EQUISTREAM XK Long-Term Hemodialysis Catheter with Preloaded Stylet 5913190Recalls & reported problems
Bard Access Systems, Inc. · EQUISTREAM XK HEMODIALYSIS CATH w/STYLET, ST, AIRGUARD,16 FR. 19cm · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00801741013836), received between Apr 16, 2019 and Mar 24, 2023 — classified as 1 injury, 7 malfunctions.
The most frequently reported specific device problem is “Defective Component” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Defective Component3
- Obstruction of Flow3
- Malposition of Device2
- Component Incompatible1
- Component Missing1
- Patient Device Interaction Problem1
Patient problems
- No Consequences Or Impact To Patient3
- No Patient Involvement3
- No Clinical Signs, Symptoms or Conditions2
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 24, 2023Obstruction of Flow; Obstruction of Flow
IT WAS REPORTED THROUGH RESULTS OF A CLINICAL TRIAL THAT FOUR MONTHS AND EIGHTEEN DAYS POST INDEX PROCEDURE COMPLETION, THE SUBJECT HAD ADVERSE EVENT OF CLOTTED CATHETER. IT WAS FURTHER REPORTED THAT, THE ADVERSE EVENT RELATIONSHIP TO STUDY DEVICE WAS POSSIBLY RELATED AND THUS THE ACTION WAS TAKEN TO REMOVE THE DEVICE. REPORTEDLY, THE CATHETER WAS REMOVED. THE PROCEDURE WAS COMPLETED EXCHANGING ANOTHER NEW CATHETER. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16610895 ↗InjuryMar 24, 2023Obstruction of Flow; Obstruction of Flow
IT WAS REPORTED THROUGH RESULTS OF A CLINICAL TRIAL THAT NINE MONTHS AND TWENTY-SIX DAYS POST INDEX PROCEDURE COMPLETION, THE SUBJECT HAD ADVERSE EVENT OBSTRUCTED CATHETER WITH FIBRIN SHEATH FORMATION. IT WAS FURTHER REPORTED THAT THE ADVERSE EVENT RELATIONSHIP TO STUDY DEVICE WAS POSSIBLY RELATED WHICH REQUIRES ADDITIONAL SURGICAL INTERVENTION AND SO ACTION WAS TAKEN TO REMOVE THE DEVICE. REPORTEDLY, THE CATHETER WAS REMOVED. THE PROCEDURE WAS COMPLETED EXCHANGING ANOTHER NEW CATHETER. THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 16609195 ↗MalfunctionJan 23, 2021Malposition of Device; Patient Device Interaction Problem; Malposition of Device; Patient Device Interaction Problem
THIS REPORT SUMMARIZES FOUR MALFUNCTIONS. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL DL900J VENA CAVA FILTER ALLEGEDLY EXPERIENCED PERFORATION AND TILT. THIS INFORMATION WAS RECEIVED FROM VARIOUS SOURCES. THE FOUR MALFUNCTIONS INVOLVED PATIENTS WITH NO CONSEQUENCES. OF THE FOUR MALFUNCTIONS, ONE MALE AND THREE FEMALES AGES WERE REPORTED RANGING FROM 44 TO 77. THE WEIGHT OF ONE PATIENT WAS (B)(6) KGS AND THE REMAINING PATIENT WEIGHTS ARE UNKNOWN.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11219798 ↗MalfunctionJun 6, 2019Defective Component; Defective Component
IT WAS REPORTED THAT DURING THE PLACEMENT OF THE DIALYSIS CATHETER, THE DILATOR ALLEGEDLY BROKE WHEN ATTEMPTING TO TAKE THE INNER STYLET OUT. REPORTEDLY, THE DEVICE WAS REMOVED AND REPLACED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8675756 ↗MalfunctionMay 7, 2019Defective Component; Defective Component
IT WAS REPORTED THAT UPON OPENING THE PACKAGE THE TIP OF THE TUNNELER WAS ALLEGEDLY BROKEN. REPORTEDLY, A NEW DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Patient Involvement
MAUDE report 8587111 ↗MalfunctionApr 25, 2019Defective Component; Defective Component
IT WAS REPORTED THAT UPON OPENING THE PACKAGE THE BLUE PART OF THE TUNNELER WAS ALLEGEDLY BROKEN. REPORTEDLY, A NEW KIT WAS OPENED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Patient Involvement
MAUDE report 8550274 ↗MalfunctionApr 18, 2019Component Incompatible; Component Missing; Component Incompatible; Component Missing
IT WAS REPORTED THAT DURING PREPARATION FOR THE CHRONIC CATHETER PLACEMENT, TWO KITS WERE OPENED AND THE PEEL-AWAY SHEATHS WERE ALLEGEDLY INCOMPATIBLE FOR THE PACKAGED CATHETER AND WERE RENDERED UNUSABLE. REPORTEDLY, ANOTHER TWO KITS WERE UNOPENED AND THE INCORRECT SIZE WAS ALLEGEDLY ABLE TO BE IDENTIFIED THROUGH THE CLEAR PACKAGING, THEREFORE, THE CUSTOMER ALLEGED THE FOUR KITS WERE PROVIDED WITH THE INCORRECT SIZE SHEATH. IT WAS FURTHER REPORTED THAT ANOTHER DEVICE WAS IMPLANTED. THERE WAS NO REPORTED PATIENT CONTACT.
Patient problems: No Patient Involvement
MAUDE report 8528525 ↗MalfunctionApr 16, 2019Obstruction of Flow; Malposition of Device; Obstruction of Flow; Malposition of Device
THIS REPORT SUMMARIZES THREE MALFUNCTION EVENTS. A REVIEW OF THE EVENTS INDICATED THAT MODEL EC500F ECLIPSE FILTER SYSTEM EXPERIENCED A MALPOSITION OF DEVICE. THESE REPORTS WERE RECEIVED FROM VARIOUS SOURCES. ALL THREE EVENTS HAD NO REPORTED PATIENT INJURY. OF THE THREE REPORTED PATIENT, 2 PATIENTS WERE MALE, ONE PATIENT WEIGHT WAS 267 LBS; HOWEVER, OTHER PATIENTS AGE, WEIGHT AND GENDER WERE NOT PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 8521072 ↗
Other Catheter, hemodialysis, implanted devices
- Palindrome 8888145014Mozarc Medical US LLC · Catheter, hemodialysis, implanted1 recall312 reports
- Palindrome 8888145015Mozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 death219 reports
- HemoStar Long-Term Hemodialysis Catheter 5833690Bard Access Systems, Inc. · Catheter, hemodialysis, implanted2 deaths290 reports
- Palindrome 8888119369PMozarc Medical US LLC · Catheter, hemodialysis, implanted3 recalls1 death14 reports
- HemoStar Long-Term Hemodialysis Catheter 5833730Bard Access Systems, Inc. · Catheter, hemodialysis, implanted1 death141 reports
- GlidePath Long-Term Hemodialysis Catheter with Preloaded Stylet 5393230Bard Access Systems, Inc. · Catheter, hemodialysis, implanted5 deaths138 reports
- Palindrome 8888123404PMozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 death17 reports
- Palindrome 8888119364PMozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 death17 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page