Nitron CryoConsole™Recalls & reported problems
Medtronic, Inc. · CONSOLE NITRON US · Class 3
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00763000593902), received between Jul 8, 2025 and Jul 24, 2025 — classified as 2 injuries.
2 other recalls cover devices cleared under the same FDA submission (P020045); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJul 24, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING A CRYO ABLATION PROCEDURE, AN ERROR MESSAGE WAS RECEIVED STATING THAT THE SYSTEM DETECTED AN UNEXPECTED TEMPERATURE DURING SYSTEM FLUSH. AN ERROR MESSAGE WAS ALSO RECEIVED STATING THAT THE SYSTEM DETECTED A HIGHER-THAN-EXPECTED FLOW RATE INDICATING A TREATMENT OVERFLOW. FIELD SERVICE ARRIVED TO PERFORM TROUBLESHOOTING AND RESOLVE THE ERROR. THE CASE WAS ABORTED UNDER GENERAL ANESTHESIA. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22589607 ↗InjuryJul 8, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING A CRYO CASE, AN ERROR MESSAGE WAS RECEIVED STATING THAT THE SYSTEM DETECTED A HIGHER THAN EXPECTED FLOW RATE INDICATING A TREATMENT OVERFLOW. THE DIRECTION OF THE COAXIAL UMBILICAL WAS SWAPPED, THE COAXIAL UMBILICAL AND CATHETER WERE REPLACED, AND THE CONSOLE WAS REBOOTED WITHOUT RESOLUTION. TEST ABLATIONS ON THE CONSOLE RESULTED IN CONTINUED N-001 SYSTEM NOTICES. THE PROCEDURE WAS ABORTED AND THE PATIENT WAS UNDER GENERAL ANESTHESIA. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22434240 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P020045. One submission often covers many models, so check the product and lot details in each record.
- Z-2777-2015 · Sep 1, 2015 · TerminatedMedtronic CryoConsole, models 106A3, 106E2, and 106A2-K For use in performing cardiac ablation procedures.
- Z-1452-06 · Aug 4, 2006 · Terminated7F Freezor Cardiac Cryoablation catheter, REF # 207F1. CryoCath Technologies Inc.
Other Cardiac ablation percutaneous catheter devices
- THERMOCOOL SMARTTOUCH SF D134805Biosense Webster Inc · Cardiac ablation percutaneous catheter1 recall176 deaths4,106 reports
- THERMOCOOL SMARTTOUCH SF D134804Biosense Webster Inc · Cardiac ablation percutaneous catheter44 deaths784 reports
- THERMOCOOL SMARTTOUCH SF D134801Biosense Webster Inc · Cardiac ablation percutaneous catheter15 deaths608 reports
- ThermoCool SmartTouch D132705Biosense Webster Inc · Cardiac ablation percutaneous catheter25 deaths574 reports
- THERMOCOOL SMARTTOUCH SF D134702Biosense Webster Inc · Cardiac ablation percutaneous catheter19 deaths366 reports
- THERMOCOOL SMARTTOUCH SF D134701Biosense Webster Inc · Cardiac ablation percutaneous catheter11 deaths278 reports
- ThermoCool SmartTouch D133602Biosense Webster Inc · Cardiac ablation percutaneous catheter8 deaths225 reports
- SMARTABLATE SAT001Biosense Webster Inc · Cardiac ablation percutaneous catheter185 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page