GlidePath Long-Term Hemodialysis Catheter with Preloaded Stylet 5397230Recalls & reported problems
Bard Access Systems, Inc. · GLIDEPATH HEMODIALYSIS CATH, ST, DL, 14.5 FR. 23cm · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Bard GlidePath dialysis catheter is a long-term implanted tube used for hemodialysis. Reports mostly describe the tube collapsing, kinking or flattening, sometimes wearing down where clamps are placed and needing replacement. There are 41 reports naming its identifier, mentioning 1 death and 1 injury. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 41
- Deaths reported
- 1
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 41 adverse event reports that list this exact device (DI 00801741012303), received between May 18, 2018 and Nov 30, 2025 — classified as 1 death, 1 injury, 39 malfunctions.
The most frequently reported specific device problem is “Collapse” (14 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Collapse14
- Material Deformation14
- Defective Component10
- Kinked10
- Fracture3
- Break1
- Contamination /Decontamination Problem1
- Deformation Due to Compressive Stress1
Patient problems
- No Clinical Signs, Symptoms or Conditions18
- No Consequences Or Impact To Patient14
- No Patient Involvement8
- Thrombosis/Thrombus1
Recent adverse event reports
The 25 most recent of 41 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 30, 2025Deformation Due to Compressive Stress; Deformation Due to Compressive Stress
A PATIENT WAS IMPLANTED A CATHETER PLACEMENT PROCEDURE USING GLIDEPATH. ON (B)(6) 2023, THE PORT WAS IMPLANTED AND TWO YEARS, FOUR MONTHS AND THIRTEEN DAYS POST A DIALYSIS CATHETER PLACEMENT, THE CATHETER WAS ALLEGEDLY FOUND FLATTENED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23677457 ↗MalfunctionOct 7, 2024Collapse; Material Deformation; Collapse; Material Deformation
IT WAS REPORTED THAT TWO MONTHS AND TWENTY-ONE DAYS POST A DIALYSIS CATHETER PLACEMENT, THE PLASTIC LUMEN ON THE CATHETER WAS ALLEGEDLY COLLAPSED. IT WAS FURTHER REPORTED THAT THE PLASTIC ALLEGEDLY WORE DOWN, SPECIFICALLY WHERE CLAMPS ARE EVEN WHEN ROTATED. REPORTEDLY, THE CATHETER WAS REPLACED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20392996 ↗MalfunctionOct 7, 2024Collapse; Material Deformation; Collapse; Material Deformation
IT WAS REPORTED THAT TWO MONTHS AND TWENTY-ONE DAYS POST A DIALYSIS CATHETER PLACEMENT, THE PLASTIC LUMEN ON THE CATHETER WAS ALLEGEDLY COLLAPSED. IT WAS FURTHER REPORTED THAT THE PLASTIC ALLEGEDLY WORE DOWN, SPECIFICALLY WHERE CLAMPS ARE EVEN WHEN ROTATED. REPORTEDLY, THE CATHETER WAS REPLACED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20392969 ↗MalfunctionOct 7, 2024Collapse; Material Deformation; Collapse; Material Deformation
IT WAS REPORTED THAT TWO MONTHS AND TWENTY-ONE DAYS POST A DIALYSIS CATHETER PLACEMENT, THE PLASTIC LUMEN ON THE CATHETER WAS ALLEGEDLY COLLAPSED. IT WAS FURTHER REPORTED THAT THE PLASTIC ALLEGEDLY WORE DOWN, SPECIFICALLY WHERE CLAMPS ARE EVEN WHEN ROTATED. REPORTEDLY, THE CATHETER WAS REPLACED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20392850 ↗MalfunctionOct 7, 2024Collapse; Material Deformation; Collapse; Material Deformation
