Skip to content

NIM 8229507Recalls & reported problems

Medtronic Xomed, Inc. · EMG TUBE 8229507 NIM CONT 7MM ROHS · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00763000882426), received between May 27, 2024 and Mar 13, 2025 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Failure to Deflate” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 29, 2022Open, Classified

    The firm received reports of events related to airway obstruction while using NIM Standard Reinforced EMG Endotracheal Tubes & NIM CONTACT Reinforced EMG Endotracheal Tubes.

    Z-1519-2022 · Root cause: Under Investigation by firm · FDA record ↗

Most reported problems

Device problems

  • Failure to Deflate1
  • Incorrect, Inadequate or Imprecise Result or Readings1
  • Obstruction of Flow1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Airway Obstruction1
  • Unintended Extubation1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 13, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT THE ENDOTRACHEAL TUBE DOES NOT RECOGNIZE VOCAL ON THE NEURO MONITOR. THE MONITOR WAS RESTARTED SEVERAL TIMES WITHOUT SUCCESS. THE ENDOTRACHEAL TUBE WAS ROTATED SEVERAL TIMES WITHOUT SUCCESS. THERE WAS NO PATIENT IMPACT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21593922 ↗
  • MalfunctionJun 21, 2024Failure to Deflate; Failure to Deflate

    IT WAS REPORTED THAT WHEN CHECKING THE PATENCY OF THE CUFF BEFORE THE PROCEDURE, THE ANESTHETIST WAS UNABLE TO DEFLATE THE CUFF. THE ISSUE OCCURRED AS OUT OF BOX, FIRST USE FAILURE. THE CUFF CAME INFLATED AND WOULD NOT DEFLATE. ANOTHER TUBE WAS USED FOR THE PROCEDURE. THERE WAS NO PATIENT INVOLVED IN THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19580320 ↗
  • InjuryMay 27, 2024Obstruction of Flow; Obstruction of Flow

    IT WAS REPORTED THAT DURING RIGHT HEMI THYROIDECTOMY ANESTHESIOLOGIST STATED TUBE BLOCKED PATIENTS AIRWAY AND NEEDED TO REDO THE INT UBATION. THE PROCEDURE WAS COMPLETED WITH BACKUP PRODUCT. PATIENT ALIVE WITH NO INJURY.

    Patient problems: Airway Obstruction; Unintended Extubation

    MAUDE report 19399284 ↗

Other STIMULATOR, NERVE devices

All STIMULATOR, NERVE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page