Skip to content

GlidePath Long-Term Hemodialysis Catheter with Preloaded Stylet 5393350Recalls & reported problems

Bard Access Systems, Inc. · GLIDEPATH HEMODIALYSIS CATH, ST, DL, 14.5 FR. 35cm · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00801741012211), received between May 3, 2018 and Nov 28, 2025 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Air/Gas in Device” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Air/Gas in Device1
  • Device Dislodged or Dislocated1
  • Fracture1

Patient problems

  • Air Embolism1
  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 28, 2025Fracture; Fracture

    A PATIENT UNDERWENT FOR DIALYSIS CATHETER PLACEMENT PROCEDURE USING GLIDEPATH. POST THE CATHETER PLACEMENT PROCEDURE, THE CATHETER CLAMP WAS ALLEGEDLY FOUND BROKEN. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23669795 ↗
  • InjuryNov 27, 2025Adverse Event Without Identified Device or Use Problem; Air/Gas in Device; Adverse Event Without Identified Device or Use Problem; Air/Gas in Device

    A PATIENT UNDERWENT A CATHETER PLACEMENT PROCEDURE BY USING GLIDEPATH. POST THE PROCEDURE, IT WAS REPORTED THAT THE PATIENT ALLEGEDLY EXPERIENCED MICROBUBBLES SEEN ON BLOODLINE. HOWEVER, THE CURRENT STATUS OF THE PATIENT WAS UNKNOWN.

    Patient problems: Air Embolism

    MAUDE report 23665242 ↗
  • MalfunctionMay 3, 2018Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE CATHETER PLACED FEMORALLY WAS SUTURED IN; HOWEVER, THE CATHETER HAS FALLEN OUT OF THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7485580 ↗

Other Catheter, hemodialysis, implanted devices

All Catheter, hemodialysis, implanted devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page