Skip to content

HexaFlow™ Irrigation Pump AFR-00005Recalls & reported problems

Medtronic, Inc. · IRRIGATION PUMP AFR-00005 HEXAFLOW US · Class 3

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00763000871895), received between Jul 8, 2025 and Nov 19, 2025 — classified as 2 injuries.

The most frequently reported specific device problem is “Device Alarm System” (1 report).

1 other recall cover devices cleared under the same FDA submission (P240013); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Alarm System1
  • Display or Visual Feedback Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 19, 2025Device Alarm System; Device Alarm System

    IT WAS REPORTED THAT DURING AN ABLATION PROCEDURE, A PUMP EXHIBITED HARDWARE ERROR 074 AND REPEATEDLY REBOOTED, RESULTING IN THE ABORTION OF THE PROCEDURE WHILE THE PATIENT WAS UNDER GENERAL ANESTHESIA. THE PATIENT WAS CARDIOVERTED AND DID NOT RECEIVE ALTERNATE THERAPY FROM ANOTHER PRODUCT. A SERVICE VISIT OCCURRED ON A LATER DATE AND THE PUMP WAS REPLACED. THE PUMP WAS TESTED TO VERIFY FUNCTIONALITY AND AN EXTENDED SELF TEST WAS COMPLETED. NO ADVERSE PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23598785 ↗
  • InjuryJul 8, 2025Display or Visual Feedback Problem; Display or Visual Feedback Problem

    IT WAS REPORTED THAT DURING A CARDIAC ABLATION PROCEDURE, THE IRRIGATION PUMP'S TOUCHSCREEN FROZE. REBOOTS DID NOT RESOLVE THE ISSUE. ABLATIONS IN THE LEFT ATRIUM WERE COMPLETED AND THE CAVOTRICUSPID ISTHMUS (CTI) LINE IN THE RIGHT ATRIUM WAS NOT COMPLETED. THE PROCEDURE WAS ABORTED AND THE PATIENT WAS UNDER GENERAL ANESTHESIA. SERVICE WAS PERFORMED AND THE IRRIGATION PUMP WAS REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22439863 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P240013. One submission often covers many models, so check the product and lot details in each record.

  • Z-1948-2026 · Mar 11, 2026 · Open, Classified
    Medtronic Sphere-9 Catheter, Model Number AFR-00001; catheter, percutaneous, cardiac ablation, for treatment of atrial flutter.

Other Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation devices

All Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page