SYNCHROMED® II 8637-20 (UDI 00763000634094)Recalls & reported problems
Medtronic, Inc. · PUMP 8637-20 SM II 20ML EMAN SYMBL · Class 3 · Implant
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Medtronic SynchroMed II implanted drug pump has 2,220 reports, which include 3 deaths and 1,210 injuries. Reports describe communication problems, the pump moving out of place, and flow problems, and some mention pain, infection or spinal fluid leaks. 1 recall is open, from November 2023, over the pump possibly not restarting drug delivery after an MRI scan if it was sounding an alarm.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 2,242
- Deaths reported
- 3
- Injuries reported
- 1,215
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2,242 adverse event reports that list this exact device (DI 00763000634094), received between Sep 14, 2022 and Sep 30, 2026 — classified as 3 deaths, 1,215 injuries, 1,024 malfunctions.
The most frequently reported specific device problem is “Communication or Transmission Problem” (487 reports).
1 FDA recall record name this device's identifier, 1 still open.
39 other recalls cover devices cleared under the same FDA submission (P860004); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Nov 13, 2023Open, Classified
Medtronic recently identified that if the SynchroMed II pump switches into telemetry mode due to electromagnetic interference (EMI) from an MRI scan, while the pump is sounding an alarm, the pump will not resume drug delivery after leaving the MRI magnetic field, which is inconsistent with the current labeling.
Z-0614-2024 · Root cause: Under Investigation by firm · FDA record ↗
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem611
- Communication or Transmission Problem487
- Migration or Expulsion of Device278
- Insufficient Device Problem Information248
- Infusion or Flow Problem239
- Material Integrity Problem162
- Insufficient Flow or Under Infusion156
- Electromagnetic Interference133
Patient problems
- No Clinical Signs, Symptoms or Conditions1,197
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available365
- Pain320
- Unspecified Infection179
- Insufficient Information146
- Cerebrospinal Fluid Leakage82
- Seroma77
- Discomfort64
Recent adverse event reports
The 25 most recent of 2,242 reports, in the reporter's own words as published by the FDA.
InjurySep 30, 2026Electromagnetic Interference; Insufficient Device Problem Information
Patient problems: Anxiety; Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Diaphoresis; Insufficient Information; Malaise; Pain
MAUDE report 50950141 ↗InjurySep 29, 2026Electromagnetic Interference; Pumping Stopped
Patient problems: Pain
MAUDE report 50940883 ↗InjurySep 29, 2026Insufficient Flow or Under Infusion
Information was received from a patient receiving Dilaudid via an implantable infusion pump for spinal pain indications. It was reported that for the past 3 years, every 4th of July, their pump 'runs dry for a whole week' and the patient goes through withdrawals.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 50936662 ↗InjurySep 29, 2026Adverse Event Without Identified Device or Use Problem
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Cognitive Changes; Pain
MAUDE report 50934611 ↗InjurySep 28, 2026Adverse Event Without Identified Device or Use Problem
On (b)(6)2026, information was received from a Manufacturer Representative (REP), regarding a patient receiving baclofen (2,000 mcg /ml, 529.5 mcg/day) via an implantable pump. It was reported that the patient reported to the emergency department with symptoms of baclofen withdrawal. There were no known environmental, external, or patient factors that may have led or contributed to the issue. The diagnostics / troubleshooting performed was the catheter was aspirated through the side port. The full system was replaced as an intervention. It was unknown if the issue was resolved at the time of this report.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 50918083 ↗MalfunctionSep 28, 2026Failure to Power Up; Charging Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50915515 ↗InjurySep 25, 2026Migration or Expulsion of Device; Insufficient Flow or Under Infusion; Obstruction of Flow; Infusion or Flow Problem
Patient problems: Pain
MAUDE report 50902981 ↗MalfunctionSep 25, 2026Display or Visual Feedback Problem; Charging Problem
Information was received from a consumer regarding a patient with an implantable pump containing unknown drug for non-malignant pain. It was reported that the patient's communicator was not charging, They patient (pt) contacted their doctor but was told to call patient services. Pt said the issue started about 6 months ago. Pt already tried to use different charger and outlet, but the communicator still won't charge. Pt confirmed no visible damage on the port of the communicator. When they powered on the communicator, they got the blue flashing light but no orange light on the battery indicator when plugged in to the charging cord. Pt stated that the same charging cord that they were using for the handset and communicator was working to their other device. Troubleshooting steps taken on the call didn't resolve the issue. Agent sent a request to replace the communicator.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50898556 ↗MalfunctionSep 25, 2026Application Program Freezes or Fails to Launch
