HemoStar Long-Term Hemodialysis Catheter 5834150Recalls & reported problems

Bard Access Systems, Inc. · HEMOSTAR HEMODIALYSIS CATH, CURVED, AIRGUARD, 14.5 FR. 19cm · Class 2 · Implant

All models of HemoStar Long-Term Hemodialysis Catheter (13)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00801741013423), received between Mar 13, 2018 and Feb 3, 2025 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Material Puncture/Hole” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Puncture/Hole2
  • Backflow1
  • Deformation Due to Compressive Stress1
  • Device Operates Differently Than Expected1
  • Difficult to Flush1
  • Fracture1
  • Loss of or Failure to Bond1
  • Product Quality Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • No Code Available1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 3, 2025Fracture; Material Puncture/Hole; Fracture; Material Puncture/Hole

    IT WAS REPORTED THAT A DIALYSIS CATHETER PLACEMENT PROCEDURE OCCURRED ON (B)(6) 2024, AND SOMETIME POST PROCEDURE, THE EXTERNAL LUMEN OF THE CATHETER ALLEGEDLY DEVELOPED A SMALL PINHOLE AND WAS EXPLANTED ON (B)(6) 2024. FOLLOW UP DETAILS PROVIDED BY THE COMPLAINANT ON (B)(6) 2024 INDICATED THAT THE PATIENT WAS UNKNOWN GENDER, 82 Y/O. THE LINE WAS REPLACED, AND THE PATIENT WAS SUBSEQUENTLY DISCHARGED WITH NO OTHER INTERVENTIONS REQUIRED FOR THIS INCIDENT. THE DEVICE IS NOT AVAILABLE FOR RETURN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21291823 ↗
  • MalfunctionJan 7, 2025Material Puncture/Hole; Material Puncture/Hole

    IT WAS REPORTED THAT ONE MONTH AND TWENTY-ONE DAYS POST A DIALYSIS CATHETER PLACEMENT, THE EXTERNAL LUMEN OF THE CATHETER ALLEGEDLY DEVELOPED A SMALL PINHOLE. REPORTEDLY, THE CATHETER WAS REPLACED, AND THE PATIENT WAS SUBSEQUENTLY DISCHARGED WITH NO OTHER INTERVENTIONS REQUIRED FOR THIS INCIDENT. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21093463 ↗
  • MalfunctionJul 26, 2022Loss of or Failure to Bond; Loss of or Failure to Bond

    IT WAS REPORTED THAT DURING A DIALYSIS CATHETER PLACEMENT PROCEDURE, THE CUFF OF THE CATHETER WAS ALLEGEDLY FAILED TO BOND. IT WAS FURTHER THAT THE CATHETER WAS REMOVED AND REPLACED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15101127 ↗
  • MalfunctionApr 21, 2022Backflow; Difficult to Flush; Suction Problem; Backflow; Difficult to Flush; Suction Problem

    IT WAS REPORTED THAT TWO DAYS POST DIALYSIS CATHETER IMPLANT, THE CATHETER ALLEGEDLY RUN IN THE REVERSE POSITION. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14165753 ↗
  • MalfunctionMar 22, 2022Reflux within Device; Reflux within Device

    IT WAS REPORTED THAT ONE DAY POST DIALYSIS CATHETER PLACEMENT PROCEDURE, THE DEVICE WAS ALLEGEDLY RUNNING IN THE REVERSE POSITION SINCE THE FIRST DIALYSIS AND WAS UNABLE TO BE RUN IN A STRAIGHT POSITION RESULTING IN POOR PATIENT CLEARANCE DUE TO HIGH RECIRCULATION. THE DEVICE WAS REPORTED TO BE REMOVED AND REPLACED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13854462 ↗
  • MalfunctionJun 17, 2021Deformation Due to Compressive Stress; Deformation Due to Compressive Stress

    IT WAS REPORTED THAT DURING A DIALYSIS CATHETER PLACEMENT, THE ARTERIAL LINE WAS ALLEGEDLY KINKED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12015273 ↗
  • MalfunctionMar 13, 2018Product Quality Problem; Device Operates Differently Than Expected

    IT WAS REPORTED THAT THE RUBBER MATERIAL USED IN THE4 HEMOSTAR CVC ON THE TWO PORTS OUTSIDE THE PATIENT'S BODY ALLEGEDLY HAS MEMORY FROM THE CLAMPS, MEANING WHEN THE CLAMPS WERE APPLIED BETWEEN DIALYSIS, USUALLY TWO DAYS BETWEEN, THE RUBBER MATERIALS KEEP THE MEMORY OF THE CLAMP AND DO NOT GO BACK TO ITS ORIGINAL ROUND SHAPE WHEN UNCLAMPED , IT REMAINS COLLAPSED, FLAT AND NARROW. THIS APPARENTLY CAUSES REDUCTION AND RESTRICTION IN BLOOD FLOW THROUGH THE PORTS OF THE CVC, IN TURN CAUSES REDUCTION IN PUMP SPEED USED IN DIALYSIS, RESULTING IN REDUCTION IN CLEARANCE IN THE DIALYSIS TREATMENT OF THE PATIENT.

    Patient problems: No Code Available

    MAUDE report 7335831 ↗