IT WAS REPORTED THAT TWO MONTHS AND TWENTY-ONE DAYS POST A DIALYSIS CATHETER PLACEMENT, THE PLASTIC LUMEN ON THE CATHETER WAS ALLEGEDLY COLLAPSED. IT WAS FURTHER REPORTED THAT THE PLASTIC ALLEGEDLY WORE DOWN, SPECIFICALLY WHERE CLAMPS ARE EVEN WHEN ROTATED. REPORTEDLY, THE CATHETER WAS REPLACED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20392840 ↗MalfunctionOct 7, 2024Collapse; Material Deformation; Collapse; Material Deformation
IT WAS REPORTED THAT TWO MONTHS AND TWENTY-ONE DAYS POST A DIALYSIS CATHETER PLACEMENT, THE PLASTIC LUMEN ON THE CATHETER WAS ALLEGEDLY COLLAPSED. IT WAS FURTHER REPORTED THAT THE PLASTIC ALLEGEDLY WORE DOWN, SPECIFICALLY WHERE CLAMPS ARE EVEN WHEN ROTATED. REPORTEDLY, THE CATHETER WAS REPLACED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20392752 ↗MalfunctionOct 7, 2024Collapse; Material Deformation; Collapse; Material Deformation
IT WAS REPORTED THAT TWO MONTHS AND TWENTY-ONE DAYS POST A DIALYSIS CATHETER PLACEMENT, THE PLASTIC LUMEN ON THE CATHETER WAS ALLEGEDLY COLLAPSED. IT WAS FURTHER REPORTED THAT THE PLASTIC ALLEGEDLY WORE DOWN, SPECIFICALLY WHERE CLAMPS ARE EVEN WHEN ROTATED. REPORTEDLY, THE CATHETER WAS REPLACED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20392730 ↗MalfunctionOct 7, 2024Collapse; Material Deformation; Collapse; Material Deformation
IT WAS REPORTED THAT TWO MONTHS AND TWENTY-ONE DAYS POST A DIALYSIS CATHETER PLACEMENT, THE PLASTIC LUMEN ON THE CATHETER WAS ALLEGEDLY COLLAPSED. IT WAS FURTHER REPORTED THAT THE PLASTIC ALLEGEDLY WORE DOWN, SPECIFICALLY WHERE CLAMPS ARE EVEN WHEN ROTATED. REPORTEDLY, THE CATHETER WAS REPLACED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20392579 ↗MalfunctionOct 7, 2024Collapse; Material Deformation; Collapse; Material Deformation
IT WAS REPORTED THAT TWO MONTHS AND TWENTY-ONE DAYS POST A DIALYSIS CATHETER PLACEMENT, THE PLASTIC LUMEN ON THE CATHETER WAS ALLEGEDLY COLLAPSED. IT WAS FURTHER REPORTED THAT THE PLASTIC ALLEGEDLY WORE DOWN, SPECIFICALLY WHERE CLAMPS ARE EVEN WHEN ROTATED. REPORTEDLY, THE CATHETER WAS REPLACED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20392444 ↗MalfunctionOct 7, 2024Collapse; Material Deformation; Collapse; Material Deformation
IT WAS REPORTED THAT SIX MONTHS POST DIALYSIS CATHETER PLACEMENT, PLASTIC ON LUMEN ON THE DEVICE WAS ALLEGEDLY COLLAPSED. IT WAS FURTHER REPORTED THAT THE PLASTIC IS WEARING DOWN, SPECIFICALLY WHERE CLAMPS ARE EVEN WHEN ROTATED, REPORTEDLY, SOMETIMES UNSTABLE TO RUN. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20382020 ↗MalfunctionOct 7, 2024Collapse; Material Deformation; Collapse; Material Deformation
IT WAS REPORTED THAT SIX MONTHS POST DIALYSIS CATHETER PLACEMENT, PLASTIC ON LUMEN ON THE DEVICE WAS ALLEGEDLY COLLAPSED. IT WAS FURTHER REPORTED THAT THE PLASTIC IS WEARING DOWN, SPECIFICALLY WHERE CLAMPS ARE EVEN WHEN ROTATED, REPORTEDLY, SOMETIMES UNSTABLE TO RUN. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20382017 ↗MalfunctionOct 7, 2024Collapse; Material Deformation; Collapse; Material Deformation
