Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00763000766504), received between May 15, 2025 and Oct 29, 2025 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Burst Container or Vessel1
  • Component Missing1
  • Difficult to Remove1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 29, 2025Component Missing; Component Missing

    IT WAS REPORTED THAT DURING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY PROCEDURE USING THE EVERCROSS 035 FOR TREATMENT OF A PLAQUE LESION IN THE MID LEFT COMMON ILIAC ARTERY, IT WAS NOTED THAT THE DEVICE WAS MISSING THE PROXIMAL BALLOON MARKER. THE SHEATH USED WAS A 7FR NON-MEDTRONIC SHEATH, AND THE VESSEL WAS BOTH PRE-DILATED WITH AN EVERCROSS 6X40 AND POST-DILATED WITH AN EVERCROSS 7X40. NO RESISTANCE WAS ENCOUNTERED WHEN ADVANCING THE DEVICE, AND THE PHYSICIAN PROCEEDED WITHOUT ISSUE. THE DEVICE WAS SAFELY REMOVED FROM THE PATIENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23412779 ↗
  • InjuryMay 15, 2025Burst Container or Vessel; Difficult to Remove; Burst Container or Vessel; Difficult to Remove

    A PATIENT PRESENTED FOR A PLANNED RIGHT LOWER EXTREMITY ANGIOGRAM WITH ANGIOPLASTY. DURING THE PROCEDURE, 2 EVERCROSS PTA BALLOON CATHETERS RUPTURED ON IRREGULAR PLAQUE LEADING TO SEEMING OCCLUSION OF THE DEEP FEMORAL BRANCH. THE FIRST EVERCROSS BALLOON AB35W06060135 HAD A PIN HOLE RUPTURE. THE SECOND EVERCROSS BALLOON AB35W06040135 ALSO HAD REMOVAL DIFFICULTIES. THERE WAS CONCERN FOR POTENTIAL RETAINED BALLOON MATERIAL. THE PROCEDURE WAS ABORTED, AND A VASCULAR SURGEON WAS CONSULTED. THE PATIENT WAS URGENTLY TAKEN TO THE OPERATING ROOM SUITE FOR INTERVENTION. NO BALLOON FRAGMENTS WERE IDENTIFIED. AT THE TIME OF THE BALLOONS RUPTURING, THE INTERVENTIONAL RADIOLOGIST WAS NOT UTILIZING THE INSUFFLATOR AND ONLY USING APPROXIMATELY 8MM HG OF HAND PRESSURE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22025716 ↗