HemoSplit Long-Term Hemodialysis Catheter 5743690Recalls & reported problems
Bard Access Systems, Inc. · HEMOSPLIT HEMODIALYSIS CATH, ST, w/MI/GUIDEWIRE, AIRGUARD, 14.5. 19cm · Class 2 · Implant
All models of HemoSplit Long-Term Hemodialysis Catheter (18)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Bard HemoSplit long-term hemodialysis catheter is an implanted catheter used to access blood for dialysis. The 51 reports mostly describe cracked or broken tubing, pinhole leaks of blood or fluid, and kinked or collapsed parts, often found before or during placement; one report describes a torn catheter with bleeding after the procedure. There are no deaths and 1 injury. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 50
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 50 adverse event reports that list this exact device (DI 00801741013263), received between Aug 14, 2018 and Apr 8, 2026 — classified as 50 malfunctions.
The most frequently reported specific device problem is “Fracture” (13 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture13
- Fluid/Blood Leak10
- Material Deformation9
- Restricted Flow rate6
- Suction Problem5
- Device Damaged Prior to Use3
- Material Opacification3
- Material Protrusion/Extrusion3
Patient problems
- No Clinical Signs, Symptoms or Conditions38
- No Consequences Or Impact To Patient10
- No Patient Involvement2
Recent adverse event reports
The 25 most recent of 50 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 8, 2026Fluid/Blood Leak; Material Puncture/Hole
ON (B)(6) 2026, A PATIENT WAS PREPPED FOR A DIALYSIS CATHETER PLACEMENT PROCEDURE USING THE HEMOSPLIT HEMODIALYSIS CATHETER. DURING THE PREPARATION, A PINHOLE WAS FOUND IN THE ARTERIAL SEGMENT OF THE Y TUBE AND ALLEGEDLY LEAKED. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24813195 ↗MalfunctionFeb 16, 2026Fluid/Blood Leak; Fracture
ON (B)(6) 2026, A PATIENT WHICH WAS SUFFERING FROM RENAL INSUFFICIENCY UNDERWENT A LONG-TERM DIALYSIS CATHETER PLACEMENT PROCEDURE USING THE HEMOSPLIT CATHETER. DURING THE PROCEDURE, STEAM LIKE AND PINHOLE LIKE LEAKAGE WAS OBSERVED AT THE CONNECTION POINT OF THE TRANSPARENT CONNECTING TUBE, WHICH INTERFERED WITH THE DIALYSIS TREATMENT. THE CATHETER WAS REMOVED AND A NEW DEVICE WAS REINSERTED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24342067 ↗MalfunctionNov 21, 2025Packaging Problem; Packaging Problem
ON(B)(6) 2025, A PATIENT UNDERWENT A DIALYSIS CATHETER PLACEMENT PROCEDURE. PRIOR TO A PLACEMENT PROCEDURE THE PACKAGING WAS ALLEGEDLY FOUND DAMAGED. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23613690 ↗MalfunctionJul 30, 2025Loss of or Failure to Bond; Fracture; Material Separation; Loss of or Failure to Bond; Fracture; Material Separation
ON (B)(6) 2025, A PATIENT UNDERWENT A DIALYSIS CATHETER PROCEDURE USING A HEMOSPLIT CATHETER. DURING THE PROCEDURE, IT WAS REPORTED THAT THE CATHETER KRAFT HAD BROKEN OFF. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22668777 ↗MalfunctionJun 19, 2025Fracture; Deformation Due to Compressive Stress; Fracture; Deformation Due to Compressive Stress
ON (B)(6) 2025, A PATIENT UNDERWENT THE DIALYSIS CATHETER PLACEMENT USING HEMOSPLIT CATHETER. IT WAS REPORTED THAT PRIOR TO THE PROCEDURE THE DIALYSIS CATHETER SHEATH TUBING WAS ALLEGEDLY FOUND FRACTURED. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22291324 ↗MalfunctionJun 17, 2025Fracture; Device Damaged Prior to Use; Material Integrity Problem; Material Split, Cut or Torn; Fracture; Device Damaged Prior to Use; Material Integrity Problem; Material Split, Cut or Torn
ON (B)(6) 2025, A PATIENT UNDERWENT A DIALYSIS CATHETER PLACEMENT PROCEDURE USING THE HEMOSPLIT CATHETER. DURING THE PREPARATION OF THE PROCEDURE, IT WAS NOTED THAT THE CATHETER WAS DAMAGED, AND A SMALL CUT WAS ALLEGEDLY FOUND IN THE TUBING. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22229371 ↗MalfunctionJun 17, 2025Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping
ON (B(6) 2025, A PATIENT UNDERWENT A DIALYSIS CATHETER PLACEMENT PROCEDURE USING THE HEMOSPLIT CATHETER. DURING THE PREPARATION OF THE PROCEDURE, A HAIR WAS FOUND IN THE STERILE PACKAGING OF THE DIALYSIS CATHETER. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22217696 ↗MalfunctionMay 9, 2025Fracture; Material Separation; Fracture; Material Separation
ON (B)(6) 2025, A PATIENT UNDERWENT HEMOSPLIT CATHETER INSERTION FOR UNKNOWN MEDICAL CONDITION. DURING INSERTION THE CATHETER BROKE OUTSIDE THE BLOOD VESSEL, AND THE BROKEN PART DID NOT ENTER THE VASCULAR SYSTEM. THE PROCEDURE WAS COMPLETED BY REPLACED WITH ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21991982 ↗MalfunctionMay 2, 2025Fluid/Blood Leak; Material Integrity Problem; Fluid/Blood Leak; Material Integrity Problem
ON (B)(6) 2025, A PATIENT UNDERWENT DIALYSIS CATHETER PLACEMENT PROCEDURE USING HEMOSPLIT DIALYSIS CATHETER. DURING THE DIALYSIS, THERE WAS LEAKAGE FROM THE VENOUS END OF THE SILICONE AND POLYURETHANE PORTION OF THE SEMI-PERMANENT DIALYSIS CATHETER. REPORTEDLY CATHETER WAS ALLEGEDLY DAMAGED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21949183 ↗MalfunctionApr 8, 2025Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
IT WAS REPORTED THAT PRIOR TO A DIALYSIS CATHETER PLACEMENT PROCEDURE; THE HAIR WAS ALLEGEDLY FOUND INSIDE THE CATHETER. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21795397 ↗MalfunctionMar 6, 2025Restricted Flow rate; Difficult to Flush; Suction Problem; Restricted Flow rate; Difficult to Flush; Suction Problem
IT WAS REPORTED THAT SOMETIMES POST A DIALYSIS CATHETER PLACEMENT, THE DEVICE ALLEGEDLY NOTED POOR CATHETER FLOW. IT WAS FURTHER REPORTED THAT THE BLOOD FLOW FLUCTUATING BY 180-200 ML/MIN, WITH MULTIPLE MACHINE ALARMS REQUIRING TREATMENT SUCH AS THROMBOLYSIS WITH UROKINASE AND CLOSURE OF THE CATHETER, AFFECTING THE ADEQUACY OF DIALYSIS. ADDITIONALLY, THERE ARE PROBLEMS DURING INFUSION AND ASPIRATION. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21536195 ↗MalfunctionMar 6, 2025Restricted Flow rate; Restricted Flow rate
IT WAS REPORTED THAT SOMETIMES POST DIALYSIS CATHETER PLACEMENT, THE DEVICE ALLEGEDLY NOTED POOR CATHETER FLOW. IT WAS FURTHER REPORTED THAT THE BLOOD FLOW FLUCTUATING, AFFECTING THE ADEQUACY OF DIALYSIS. DURING MULTIPLE DIALYSIS SESSIONS. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21534442 ↗MalfunctionMar 5, 2025Restricted Flow rate; Restricted Flow rate
IT WAS REPORTED THAT FOUR MONTHS AND EIGHTEEN DAYS POST DIALYSIS CATHETER PLACEMENT, THE DEVICE ALLEGEDLY NOTED POOR CATHETER FLOW. IT WAS FURTHER REPORTED THAT THE BLOOD FLOW FLUCTUATING, WITH MULTIPLE MACHINE ALARMS REQUIRING TREATMENT SUCH AS THROMBOLYSIS WITH UROKINASE AND CLOSURE OF THE CATHETER, AFFECTING THE ADEQUACY OF DIALYSIS. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21525798 ↗MalfunctionMar 5, 2025Restricted Flow rate; Restricted Flow rate
IT WAS REPORTED THAT FIVE MONTHS AND TWELVE DAYS POST DIALYSIS CATHETER PLACEMENT, THE DEVICE ALLEGEDLY NOTED POOR CATHETER FLOW. IT WAS FURTHER REPORTED THAT THE BLOOD FLOW FLUCTUATING, AFFECTING THE ADEQUACY OF DIALYSIS. DURING MULTIPLE DIALYSIS SESSIONS. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21525353 ↗MalfunctionFeb 4, 2025Fracture; Device Damaged Prior to Use; Fracture; Device Damaged Prior to Use
IT WAS REPORTED THAT PRIOR TO A DIALYSIS CATHETER PLACEMENT PROCEDURE, THE TEAR OFF SHEATH ALLEGEDLY BREAKS AFTER UNPACKING. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21302433 ↗MalfunctionNov 25, 2024Restricted Flow rate; Difficult to Flush; Obstruction of Flow; Deformation Due to Compressive Stress; Restricted Flow rate; Difficult to Flush; Obstruction of Flow; Deformation Due to Compressive Stress