IT WAS REPORTED THAT SIX MONTHS POST DIALYSIS CATHETER PLACEMENT, PLASTIC ON LUMEN ON THE DEVICE WAS ALLEGEDLY COLLAPSED. IT WAS FURTHER REPORTED THAT THE PLASTIC IS WEARING DOWN, SPECIFICALLY WHERE CLAMPS ARE EVEN WHEN ROTATED, REPORTEDLY, SOMETIMES UNSTABLE TO RUN. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20382016 ↗MalfunctionOct 7, 2024Collapse; Material Deformation; Collapse; Material Deformation
IT WAS REPORTED THAT SIX MONTHS POST DIALYSIS CATHETER PLACEMENT, PLASTIC ON LUMEN ON THE DEVICE WAS ALLEGEDLY COLLAPSED. IT WAS FURTHER REPORTED THAT THE PLASTIC IS WEARING DOWN, SPECIFICALLY WHERE CLAMPS ARE EVEN WHEN ROTATED, REPORTEDLY, SOMETIMES UNSTABLE TO RUN. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20381969 ↗MalfunctionOct 7, 2024Collapse; Material Deformation; Collapse; Material Deformation
IT WAS REPORTED THAT SIX MONTHS POST DIALYSIS CATHETER PLACEMENT, PLASTIC ON LUMEN ON THE DEVICE WAS ALLEGEDLY COLLAPSED. IT WAS FURTHER REPORTED THAT THE PLASTIC IS WEARING DOWN, SPECIFICALLY WHERE CLAMPS ARE EVEN WHEN ROTATED, REPORTEDLY, SOMETIMES UNSTABLE TO RUN. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20381732 ↗MalfunctionOct 7, 2024Collapse; Material Deformation; Collapse; Material Deformation
IT WAS REPORTED THAT SIX MONTHS POST DIALYSIS CATHETER PLACEMENT, PLASTIC ON LUMEN ON THE DEVICE WAS ALLEGEDLY COLLAPSED. IT WAS FURTHER REPORTED THAT THE PLASTIC IS WEARING DOWN, SPECIFICALLY WHERE CLAMPS ARE EVEN WHEN ROTATED, REPORTEDLY, SOMETIMES UNSTABLE TO RUN. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20381730 ↗MalfunctionDec 9, 2023Fracture; Premature Separation; Fracture; Premature Separation
IT WAS REPORTED THAT DURING PREPARATION OF A PORT PLACEMENT PROCEDURE, THE AIR GUARD VALVE WAS ALLEGEDLY CRACKED AND BEGINNING TO PEEL APART. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18298133 ↗MalfunctionOct 21, 2023Fracture; Fracture
IT WAS REPORTED THAT THREE MONTHS AND TWENTY-SEVEN DAYS POST DIALYSIS CATHETER PLACEMENT, THE CATHETER ALLEGEDLY SPLIT IN THE ARTERIAL PORTION OF THE EXTERNAL LUMEN. REPORTEDLY, THE CATHETER WAS REPLACED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17982143 ↗InjuryFeb 13, 2023Obstruction of Flow; Expulsion; Obstruction of Flow; Expulsion
IT WAS REPORTED THROUGH RESULTS OF A CLINICAL TRIAL THAT TWO MONTHS AND FOUR DAYS POST INDEX PROCEDURE, THE SUBJECT HAD SERIOUS ADVERSE EVENT OF CLOTTED CATHETER AND THE RELATIONSHIP TO STUDY DEVICE WAS POSSIBLY RELATED. IT WAS FURTHER REPORTED THAT, THE SUBJECT HAD SERIOUS ADVERSE EVENT OF EXPOSED CUFF CATHETER. THE RELATIONSHIP TO STUDY DEVICE WAS POSSIBLY RELATED AND ASSOCIATED WITH DEVICE FAILURE. REPORTEDLY, THE CATHETER WAS REMOVED AND THE REASON FOR REMOVAL WAS THROMBOSIS. THE PROCEDURE WAS COMPLETED BY EXCHANGING ANOTHER CATHETER. HOWEVER, THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.