Information was received from a patient, receiving an unknown drug via an implantable infusion pump for non-malignant pain indications for use. It was reported that the patient's handset was lagging at 90% and it was taking about 10 minutes to deliver a bolus. The patient mentioned that they had met the manufacturer's representative (Rep) at the doctor's office and were instructed to call patient services. When asked for event date, the patient stated that the issue has been ongoing for a while since they had the pump. The Agent reviewed information and walked the patient through clearing the data. The patient was able to connect to the pump with no lag. The patient mentioned that they just delivered a bolus before they called. The Agent reviewed best practices and instructed the patient to call back if they needed further assistance.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50897315 ↗MalfunctionSep 23, 2026Application Program Freezes or Fails to Launch
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50876305 ↗InjurySep 23, 2026Insufficient Flow or Under Infusion; Obstruction of Flow; Infusion or Flow Problem
Patient problems: Insufficient Information
MAUDE report 50874024 ↗MalfunctionSep 22, 2026Failure to Power Up; Charging Problem; Communication or Transmission Problem; Protective Measures Problem
Information was received from the patient receiving intrathecal Morphine (unknown) via an implantable pump for spinal pain indications. The patient reported that the communicator would not charge and now it was dead. Patient said she uses it every day several times a day and needed it by tomorrow. Agent asked what happens when patient attempts a bolus and patient said they get a message to deliver bolus but could not. Patient also said the charging port was broken and they could not get the cord to go in. Patient tried a new cord and that did not resolve the issue. The issue was not resolved through troubleshooting. A request was sent to the repair department. Patient called back stating they still did not get their new equipment and they had had several hours without pain medicine now.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50858201 ↗MalfunctionSep 18, 2026Data Problem
Additional information was received from the manufacturing representative (rep). The rep reported that the cause of the non-critical pump memory error was not determined. No actions were taken to resolve the issue. There were results to provide regarding diagnosti cs or troubleshooting. It was just the patient reporting that they saw an error. The issue was resolved.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50828920 ↗InjurySep 17, 2026Migration or Expulsion of Device
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Emotional Changes; Shaking/Tremors
MAUDE report 50812653 ↗InjurySep 17, 2026Positioning Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50811501 ↗InjurySep 16, 2026Insufficient Device Problem Information
Information was received from a patient's representative regarding an external device to an implantable pump infusing unknown drug and unknown Morphine for non-malignant pain. It was reported that the patient stated they had back surgery on Thursday ((b)(6) 2026).The historical event reported. The patient stated that last Thursday ((b)(6) 2026), they had to have a catheter revision. The patient also stated they had to have the "tubing rerouted".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50792018 ↗MalfunctionSep 15, 2026Failure to Power Up; Adverse Event Without Identified Device or Use Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50775367 ↗MalfunctionSep 14, 2026Failure to Power Up; Charging Problem; Communication or Transmission Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50763383 ↗MalfunctionSep 14, 2026Communication or Transmission Problem
Information was received from the patient receiving intrathecal Dilaudid (hydromorphone) via an implantable pump for non-malignant pain. It was reported that for the past 2 months or maybe more the patient had been having issues with their handset. Patient stated that the handset was getting stuck at 90% when it was loading and they had to wait 2 minutes and 13 seconds for the handset to respond to the request for a bolus. During the call the patient service specialist walked the patient through clearing data services and connecting to the pump. Patient was able to successfully connect and request/receive a bolus. Agent reviewed proper communicator positioning. Patient would monitor the issue and call back if needed.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50755819 ↗MalfunctionSep 14, 2026Failure to Power Up; Insufficient Device Problem Information
Information was received from the patient receiving intrathecal medication via an implantable pump for non-malignant pain. The patient reported that for the past couple of weeks the communicator would not work consistently. The patient stated it would just cut off while she was trying to get a bolus. The patient stated now the communicator would not come on and it would not respond to the charging cord. Patient did try different charging cords but that did not resolve the issue. See request to repair for the communicator.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50754644 ↗MalfunctionSep 14, 2026Pumping Stopped
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50752895 ↗MalfunctionSep 11, 2026Failure to Power Up; Charging Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50735177 ↗MalfunctionSep 10, 2026Communication or Transmission Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50713808 ↗MalfunctionSep 9, 2026Electromagnetic Interference
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50697870 ↗MalfunctionSep 9, 2026Communication or Transmission Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50694164 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P860004. One submission often covers many models, so check the product and lot details in each record.
