Skip to content

NIM Contact 8229507Recalls & reported problems

Medtronic Xomed, Inc. · EMG TUBE 8229507 NIM CONT 7MM ROHS · Class 2

All models of NIM Contact (6)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00763000745851), received between Mar 18, 2024 and Mar 18, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 29, 2022Class IOpen, Classified

    The firm received reports of events related to airway obstruction while using NIM Standard Reinforced EMG Endotracheal Tubes & NIM CONTACT Reinforced EMG Endotracheal Tubes.

    Z-1519-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break1

Patient problems

  • Hemorrhage/Blood Loss/Bleeding1
  • Laceration(s)1
  • Unintended Extubation1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 18, 2024Break; Break

    IT WAS REPORTED THAT DURING THE THYROIDECTOMY PROCEDURE NIM TUBE STYLET WAS PUT INTO NIM TUBE TO FACILITATE THE CMAC INTUBATION. UPON CURVATURE, WE DIDN'T NOTICE THE NIM WIRE DETACH. THE WIRE WAS CAUGHT IN THE HYPOPHARYNX AND PREVENTED THE NIM TUBE FROM GOING INTO THE TRACHEA. THE NIM TUBE WAS REMOVED AND A REGULAR 7.0 ET TUBE WAS USED INSTEAD TO INTUBATE THE PATIENT. UPON REMOVAL OF THE NIM TUBE, FRESH BLOOD WAS NOTICED AT THE INSERTION TIP OF THE NIM TUBE. THE WIRE WAS CAUGHT IN THE HYPOPHARYNX AND CAUSED BLEEDING. THE PATIENT IS OK FOR THE TIME BEING ACCORDING TO THE CHARGE NURSE. THE EVENT WAS RESULTED IN A PROCEDURE DELAY. THE PATIENT WAS ALIVE - WITH INJURY.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Laceration(s); Unintended Extubation

    MAUDE report 18919359 ↗

Other STIMULATOR, NERVE devices

All STIMULATOR, NERVE devices →

Sources and updates