Skip to content

GlidePath Long-Term Hemodialysis Catheter with Preloaded Stylet 5397310Recalls & reported problems

Bard Access Systems, Inc. · GLIDEPATH HEMODIALYSIS CATH, ST, DL, 14.5 FR. 31cm · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00801741012327), received between Feb 29, 2016 and Oct 21, 2025 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Malposition of Device” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Malposition of Device1
  • Obstruction of Flow1
  • Suction Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Known Impact Or Consequence To Patient1
  • Thrombosis/Thrombus1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 21, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH LITIGATION PROCESS THAT SOMETIME LATER PORT PLACEMENT PROCEDURE, THE PATIENT ALLEGEDLY HAD BLOOD CLOT. HOWEVER, THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.

    Patient problems: Thrombosis/Thrombus

    MAUDE report 23345293 ↗
  • MalfunctionMar 25, 2022Suction Problem; Obstruction of Flow; Suction Problem; Obstruction of Flow

    IT WAS REPORTED THAT DURING A CATHETER PLACEMENT PROCEDURE, WHERE WAS A LOT OF CLOT COMING OUT AND WOULD NOT PULL FREELY. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13897725 ↗
  • InjuryFeb 29, 2016Malposition of Device

    PER SALES REPRESENTATIVE, FACILITY REPORTED THAT DURING PLACEMENT, THE CATHETER PENETRATED THE MEDIASTINUM. PHYSICIAN WAS ABLE TO PULL THE CATHETER BACK AND USED A WIRE TO CONTINUE ADVANCING THE CATHETER FOR PLACEMENT. NO PATIENT INJURY REPORTED OR ADDITIONAL MEDICAL INTERVENTION REQUIRED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 5466279 ↗

Other Catheter, hemodialysis, implanted devices

All Catheter, hemodialysis, implanted devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page