- Z-2201-2026 · Apr 8, 2026 · Open, ClassifiedMedtronic SynchroMed Flex Infusion Mode - A810 Clinician Programmer Software Application, loaded on the following devices: 1. PROG CT900A CLINICIAN TABLET US, Model Number CT900
- Z-1506-2025 · Jan 9, 2025 · Open, ClassifiedA820 myPTM Software Application associated with Medtronic SynchroMed Pump and Infusion System.
- Z-2352-2024 · May 10, 2024 · Open, ClassifiedMedtronic Ascenda Intrathecal Catheter, Models: a) 8780 (114 cm); b) 8781 (140 cm); c) 8784 (catheter pump revision kit)
- Z-0615-2024 · Nov 13, 2023 · Open, ClassifiedMedtronic SynchroMed II, Model 8637-40, Programmable pump
- Z-0522-2024 · Jul 13, 2020 · CompletedMedtronic SynchroMed, Model A10
- Z-0508-2020 · Oct 11, 2019 · TerminatedSynchromed II programmable pump, Model numbers 8637-20 and 8637-40
- Z-0537-2018 · Sep 21, 2017 · TerminatedMedtronic Ascenda(R) Intrathecal Catheter, Models: (a) 8780, UDI: 00643169202122, 00643169783027 (b) 8781, UDI: 00643169202139 (c) 8784, UDI: 00643169202153 The Ascenda Intrath
- Z-0896-2018 · Aug 9, 2017 · Open, ClassifiedMedtronic Synchromed II Model 8637 supplied in 20mL and 40 mL reservoir size and a catheter. Product Usage: The implantable Model 8637 SynchroMed II programmable pump is part of a
- Z-0788-2017 · Oct 3, 2016 · TerminatedMedtronic SynchroMed II Implantable Drug Infusion System. This system includes: Model 8870 Software Application Card, Model 8840 N'Vision Clinician Programmer, and Model 8637 Sy
- Z-1271-2016 · Feb 9, 2016 · TerminatedMedtronic, Model 8781 Ascenda" Intrathecal Catheter kit, Rx Only, Sterile Ethylene Oxide.
- Z-1681-2015 · Apr 10, 2015 · TerminatedMedtronic SynchroMed II Implantable Infusion Pumps, models 8637-20 and 8637-40. Sterile. Product Usage: The implantable Medtronic SynchroMed II programmable pumps are part of an
- Z-1062-2011 · Jan 14, 2011 · TerminatedMedtronic SynchroMed EL (Model No: 8626 and 8627) implantable drug pump The SynchroMed EL Infusion System is indicated for use when patient therapy requires the chronic infusion
- Z-1060-2011 · Jan 14, 2011 · TerminatedMedtronic SynchroMed II (Model No: 8637) implantable drug pump The SynchroMed II Programmable Pump is indicated when patient therapy requires the chronic infusion of the follow
- Z-1061-2011 · Jan 14, 2011 · TerminatedMedtronic SynchroMed implantable drug pump Refill Kits (Model No: 8551, 8555, 8561, 8562, 8564, 8565, 8566 The Model 8551, 8555, 8561, 8562, 8564, 8565, and 8566 Refill Kits ar
- Z-2276-2009 · Jul 1, 2009 · TerminatedMedtronic SynchroMed II Programmable Pump, 8637, Rx Only, Sterilized by ethylene oxide gas, Medtronic inc., Minneapolis, MN 55432-5604. (The SynchroMed II Pump is supplied in 20 m
…and 24 more.
Other Pump, infusion, implanted, programmable devices
- SYNCHROMED® II 8637-40Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE1 recall7 deaths4,929 reports
- SYNCHROMED® II 8637-20Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE2 recalls5 deaths4,303 reports
- SYNCHROMED ® II 8637-40Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE1 recall6 deaths4,119 reports
- SYNCHROMED® II 8637-40Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE2 recalls10 deaths3,054 reports
- SYNCHROMED ® II 8637-20Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE1 recall4 deaths3,084 reports
- SYNCHROMED® II 8637-20Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE2 recalls1 death2,588 reports
- SynchroMed® II 8637-40Medtronic, Inc. · PUMP, INFUSION, IMPLANTED, PROGRAMMABLE1 recall6 deaths2,291 reports
- SYNCHROMED® II 8637-40Medtronic, Inc. · Pump, infusion, implanted, programmable1 recall4 deaths1,964 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.