Skip to content

HexaPulse™ PF Generator AFR-00008 (UDI 00763000998363)Recalls & reported problems

Medtronic, Inc. · PF GEN AFR-00008 HEXAPULSE US PLX · Class 3

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00763000998363), received between Feb 23, 2026 and Feb 23, 2026 — classified as 1 injury.

1 other recall cover devices cleared under the same FDA submission (P240013); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 23, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING A CARDIAC ABLATION PROCEDURE, AN ERROR MESSAGE STATING "NO SOUND DURING ABLATION" WAS DISPLAYED WHEN ATTEMPTING TO DELIVER A PULSED FIELD (PF) LESION ON THE POSTERIOR WALL/FLOOR LINE AFTER COMPLETION OF THE PULMONARY VEIN ISOLATION (PVI). THE ERROR PREVENTED FURTHER ABLATION AND COULD NOT BE CLEARED. THE SYSTEM WAS RESTARTED, AND THE POWER ON TEST RESULT INDICATED "RFG (RADIOFREQUENCY GENERATOR) TESTS PASSED, PFG (PULSED FIELD GENERATOR) WAS POWERED OFF." UPON A SECOND RESTART, THE SAME ERROR REGARDING PFG POWERED OFF WAS OBSERVED. DUE TO TIME CONSTRAINTS, THE PHYSICIAN DECIDED NOT TO RESTART THE SYSTEM AGAIN, AND THE PROCEDURE WAS ABORTED. WHILE THE PATIENT WAS UNDER GENERAL ANESTHESIA. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24421221 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P240013. One submission often covers many models, so check the product and lot details in each record.

  • Z-1948-2026 · Mar 11, 2026 · Open, Classified
    Medtronic Sphere-9 Catheter, Model Number AFR-00001; catheter, percutaneous, cardiac ablation, for treatment of atrial flutter.

Other Percutaneous Cardiac Ablation Catheter for Treatment of Atrial Fibrillation with Irreversible Electroporation devices

All Percutaneous Cardiac Ablation Catheter for Treatment of Atrial Fibrillation with Irreversible Electroporation devices →

Sources and updates