IT WAS REPORTED THAT PRIOR TO A DIALYSIS CATHETER PLACEMENT PROCEDURE, THE PIPE WORK OF THE DEVICE WAS ALLEGEDLY NOT CLEAR. IT WAS FURTHER REPORTED THAT THERE WAS A LOT OF RESISTANCE TO THE FLOW OF WATER. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20769263 ↗MalfunctionNov 1, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT DURING A DIALYSIS CATHETER PLACEMENT PROCEDURE, THE INTRODUCER NEEDLE WAS ALLEGEDLY FOUND TO BE LEAKING, AND THE PROCEDURE COULD NOT BE COMPLETED. THE PROCEDURE WAS COMPLETED BY USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20585453 ↗MalfunctionOct 9, 2024Degraded; Degraded
IT WAS REPORTED THAT SOMETIME POST A DIALYSIS CATHETER PLACEMENT, THE CATHETER WAS ALLEGEDLY DISSOLVING WHICH RESULTED IN A DYSFUNCTIONAL CATHETER, WHICH PREVENTED THE COMPLETION OF NORMAL HEMODIALYSIS TREATMENTS. REPORTEDLY, THE CATHETER WAS REMOVED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20405421 ↗MalfunctionSep 29, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT SOMETIME POST A DIALYSIS CATHETER PLACEMENT, THE CATHETER ALLEGEDLY FOUND TO HAVE A BLOOD LEAK. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20330496 ↗MalfunctionJul 9, 2024Degraded; Restricted Flow rate; Degraded; Restricted Flow rate
IT WAS REPORTED THAT SOMETIMES POST A DIALYSIS CATHETER PLACEMENT, THE CATHETER ALLEGEDLY HAD DETERIORATED AT THE CONNECTION BETWEEN THE ARTERIAL AND VENOUS ENDS OF THE CATHETER RESULTING IN INSUFFICIENT BLOOD FLOW THROUGH THE CATHETER. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19698679 ↗MalfunctionJun 26, 2024Fluid/Blood Leak; Suction Problem; Fluid/Blood Leak; Suction Problem
IT WAS REPORTED THAT DURING A DIALYSIS CATHETER PLACEMENT PROCEDURE VIA THE JUGULAR VEIN, THE CATHETER ALLEGEDLY LEAKED AND THE NEGATIVE PRESSURE ALLEGEDLY FAILED. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19611970 ↗MalfunctionDec 28, 2023Air/Gas in Device; Air/Gas in Device
IT WAS REPORTED THAT DURING A DIALYSIS CATHETER PLACEMENT PROCEDURE, THE AIR LEAKAGE WAS ALLEGEDLY FOUND AFTER SUCTION AFTER CATHETERIZATION. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18411642 ↗MalfunctionJul 17, 2023Fracture; Device Damaged Prior to Use; Fracture; Device Damaged Prior to Use
IT WAS REPORTED THAT PRIOR TO A DIALYSIS CATHETER PLACEMENT PROCEDURE, THE WHITE RUBBER TUBE OF THE TUNNELLER INSIDE THE DIALYSIS CATHETER WAS ALLEGEDLY FOUND TO BE RUPTURED. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17334999 ↗MalfunctionDec 29, 2022Fracture; Fracture
IT WAS REPORTED THAT SOMETIME POST DIALYSIS CATHETER PLACEMENT, THE CLIP WAS ALLEGEDLY DAMAGED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16066288 ↗MalfunctionJul 8, 2022Suction Problem; Air/Gas in Device; Suction Problem; Air/Gas in Device
IT WAS REPORTED THAT DURING A DIALYSIS CATHETER PLACEMENT PROCEDURE, THE INTRODUCER ALLEGEDLY HAD SUCTION ISSUE. IT WAS FURTHER REPORTED THAT THE PLASTIC PART OF THE NEEDLE LEAKED AIR. THE PROCEDURE WAS COMPLETED BY USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14953045 ↗
Other Catheter, hemodialysis, implanted devices
- Palindrome 8888145014Mozarc Medical US LLC · Catheter, hemodialysis, implanted1 recall312 reports
- Palindrome 8888145015Mozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 death219 reports
- HemoStar Long-Term Hemodialysis Catheter 5833690Bard Access Systems, Inc. · Catheter, hemodialysis, implanted2 deaths290 reports
- Palindrome 8888119369PMozarc Medical US LLC · Catheter, hemodialysis, implanted3 recalls1 death14 reports
- HemoStar Long-Term Hemodialysis Catheter 5833730Bard Access Systems, Inc. · Catheter, hemodialysis, implanted1 death141 reports
- GlidePath Long-Term Hemodialysis Catheter with Preloaded Stylet 5393230Bard Access Systems, Inc. · Catheter, hemodialysis, implanted5 deaths138 reports
- Palindrome 8888123404PMozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 death17 reports
- Palindrome 8888119364PMozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 death17 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page