Skip to content

ESTEEM 423665Recalls & reported problems

Convatec, Purchasing Department · N/A · Class 1

All models of ESTEEM (52)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00768455204144), received between Dec 11, 2025 and Dec 11, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Sharp Edges” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Feb 19, 2024Class IIOpen, Classified

    Convatec Inc is conducting a Voluntary Medical Device Recall (removal) for specific lots of the EsteemBody Soft Convex, Drainable Pouch due to inconsistency in the strength of the filter weld.

    Z-1350-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Sharp Edges1

Patient problems

  • Hemorrhage/Blood Loss/Bleeding1
  • Laceration(s)1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 11, 2025Sharp Edges; Sharp Edges

    TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Laceration(s)

    MAUDE report 23777084 ↗

FDA clearance and classification

The FDA regulates product code FON (bag, drainage, with adhesive, ostomy) under 21 CFR 876.5900 as a Class I device. 21 CFR 876.5900 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the ESTEEM 423665

Has the ESTEEM 423665 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1350-2024 (Class II), began on Feb 19, 2024: Convatec Inc is conducting a Voluntary Medical Device Recall (removal) for specific lots of the EsteemBody Soft Convex, Drainable Pouch due to inconsistency in the strength of the filter weld. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the ESTEEM 423665?

The FDA's MAUDE database lists 1 adverse event report for this device, received on Dec 11, 2025. That report lists device problem “Sharp Edges” (1); patient problems “Hemorrhage/Blood Loss/Bleeding” (1) and “Laceration(s)” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the ESTEEM 423665?

None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the ESTEEM 423665 FDA approved?

Devices of this type (product code FON) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Convatec, Purchasing Department.

Who makes the ESTEEM 423665, and is it still sold?

The FDA's device database lists Convatec, Purchasing Department as the company responsible for it (brand ESTEEM, model 423665). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Bag, drainage, with adhesive, ostomy devices

All Bag, drainage, with adhesive, ostomy devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page