HemoStar Long-Term Hemodialysis Catheter 5833350Recalls & reported problems
Bard Access Systems, Inc. · HEMOSTAR HEMODIALYSIS CATH, ST, AIRGUARD, 14.5 FR. 35cm · Class 2 · Implant
All models of HemoStar Long-Term Hemodialysis Catheter (13)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 21
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 21 adverse event reports that list this exact device (DI 00801741013362), received between Apr 4, 2018 and Apr 30, 2025 — classified as 21 malfunctions.
The most frequently reported specific device problem is “Fracture” (9 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture9
- Fluid/Blood Leak3
- Air/Gas in Device2
- Detachment of Device or Device Component2
- Tear, Rip or Hole in Device Packaging2
- Component Missing1
- Deformation Due to Compressive Stress1
- Dent in Material1
Patient problems
- No Clinical Signs, Symptoms or Conditions9
- No Consequences Or Impact To Patient6
- No Patient Involvement5
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 21 most recent of 21 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 30, 2025Device Markings/Labelling Problem; Device Markings/Labelling Problem
ON (B)(6) 2025, A PATIENT WAS UNDERGOING A DIALYSIS CATHETER PLACEMENT PROCEDURE. IT WAS REPORTED THAT THE CATHETER PACKAGING DID NOT HAVE ANY LABEL STICKERS SHOWING THE LOT NUMBER OR THE EXPIRATION DATE OF THE PRODUCT. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21932679 ↗MalfunctionDec 28, 2023Difficult to Remove; Failure to Advance; Deformation Due to Compressive Stress; Physical Resistance/Sticking; Difficult to Remove; Failure to Advance; Deformation Due to Compressive Stress; Physical Resistance/Sticking
IT WAS REPORTED THAT DURING A DIALYSIS CATHETER PLACEMENT PROCEDURE, THE CATHETER ALLEGEDLY CANNOT BE PASSED INTO THE GUIDEWIRE. IT WAS FURTHER REPORTED THAT THE GUIDEWIRE ALLEGEDLY STUCK ON THE PUNCTURE NEEDLE AND ALLEGEDLY COULD NOT BE REMOVED. IT WAS FURTHER REPORTED THAT THE GUIDEWIRE WAS ALLEGEDLY BENT. REPORTEDLY, THE PATIENT EXPERIENCED ALLEGEDLY BLEEDING AND LOCAL COMPRESSION WAS DONE TO STOP THE BLEEDING. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18408201 ↗MalfunctionOct 11, 2023Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED THAT SOMETIME POST A DIALYSIS CATHETER PLACEMENT, THE CATHETER TANK BODY WAS ALLEGEDLY DAMAGED. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17911109 ↗MalfunctionOct 10, 2023Fracture; Fracture
IT WAS REPORTED THAT SOMETIME POST A DIALYSIS CATHETER PLACEMENT, THE CLAMP WAS ALLEGEDLY BROKEN. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17903566 ↗MalfunctionSep 20, 2023Fracture; Fracture
IT WAS REPORTED THAT SOMETIME POST DIALYSIS CATHETER PLACEMENT, THE CLAMP WAS ALLEGEDLY BROKEN. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17778510 ↗MalfunctionJun 28, 2023Fracture; Air/Gas in Device; Fracture; Air/Gas in Device
IT WAS REPORTED THAT DURING A PORT PLACEMENT PROCEDURE, INTRODUCER NEEDLE ALLEGEDLY RUPTURED. IT WAS FURTHER REPORTED THAT INTRODUCER NEEDLE ALLEGEDLY LEAKED AIR. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17215765 ↗MalfunctionJun 27, 2023Fluid/Blood Leak; Fracture; Air/Gas in Device; Fluid/Blood Leak; Fracture; Air/Gas in Device
IT WAS REPORTED THAT DURING A DIALYSIS CATHETER PLACEMENT PROCEDURE, THE INTRODUCER NEEDLE WAS ALLEGEDLY RUPTURED. IT WAS FURTHER REPORTED THAT THE INTRODUCER NEEDLE ALLEGEDLY LEAKED AIR. FURTHERMORE, SALINE ALLEGEDLY LEAKED IN THE PINK JOINT OF THE INTRODUCER NEEDLE WHEN INJECTED. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17206738 ↗MalfunctionMay 24, 2023Detachment of Device or Device Component; Material Deformation; Detachment of Device or Device Component; Material Deformation
IT WAS REPORTED THAT DURING THE PREPARATION OF DIALYSIS CATHETER PLACEMENT PROCEDURE, UPON UNPACKING OF THE CATHETER, A PROBLEM WAS ALLEGEDLY FOUND WITH THE SHEATH WHICH HAD UNEVEN RUBBER. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16990442 ↗MalfunctionApr 15, 2022Material Opacification; Stretched; Material Protrusion/Extrusion; Material Opacification; Stretched; Material Protrusion/Extrusion
IT WAS REPORTED THAT APPROXIMATELY ONE YEAR POST CATHETER PLACEMENT, THE DEVICE ALLEGEDLY HAD WHITE OPACIFICATION OF THE TRANSPARENT PART. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14119525 ↗MalfunctionOct 26, 2020Fracture; Fracture
IT WAS REPORTED THAT DURING A CATHETER PLACEMENT, THE NEEDLE WAS ALLEGEDLY FRACTURED AND PUNCTURE FAILURE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10734439 ↗MalfunctionSep 8, 2020Fracture; Suction Problem; Fracture; Suction Problem
IT WAS REPORTED THAT DURING A CATHETER PLACEMENT, THE PLASTIC CONNECTOR OF INTRODUCER NEEDLE WAS BROKEN. IT WAS FURTHER REPORTED THAT, ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10500622 ↗MalfunctionSep 1, 2020Fracture; Fracture
IT WAS REPORTED THAT DURING IMPLANTATION PROCEDURE, THE PLASTIC FITTING OF THE INTRODUCER NEEDLE WAS ALLEGEDLY BROKEN. REPORTEDLY THE ALLEGED ISSUE LEADS TO AIR LEAKS DURING ASPIRATION. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10475984 ↗MalfunctionApr 16, 2020Fracture; Fracture
IT WAS REPORTED THAT PRIOR TO A HEMODIALYSIS PROCEDURE, INTRODUCTION NEEDLE OF THE DIALYSIS CATHETER WAS ALLEGEDLY FOUND TO BE FRACTURED. REPORTEDLY, THE INTRODUCTION NEEDLE WAS REPLACED AND THE PROCEDURE WAS COMPLETED. THERE WAS NO REPORTED PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 9970108 ↗MalfunctionFeb 5, 2020Device Damaged Prior to Receipt by User; Tear, Rip or Hole in Device Packaging; Device Damaged Prior to Receipt by User; Tear, Rip or Hole in Device Packaging
IT WAS REPORTED THAT UPON OPENING THE DIALYSIS CATHETER PACKAGE, THE PACKAGE WAS ALLEGEDLY TORN. THERE WAS NO PATIENT CONTACT.