Patient problems: Thrombosis/Thrombus
MAUDE report 16362156 ↗MalfunctionNov 10, 2021Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT WAS REPORTED THAT DURING PREPARATION OF A PROCEDURE, THE FOREIGN PARTICLE WAS ALLEGEDLY FOUND IN THE DEVICE. THE PROCEDURE WAS COMPLETED BY USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12784298 ↗MalfunctionMar 17, 2020Fracture; Fracture
IT WAS REPORTED THAT DURING DIALYSIS CATHETER INSERTION, THE CATHETER TIP ALLEGEDLY BROKE OFF DURING THE TUNNELING PROCESS. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9845427 ↗MalfunctionMar 26, 2019Defective Component; Defective Component
IT WAS REPORTED THAT DURING PREPARATION FOR CHRONIC CATHETER PLACEMENT, THE WHITE PLASTIC TUNNELER TIP ALLEGEDLY WAS CRACKED AND THE TIP BROKE OFF. ANOTHER DEVICE WAS USED TO COMPLETE PLACEMENT. THERE WAS NO REPORTED PATIENT CONTACT.
Patient problems: No Patient Involvement
MAUDE report 8452078 ↗MalfunctionMar 25, 2019Defective Component; Defective Component
IT WAS REPORTED THAT DURING PREPARATION FOR CHRONIC CATHETER PLACEMENT, THE WHITE PLASTIC TUNNELER TIP ALLEGEDLY WAS CRACKED AND THE TIP BROKE OFF. ANOTHER DEVICE WAS USED TO COMPLETE PLACEMENT. THERE WAS NO REPORTED PATIENT CONTACT.
Patient problems: No Patient Involvement
MAUDE report 8450235 ↗MalfunctionMar 25, 2019Defective Component; Defective Component
IT WAS REPORTED THAT DURING PREPARATION FOR CHRONIC CATHETER PLACEMENT, THE WHITE PLASTIC TUNNELER TIP ALLEGEDLY WAS CRACKED AND THE TIP BROKE OFF. ANOTHER DEVICE WAS USED TO COMPLETE PLACEMENT. THERE WAS NO REPORTED PATIENT CONTACT.
Patient problems: No Patient Involvement
MAUDE report 8450211 ↗MalfunctionMar 25, 2019Defective Component; Defective Component
IT WAS REPORTED THAT DURING PREPARATION FOR CHRONIC CATHETER PLACEMENT, THE WHITE PLASTIC TUNNELER TIP ALLEGEDLY WAS CRACKED AND THE TIP BROKE OFF. ANOTHER DEVICE WAS USED TO COMPLETE PLACEMENT. THERE WAS NO REPORTED PATIENT CONTACT.
Patient problems: No Patient Involvement
MAUDE report 8450142 ↗MalfunctionFeb 26, 2019Defective Component; Defective Component
IT WAS REPORTED THAT UPON OPENING THE CHRONIC CATHETER KIT, THE WHITE PLASTIC TUNNELER TIP WAS ALLEGEDLY CRACKED. THERE WAS NO REPORTED PATIENT CONTACT.
Patient problems: No Patient Involvement
MAUDE report 8372404 ↗
Other Catheter, hemodialysis, implanted devices
- Palindrome 8888145014Mozarc Medical US LLC · Catheter, hemodialysis, implanted1 recall312 reports
- Palindrome 8888145015Mozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 death219 reports
- HemoStar Long-Term Hemodialysis Catheter 5833690Bard Access Systems, Inc. · Catheter, hemodialysis, implanted2 deaths290 reports
- Palindrome 8888119369PMozarc Medical US LLC · Catheter, hemodialysis, implanted3 recalls1 death14 reports
- HemoStar Long-Term Hemodialysis Catheter 5833730Bard Access Systems, Inc. · Catheter, hemodialysis, implanted1 death141 reports
- GlidePath Long-Term Hemodialysis Catheter with Preloaded Stylet 5393230Bard Access Systems, Inc. · Catheter, hemodialysis, implanted5 deaths138 reports
- Palindrome 8888123404PMozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 death17 reports
- Palindrome 8888119364PMozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 death17 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