AND COMPLETED USING BACKUP PRODUCT. THERE WAS NO PATIENT INJURY. ANALYSIS OF THE PRODUCTS CONFIRMED THAT THE BLADES WERE BROKEN AND DETACHED WHEN RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19649998 ↗
  • MalfunctionJul 1, 2024Break; Material Deformation; Noise, Audible; Break; Material Deformation; Noise, Audible

    ADDITIONAL INFORMATION RECEIVED STATING THAT THE BLADES WERE BROKEN WHILE REMOVING FROM THE HANDPIECE AND NO FRAGMENTS CAME INTO THE CONTACT OF PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19647232 ↗
  • MalfunctionFeb 16, 2024Break; Break

    IT WAS REPORTED THAT BLADE TIP CAME APART DURING SURGERY. THE PROCEDURE COMPLETED WITH THE BACK UP DEVICE. THERE WAS NO DELAY IN THE SURGERY AND NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18719429 ↗
  • MalfunctionApr 7, 2023Break; Break

    IT WAS REPORTED THAT DURING PROCEDURE, BLADE COME APART. NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16702334 ↗
  • MalfunctionApr 6, 2023Break; Break

    IT WAS REPORTED THAT DURING PROCEDURE, THE BLADE CAME APART. THERE WAS NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16694017 ↗
  • MalfunctionMar 29, 2023Break; Break

    IT WAS REPORTED BY HEALTH CARE PROFESSIONAL THAT DURING PROCEDURE SHAVER TIP POPPED OUT OF ITS PLACE AND ROTATE. THERE WAS NO REPORTED PATIENT IMPACT. SHAVER FAILED TO LINE UP. THERE WAS NO REPORTED PATIENT INVOLVEMENT. ADDITIONAL INFORMATION RECEIVED AFTER FOLLOW UP THAT THE DISTAL BLADE AND SUCTION PORT WAS LOOSE, PARTIALLY DETACHED AND SPONTANEOUSLY ROTATING WHILE IN USING AND WAS NOT SHAVING TISSUE. THE BLADE TIP AND SUCTION PORT WAS PARTIALLY DETACHED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16633764 ↗
  • MalfunctionMar 10, 2023Break; Noise, Audible; Break; Noise, Audible

    A HEALTH CARE PROFESSIONAL REPORTED THAT, DURING FUNCTIONAL ENDOSCOPIC SINUS SURGERY WITH NASAL POLYPECTOMY PROCEDURE PHYSICIAN WAS USING THE BLADE IN THE PATIENT¿S MAXILLARY SINUS. A NOISE WAS HEARD AND THEY IMMEDIATELY NOTICED THAT THE TIP OF THE METAL SHAFT HAD BROKEN. THEY REMOVED THE BROKEN PIECE AND CONTINUED WITH THE PROCEDURE. THE PROCEDURE WAS COMPLETED WITH BACKUP PRODUCT. THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION RECEIVED AFTER FOLLOW UP THAT FRAGMENTS WAS DETACHED FROM THE BROKEN DEVICE AND PHYSICIAN VISUALLY REMOVED THE BROKEN TIP FROM THE PATIENT¿S SINUS. THERE IS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16520632 ↗

Other Neurological stereotaxic Instrument devices

All Neurological stereotaxic Instrument devices →