NIM Essence NIMEID080Recalls & reported problems

Medtronic Xomed, Inc. · ESSENCE EMG TUBE NIMEID080 5PK 8MM CE · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00763000821814), received between Aug 7, 2026 and Aug 13, 2026 — classified as 2 injuries, 5 malfunctions.

The most frequently reported specific device problem is “Difficult to Insert” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 7 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520267
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Insert6
  • Incorrect, Inadequate or Imprecise Result or Readings6
  • Inappropriate Audible Prompt/Feedback1

Patient problems

  • Insufficient Information4
  • Unintended Extubation2
  • Unspecified Tissue Injury2
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 13, 2026Inappropriate Audible Prompt/Feedback

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50231734 ↗
  • MalfunctionAug 11, 2026Difficult to Insert; Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: Insufficient Information

    MAUDE report 50204995 ↗
  • MalfunctionAug 11, 2026Difficult to Insert; Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: Insufficient Information

    MAUDE report 50204972 ↗
  • InjuryAug 8, 2026Difficult to Insert; Incorrect, Inadequate or Imprecise Result or Readings

    It was reported that Patient having the removal of a right hemi thyroid. The anesthetist used the Nim Essence tubes and spent over 30 minutes trying to intubate. The anesthetist had the Nim Vital on the setup electrode check screen and they said they spent all this time trying to achieve the green confirmation to proceed ticks, Every slight movement would mean that the ticks would go back to the red cross. So small adjustments slightly left then 5mm slight right to get the tube to be sat in the correct position and the electrode check kept changing from green ticks back to red ticks with no room for any slight error.  They also re-intubated the patient with a second tube to have the same issue that only resolved when they used the Nim Flex tube. There was 30 min delay in the procedure. The patient suffered soft palet injury due to repeat re-intubation.

    Patient problems: Unintended Extubation; Unspecified Tissue Injury

    MAUDE report 50187129 ↗
  • InjuryAug 8, 2026Difficult to Insert; Incorrect, Inadequate or Imprecise Result or Readings

    It was reported that Patient having the removal of a right hemi thyroid. The anesthetist used the Nim Essence tubes and spent over 30 minutes trying to intubate. The anesthetist had the Nim Vital on the setup electrode check screen and they said they spent all this time trying to achieve the green confirmation to proceed ticks, Every slight movement would mean that the ticks would go back to the red cross. So small adjustments slightly left then 5mm slight right to get the tube to be sat in the correct position and the electrode check kept changing from green ticks back to red ticks with no room for any slight error.  They also re-intubated the patient with a second tube to have the same issue that only resolved when they used the Nim Flex tube. There was 30 min delay in the procedure. The patient suffered soft palet injury due to repeat re-intubation.

    Patient problems: Unintended Extubation; Unspecified Tissue Injury

    MAUDE report 50187128 ↗
  • MalfunctionAug 7, 2026Difficult to Insert; Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: Insufficient Information

    MAUDE report 50171814 ↗
  • MalfunctionAug 7, 2026Difficult to Insert; Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: Insufficient Information

    MAUDE report 50171153 ↗

Other STIMULATOR, NERVE devices

All STIMULATOR, NERVE devices →