Skip to content

HexaGen™ RF Generator AFR-00004Recalls & reported problems

Medtronic, Inc. · GENERATOR AFR-00004 HEXAGEN RF US · Class 3

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00763000871864), received between Sep 10, 2025 and Mar 5, 2026 — classified as 2 injuries.

1 other recall cover devices cleared under the same FDA submission (P240013); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 5, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING A CARDIAC ABLATION PROCEDURE, A CATHETER TEMPERATURE TOO LOW ERROR OCCURRED. THE CATHETER AND CATHETER CABLE WERE REPLACED, REBOOTING OF THE STACK AND WORKSTATION PERFORMED, AND THE LINK CABLES WERE RECONNECTED, BUT NONE OF THESE TROUBLESHOOTING STEPS RESOLVED THE ERROR. THE CASE WAS ABORTED. TECHNICAL SERVICES RECOMMENDED REPLACEMENT OF THE RADIOFREQUENCY GENERATOR COVERED UNDER WARRANTY. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24523871 ↗
  • InjurySep 10, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS FURTHER REPORTED THAT PULSED FIELD ABLATIONS WERE BEING UTILIZED AROUND THE LEFT PULMONARY VEINS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23017142 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P240013. One submission often covers many models, so check the product and lot details in each record.

  • Z-1948-2026 · Mar 11, 2026 · Open, Classified
    Medtronic Sphere-9 Catheter, Model Number AFR-00001; catheter, percutaneous, cardiac ablation, for treatment of atrial flutter.

Other Percutaneous Cardiac Ablation Catheter for Treatment of Atrial Fibrillation with Irreversible Electroporation devices

All Percutaneous Cardiac Ablation Catheter for Treatment of Atrial Fibrillation with Irreversible Electroporation devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page