NanoCross™ Elite AB14W040150150 (UDI 00763000758004)Recalls & reported problems
Medtronic, Inc. · PTA AB14W040150150 NANOCROSS 014 V06 · Class 2
All models of NanoCross Elite (31)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00763000758004), received between Nov 3, 2025 and Nov 12, 2025 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Deflation Problem” (2 reports).
1 other recall cover devices cleared under the same FDA submission (K141118); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Deflation Problem2
- Compatibility Problem1
- Deformation Due to Compressive Stress1
- Detachment of Device or Device Component1
- Difficult to Remove1
- Inflation Problem1
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Insufficient Information1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryNov 12, 2025Deflation Problem; Inflation Problem; Deformation Due to Compressive Stress; Detachment of Device or Device Component; Deflation Problem; Inflation Problem; Deformation Due to Compressive Stress; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY PROCEDURE USING THE NANOCROSS 014, DEFLATION DIFFICULTIES OCCURRED WHEN THE BALLOON WOULD NOT DEFLATE. A SMALL KINK IN THE CATHETER WAS NOTICED AFTER REMOVAL. THE PROCEDURE WAS ABORTED. THE TARGET LESION WAS LOCATED IN THE PERIPHERAL ARTERIES, SPECIFICALLY THE COMMON ILIAC ARTERY, SUPERFICIAL FEMORAL ARTERY, AND POPLITEAL ARTERY, WITH AN ARTERY DIAMETER OF 6MM AND CALCIFICATION PRESENT. THE VESSEL WAS PRE-DILATED AND THE DEVICE WAS PASSED THROUGH A PREVIOUSLY DEPLOYED STENT. THE BALLOON WAS INFLATED USING AN INFLATION DEVICE, AND DEFLATION WAS ATTEMPTED BY SNARING THROUGH A TRAILBLAZER. NO ABNORMALITIES WERE REPORTED IN RELATION TO ANATOMY, AND INSTRUCTIONS FOR USE WERE FOLLOWED WITHOUT ISSUE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 23533172 ↗MalfunctionNov 3, 2025Deflation Problem; Difficult to Remove; Compatibility Problem; Material Deformation; Deflation Problem; Difficult to Remove; Compatibility Problem; Material Deformation
IT WAS REPORTED THAT DURING TREATMENT OF A SOFT TISSUE LESION IN THE MID LEFT POSTERIOR TIBIAL ARTERY USING THE PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY NANOCROSS 014, THE BALLOON INFLATED AND TRACKED INTO THE BODY WITHOUT ISSUE, BUT DEFLATION DIFFICULTIES OCCURRED AS THE BALLOON WOULD NOT FULLY DEFLATE FOR APPROXIMATELY TWO MINUTES. THE BALLOON CATHETER APPEARED TO STRETCH DURING REMOVAL AND BECAME STUCK ON THE WIRE, REQUIRING BOTH THE BALLOON CATHETER AND WIRE TO BE REMOVED TOGETHER. AFTER REMOVAL FROM THE BODY, THE WIRE COULD NOT BE SEPARATED FROM THE CATHETER, AS IT WAS STUCK INSIDE, AND THE BALLOON CATHETER APPEARED STRETCHED AND DETACHED. THE PROCEDURE WAS COMPLETED USING A NEW DEVICE OF THE SAME MAKE AND MODEL. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23450685 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K141118. One submission often covers many models, so check the product and lot details in each record.
- Z-0904-2025 · Dec 10, 2024 · Open, ClassifiedMedtronic NanoCross Elite PTA catheter, REF numbers amd sizes: 1) AB14W050020090, Size: 5MM X 20MM X 90CM; 2) AB14W050020150, Size: 5MM X 20MM X 150CM; 3) AB14W050040090,
Other Catheter, percutaneous devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page