Patient problems: No Patient Involvement
MAUDE report 9671250 ↗MalfunctionOct 8, 2019Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THE INTRODUCER NEEDLE ALLEGEDLY LEAKED AIR. THERE WAS NO PATIENT CONTACT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9166497 ↗MalfunctionOct 8, 2019Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging
IT WAS REPORTED THAT THE DIALYSIS KIT PACKAGING BAG WAS ALLEGEDLY DAMAGED. THERE WAS NO PATIENT CONTACT.
Patient problems: No Patient Involvement
MAUDE report 9166491 ↗MalfunctionOct 1, 2019Fluid/Blood Leak; Fracture; Fluid/Blood Leak; Fracture
IT WAS REPORTED THAT DURING A DIALYSIS CATHETER PLACEMENT THE INTRODUCER NEEDLE WAS INSERTED UNDER ULTRASOUND GUIDANCE AND ALLEGEDLY FLUID WAS LEAKING FROM THE HUB DURING ASPIRATION. REPORTEDLY, IT WAS IDENTIFIED THAT THE HUB ALLEGEDLY RUPTURED. A NEW KIT WAS OPENED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9142052 ↗MalfunctionAug 22, 2019Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
IT WAS REPORTED THAT UPON OPENING THE DIALYSIS CATHETER PACKAGE, FOREIGN MATERIAL WAS ALLEGEDLY PRESENT AND THE CATHETER WAS NOT USED. REPORTEDLY, ANOTHER CATHETER WAS OPENED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Patient Involvement
MAUDE report 8923246 ↗MalfunctionMay 6, 2019Component Missing; Component Missing
IT WAS REPORTED THAT THE GUIDEWIRE WAS ALLEGEDLY MISSING. THERE WAS NO PATIENT CONTACT.
Patient problems: No Patient Involvement
MAUDE report 8582774 ↗MalfunctionSep 12, 2018Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT THE CATHETER DISCONNECTED AT THE ARTERIAL END CONNECTOR. AFTER INSERTION IT WAS FOUND THAT THE CONNECTION BETWEEN THE CATHETER CONNECTOR AND THE CATHETER WAS LOOSE, THE CATHETER AND CONNECTER WERE ADJUSTED AND THE CONNECTOR COMPLETELY DETACHED. THE DEVICE WAS IMMEDIATELY REMOVED AND ANOTHER CATHETER WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7868365 ↗MalfunctionApr 4, 2018Dent in Material
IT WAS REPORTED BY THE NURSE THAT DURING THE USE OF THE DEVICE FOR THE THIRD TIME AFTER CATHETERIZATION, IT WAS FOUND THAT THE INDENTATION AT THE THUMB CLIP OF THE CATHETER COULD NOT BE REBOUNDED, RESULTING IN NARROW TUBING WHICH AFFECTS NORMAL BLOOD FLOW, THEN THE VENOUS PRESSURE INCREASED RESULTING IN THE DEVICE ALARMING. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7395995 ↗
Other Catheter, hemodialysis, implanted devices
- Palindrome 8888145014Mozarc Medical US LLC · Catheter, hemodialysis, implanted1 recall312 reports
- Palindrome 8888145015Mozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 death219 reports
- HemoStar Long-Term Hemodialysis Catheter 5833690Bard Access Systems, Inc. · Catheter, hemodialysis, implanted2 deaths290 reports
- Palindrome 8888119369PMozarc Medical US LLC · Catheter, hemodialysis, implanted3 recalls1 death14 reports
- HemoStar Long-Term Hemodialysis Catheter 5833730Bard Access Systems, Inc. · Catheter, hemodialysis, implanted1 death141 reports
- GlidePath Long-Term Hemodialysis Catheter with Preloaded Stylet 5393230Bard Access Systems, Inc. · Catheter, hemodialysis, implanted5 deaths138 reports
- Palindrome 8888123404PMozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 death17 reports
- Palindrome 8888119364PMozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 death17 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page